510(k) Report

510(k) · Cleared

Cleared

Color Enhancement Function for the Ssa-380A

K963705510(k) TraditionalCleared Dec 2, 199677-day reviewClass II21 CFR 892.1550

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Toshiba America Medical Systems, In.C · IYN System, Imaging, Pulsed Doppler, Ultrasonic

Cleared on Dec 2, 1996 as substantially equivalent to K933743 (Toshiba America Medical Systems, In.C, 1994) after a 77-day review. Its family tree reaches back 1 generation to 1994. 3 of the 405 devices in its family, counting itself, have a Class I recall. 403 later devices descend from it.

Decided
Dec 2, 1996
Review
77 days
Class
Class II
Product code
IYN
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K933743 (1994). 403 later devices descend from it, 2 citing it directly. 3 of the 405 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 3 descendants cleared2005: 4 descendants cleared2001: 2 descendants cleared2002: 1 descendant cleared2003: 2 descendants cleared2007: 7 descendants cleared2009: 11 descendants cleared2004: 2 descendants cleared2006: 1 descendant cleared2013: 6 descendants cleared2021: 24 descendants cleared2012: 12 descendants cleared2015: 13 descendants cleared2016: 18 descendants cleared2011: 5 descendants cleared2014: 18 descendants cleared2022: 28 descendants cleared2010: 2 descendants cleared2008: 6 descendants cleared2020: 35 descendants cleared, 1 with a Class I recall12019: 18 descendants cleared2025: 48 descendants cleared, 2 with a Class I recall22023: 30 descendants cleared2018: 28 descendants cleared2017: 22 descendants cleared2026: 32 descendants cleared2024: 25 descendants clearedK933743 · SSA-380A · Toshiba America Medical Systems, In.C · 19941990200020102020
Descendants, by generation 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 6 descendants4 generations forward: 10 descendants5 generations forward: 15 descendants6 generations forward: 26 descendants7 generations forward: 35 descendants8 generations forward: 38 descendants, 1 with a Class I recall19 generations forward: 46 descendants10 generations forward: 45 descendants, 2 with a Class I recall211 generations forward: 35 descendants12 generations forward: 38 descendants13 generations forward: 27 descendants14 generations forward: 24 descendants15 generations forward: 19 descendants16 generations forward: 12 descendants17 generations forward: 8 descendants18 generations forward: 7 descendants19 generations forward: 4 descendants20 generations forward: 3 descendants K933743 · SSA-380A · Toshiba America Medical Systems, In.C · 19940+5+10+15+20
  1. Root · 1 generation backK933743 SSA-380AToshiba America Medical Systems, In.C · 1994
  2. This deviceK963705 Color Enhancement Function for the Ssa-380AToshiba America Medical Systems, In.C · 1996
  3. Descendant · cites it directly · 212 of its ownK991858 Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C · 1999
  4. Descendant · 2 generations on · 211 of its ownK013633 Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C · 2001
  5. Descendant · 3 generations on · 210 of its ownK022400 Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2002 · Recalled
  6. Descendant · 3 generations on · 209 of its ownK032281 Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003
  7. Descendant · 4 generations on · 205 of its ownK063130 Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · Recalled

