Smith & Nephew Competitor Duo Knee Femoral

K052265510(k) AbbreviatedCleared Dec 27, 2005130-day reviewClass II21 CFR 888.3560Recalls 42

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · NPJ Prosthesis, Knee Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Dec 27, 2005 as substantially equivalent to K042896 (Smith & Nephew, Inc., 2005) and 2 other predicates after a 130-day review. Its family tree reaches back 3 generations to 1996. 2 later devices descend from it. It has been recalled 42 times.

Decided
Dec 27, 2005
Review
130 days
Class
Class II
Product code
NPJ
Recalls
42
Predicates
3

Its longest line of descent runs back 3 generations to K962137 (1996). 7 ancestors in all, the oldest K951987 (1995). 2 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2005: 1 other ancestor1996: 2 other ancestors1995: 1 other ancestor2006: 1 descendant cleared2012: 1 descendant clearedK962137 · Genesis II Constrained System · Smith & Nephew Richards, Inc. · 1996 · recalledK041106 · Revision Knee System · Smith & Nephew, Inc. · 2004 · recalledK043440 · Smith & Nephew, Inc. Revision Knee System · Smith & Nephew, Inc. · 2005 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors2 generations back: 2 other ancestors 1 generation forward: 1 descendant2 generations forward: 1 descendant K043440 · Smith & Nephew, Inc. Revision Knee System · Smith & Nephew, Inc. · 2005 · recalledK041106 · Revision Knee System · Smith & Nephew, Inc. · 2004 · recalledK962137 · Genesis II Constrained System · Smith & Nephew Richards, Inc. · 1996 · recalled0
  1. Root · 3 generations backK962137 Genesis II Constrained SystemSmith & Nephew Richards, Inc. · 1996 · Recalled
  2. 2 generations backK041106 Revision Knee SystemSmith & Nephew, Inc. · 2004 · Recalled
  3. 1 generation backK043440 Smith & Nephew, Inc. Revision Knee SystemSmith & Nephew, Inc. · 2005 · Recalled
  4. This deviceK052265 Smith & Nephew Competitor Duo Knee FemoralSmith & Nephew, Inc. · 2005
  5. Descendant · cites it directly · 1 of its ownK061569 Smith & Nephew Competitor Deuce Femoral ComponentsSmith & Nephew, Inc. · 2006
  6. Descendant · 2 generations on · 0 of its ownK113038 Smith & Nephew Inc. Unicompartmental and Biocompartmental Knee…Smith & Nephew, Inc. · 2012

Most influential descendants

Showing 2 of 2 · See all
  • K061569Smith & Nephew Competitor Deuce Femoral ComponentsSmith & Nephew, Inc. · 2006 · cites it directly1descendant
  • K113038Smith & Nephew Inc. Unicompartmental and Biocompartmental Knee System InstrumentsSmith & Nephew, Inc. · 2012 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K962137Genesis II Constrained SystemSmith & Nephew Richards, Inc.19961
−2K041106Revision Knee SystemSmith & Nephew, Inc.20041
−1K043440Smith & Nephew, Inc. Revision Knee SystemSmith & Nephew, Inc.20057
0 (this device)K052265Smith & Nephew Competitor Duo Knee FemoralSmith & Nephew, Inc.2005—
+1K061569Smith & Nephew Competitor Deuce Femoral ComponentsSmith & Nephew, Inc.2006—
+2K113038Smith & Nephew Inc. Unicompartmental and Biocompartmental Knee System…Smith & Nephew, Inc.2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510
199620
200510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201210
FIG. 1 — Lineage of K052265, 1996–2012Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NPJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K042896Smith & Nephew Hybrid Knee Femoral
Summary page 1
…the existing Hybrid designs of the Competitor Duo femoral component are based upon Femoral Component subject of K042896. C. SUBSTANTIAL EQUIVALENCE INFORMATION:…
Smith & Nephew, Inc.2005—
*K962557Genesis II Zirconium Femoral Component/Profix Zirconium Femoral Component
Summary page 1
…K042896) K043440) • Genesis II and Legion (Revision) Knee Femorals (Oxinium Material) (K962557 and Public Health Se…
Smith & Nephew Richards, Inc.19961
*K043440Smith & Nephew, Inc. Revision Knee System
Summary page 1
…s, overall indications, and materials: * Smith & Nephew Hybrid Knee Femoral (K042896) K043440) • Genesis II and Legion (Revision) Knee Femorals (Oxinium Material) (K962557 and…
Smith & Nephew, Inc.20057

Also cleared against K042896: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061569Smith & Nephew Competitor Deuce Femoral ComponentsSmith & Nephew, Inc.20061

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 23, 2009Not classifiedZ-0925-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0926-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0928-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0929-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0930-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0932-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0933-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0934-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0937-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0942-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0944-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0945-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0949-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0951-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0952-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0954-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0957-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0959-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0938-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0946-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0953-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0960-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0961-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0962-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0936-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0940-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0943-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0955-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0958-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0964-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0924-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0927-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0931-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0935-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0939-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0941-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0947-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0948-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0950-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0956-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0963-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.
Jan 23, 2009Not classifiedZ-0965-2010Device DesignThe firm has received complaints of the base plates breaking in situ. If a breakage occurs, the broken baseplate must be revised and/or changed. If left unrevised a broken base could lead to instability and premature wear of the construct.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 9 years and passes through 2 companies; a new generation was cleared every 10 months on median.
  • 5 of the 7 ancestors have a recall of any class.
  • Of its descendants, K061569 (Smith & Nephew, Inc., 2006) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K042896.

How does its category compare?

All NPJ devices

NPJ Prosthesis, Knee Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NPJ

1in 2015

2007: 12009: 32013: 12015: 1
2007high 3 (2009) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NPJ

76in 2009

2009: 76 days median
2007high 76 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · NPJ

1in 2016

2011: 12015: 32016: 12026 (partial year): 1
2007high 3 (2015) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NPJ

47in 2025

2008: 12009: 32010: 132011: 322012: 722013: 982014: 752015: 552016: 932017: 1502018: 1412019: 912020: 652021: 512022: 322023: 522024: 442025: 47
2007high 150 (2017) · low 0 (2007)2025

MAUDE reports naming product code NPJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—00
20080—01
200937603
20100—013
20110—132
20120—072
20131—098
20140—075
20151—355
20160—193
20170—0150
20180—0141
20190—091
20200—065
20210—051
20220—032
20230—052
20240—044
20250—047
20260—1—

What is it indicated for?

The Competitor Duo Knee Femoral Components are intended to be used for those patients whereby conditions exist that can not be solely addressed by a device that treats a single compartment (i.e. unicondylar or patellofemoral prosthesis) of the knee. Indications include: * post-traumatic arthritis; * degenerative arthritis; and * failed osteotomies, hemiarthroplasties; and unicompartmental replacement These indications will be used for the Competitor Duo…

Read the full statement (103 words)

The Competitor Duo Knee Femoral Components are intended to be used for those patients whereby conditions exist that can not be solely addressed by a device that treats a single compartment (i.e. unicondylar or patellofemoral prosthesis) of the knee.

Indications include:

* post-traumatic arthritis;

* degenerative arthritis; and

* failed osteotomies, hemiarthroplasties; and unicompartmental replacement

These indications will be used for the Competitor Duo Knee Femoral Components, whereby the medial condyle and patellofemoral regions have been affected by one or more of these conditions

The Competitor Duo Knee Femoral Components are single use only and are intended for implantation only with bone cement

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K052265: Smith & Nephew Competitor Duo Knee Femoral.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K052265

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026