Showing 5 of 403 descendants. See all descendants

Descendants cleared per year

1999: 32001: 22002: 12003: 22004: 22005: 42006: 12007: 72008: 62009: 112010: 22011: 52012: 122013: 62014: 182015: 132016: 182017: 222018: 282019: 182020: 352021: 242022: 282023: 302024: 252025: 482026: 32
1999high 48 (2025) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 403 · See all
  • K991858Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C · 1999 · cites it directly212descendants
  • K013633Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C · 2001 · 2 generations on211descendants
  • K022400Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2002 · 3 generations on210descendants
  • K032281Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C · 2003 · 3 generations on209descendants
  • K063130Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation · 2006 · 4 generations on205descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K933743SSA-380AToshiba America Medical Systems, In.C1994—
0 (this device)K963705Color Enhancement Function for the Ssa-380AToshiba America Medical Systems, In.C1996—
+1K991858Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C1999—
+2K013633Ssa-770A, Aplio Ultrasound SystemToshiba America Medical Systems, In.C2001—
+3K022400Ssa-700A, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C20021
+3K032281Ssa-770A - Version 4, Aplio Diagnostic Ultrasound SystemToshiba America Medical Systems, In.C2003—
+4K063130Aplio XG Diagnostic Ultrasound System, Model Ssa-790A, Version 1.0Toshibamedical Systems Corporation20061

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199930
200120
200210
200320
200420
200540
200610
200770
200860
2009110
201020
201150
2012120
201360
2014180
2015130
2016180
2017220
2018280
2019180
2020351
2021240
2022280
2023300
2024250
2025482
2026320
FIG. 1 — Lineage of K963705, 1994–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IYN device

  1. #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K933743SSA-380A
Summary page 1
…on of the SSA-380A Substantial Equivalence Summary: Color Angio is a post-processing modification to the Power Mode display that was cleared with the SSA-380A system in K933743. It is an amplitude based color display subset of color doppler imaging that performs frame addition to accentuate minute vascula…
Toshiba America Medical Systems, In.C1994—

Also cleared against K933743: 6 other devices →

What descends from it? (403)

See all 403

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K991858Ssa-390A/Power Vision 8000Toshiba America Medical Systems, In.C1999212
K991710Ssa-370A/Powervision 6000Toshiba America Medical Systems, In.C1999203

Has anything in its family been recalled? (3 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K193517ViziShot 2 FLEXOlympus Surgical Technologies America · 2020 · 8 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K242670Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 10 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall
  • K242519Lumify Diagnostic Ultrasound SystemPhilips Ultrasound · 2025 · 10 generations on · ◆ in the 2025 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K991858 (Toshiba America Medical Systems, In.C, 1999) has gone furthest, with 212 descendants of its own.
  • 6 other devices also cite its predicate K933743.

How does its category compare?

All IYN devices

IYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IYN

57in 2025

2007: 132008: 252009: 182010: 292011: 392012: 432013: 512014: 562015: 542016: 602017: 642018: 592019: 472020: 572021: 462022: 532023: 442024: 502025: 572026 (partial year): 44
2007high 64 (2017) · low 13 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IYN

124in 2025

2007: 26 days median2008: 18 days median2009: 29 days median2010: 67 days median2011: 46 days median2012: 37 days median2013: 60 days median2014: 39 days median2015: 55 days median2016: 56 days median2017: 83 days median2018: 85 days median2019: 79 days median2020: 72 days median2021: 105 days median2022: 105 days median2023: 115 days median2024: 119 days median2025: 124 days median2026 (partial year): 107 days median
2007high 124 (2025) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IYN

10in 2025

2007: 162008: 122009: 82010: 102011: 82012: 32013: 82014: 22015: 32016: 22018: 72020: 42021: 82022: 102023: 42024: 12025: 102026 (partial year): 2, 1 Class I
2007high 16 (2007) · low 0 (2017)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · IYN

610in 2025

2007: 32008: 22009: 92010: 172011: 182012: 202013: 182014: 482015: 1162016: 1452017: 1462018: 852019: 592020: 672021: 2132022: 1862023: 2292024: 2662025: 610
2007high 610 (2025) · low 2 (2008)2025

MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071326163
20082518122
2009182989
201029671017
20113946818
20124337320
20135160818
20145639248
201554553116
201660562145
201764830146
20185985785
20194779059
20205772467
2021461058213
20225310510186
2023441154229
2024501191266
20255712410610
2026441072 (1)—
Cite this page

510(k) Report. “K963705: Color Enhancement Function for the Ssa-380A.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K963705

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026