510(k) · Cleared
Recalled · Class II · 2013Eztest
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Sgm Biotech, Inc. · FRC Indicator, Biological Sterilization Process
Cleared on Feb 25, 1997 as substantially equivalent to K930682 (Sgm Biotech, Inc., 1994) and 2 other predicates after a 153-day review. Its family tree reaches back 1 generation to 1980. 12 later devices descend from it. It has been recalled once.
- Decided
- Feb 25, 1997
- Review
- 153 days
- Class
- Class II
- Product code
- FRC
- Recalls
- 1
- Predicates
- 3
Its longest line of descent runs back 1 generation to K801109 (1980). 3 ancestors in all. 12 later devices descend from it, 5 citing it directly.
- Root · 1 generation backK801109 Proof (Biological/Chemical Indicator)Amsco Co. · 1980
- This deviceK963841 EztestSgm Biotech, Inc. · 1997
- Descendant · cites it directly · 3 of its ownK070595 Sporview Steam Self-Contained Biological IndicatorSpsmedical Supply Corp. · 2007
- Descendant · 2 generations on · 2 of its ownK111515 Spsmedical Sporview Steam BISpsmedical Supply Corp. · 2012
- Descendant · cites it directly · 2 of its ownK071895 Verify 275 Degree F. 3-10 Indicator; 10 Indicator; 3 IndicatorSTERIS Corporation · 2007
- Descendant · cites it directly · 2 of its ownK041386 Raven Protest - SteamRaven Biological Laboratories, Inc. · 2004
- Descendant · 2 generations on · 1 of its ownK122024 Spsmedical Sporview 10 Steam Self Contained Biological IndicatorSpsmedical Supply Corp. · 2013
Showing 5 of 12 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 12 · See all- K070595Sporview Steam Self-Contained Biological IndicatorSpsmedical Supply Corp. · 2007 · cites it directly3descendants
- K111515Spsmedical Sporview Steam BISpsmedical Supply Corp. · 2012 · 2 generations on2descendants
- K071895Verify 275 Degree F. 3-10 Indicator; 10 Indicator; 3 IndicatorSTERIS Corporation · 2007 · cites it directly2descendants
- K041386Raven Protest - SteamRaven Biological Laboratories, Inc. · 2004 · cites it directly2descendants
- K122024Spsmedical Sporview 10 Steam Self Contained Biological IndicatorSpsmedical Supply Corp. · 2013 · 2 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K801109 | Proof (Biological/Chemical Indicator) | Amsco Co. | 1980 | — |
| 0 (this device) | K963841 | Eztest | Sgm Biotech, Inc. | 1997 | — |
| +1 | K070595 | Sporview Steam Self-Contained Biological Indicator | Spsmedical Supply Corp. | 2007 | — |
| +2 | K111515 | Spsmedical Sporview Steam BI | Spsmedical Supply Corp. | 2012 | — |
| +1 | K071895 | Verify 275 Degree F. 3-10 Indicator; 10 Indicator; 3 Indicator | STERIS Corporation | 2007 | — |
| +1 | K041386 | Raven Protest - Steam | Raven Biological Laboratories, Inc. | 2004 | — |
| +2 | K122024 | Spsmedical Sporview 10 Steam Self Contained Biological Indicator | Spsmedical Supply Corp. | 2013 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1990 | 1 | 0 |
| 1994 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2004 | 1 | 0 |
| 2007 | 2 | 0 |
| 2009 | 2 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 2 | 0 |
| 2013 | 1 | 0 |
| 2017 | 1 | 0 |
| 2020 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in FRC? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FRC device
- #1 K250044 Celerity 5 HP Biological Indicator (LCB052)…score 14.5 · 10 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (4)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K930682 | Eztest-SteamSummary page 2…r steam [as required by Section 807.92(a)(2)] 11-2 27 SUBSTANTIALLY EQUIVALENT TO THE FOLLOWING LEGALLY MARKETED DEVICES: EZTest® - SGM Biotech, Inc. (K930682) ATITEST® - PyMaH Corp. (KS02344) ATTEST® - 3M Co. (pre 1976 device) PROOF® - AMSCO (K801109) [as required by Section 807.92(a)(… | 1994 |
| * | K902344 | Ati Disposable Biological Test Pack for Steam SterSummary page 5…this device. The difference is the extension of the label claims for use in high temperature saturated steam processes of 132°C, 134°C and 135°C. B. ATITest® PyMaH/ATI (K902344): This device is exactly the same as the PyMaH ATITest® disclosed in 510(k) notification K302344. The ATITest® was developed by… | 1990 |
| * | K801109 | Proof (Biological/Chemical Indicator)Summary page 2…ENT TO THE FOLLOWING LEGALLY MARKETED DEVICES: EZTest® - SGM Biotech, Inc. (K930682) ATITEST® - PyMaH Corp. (KS02344) ATTEST® - 3M Co. (pre 1976 device) PROOF® - AMSCO (K801109) [as required by Section 807.92(a)(3)] DESCRIPTION OF DEVICE: EZTest® - Steam [as required by Section 807.92(a)(4)] The biolo… | 1980 |
| * | — | Attest® Named without a submission number |
What descends from it? (12)
See all 125 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 4, 2013 | Class II | Z-1867-2013 | Nonconforming Material/Component | Mesa EZTest , Self-contained steam biological indicator, are recalled because Mesa recently determined that the read out time is not in specification with the label claims. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All FRC devicesFRC Indicator, Biological Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FRC
6in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FRC
29in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · FRC
8so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · FRC
9in 2025
MAUDE reports naming product code FRC, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 83 | 0 | 100 |
| 2008 | 1 | — | 0 | 60 |
| 2009 | 6 | 183 | 1 | 45 |
| 2010 | 4 | 165 | 0 | 57 |
| 2011 | 6 | 118 | 0 | 37 |
| 2012 | 9 | 178 | 2 | 46 |
| 2013 | 4 | 260 | 2 | 39 |
| 2014 | 13 | 174 | 0 | 51 |
| 2015 | 3 | 130 | 0 | 182 |
| 2016 | 2 | — | 0 | 482 |
| 2017 | 10 | 130 | 0 | 182 |
| 2018 | 18 | 97 | 0 | 22 |
| 2019 | 5 | 88 | 0 | 30 |
| 2020 | 10 | 150 | 0 | 16 |
| 2021 | 5 | 107 | 0 | 10 |
| 2022 | 5 | 193 | 4 | 9 |
| 2023 | 6 | 190 | 1 | 20 |
| 2024 | 8 | 105 | 0 | 22 |
| 2025 | 6 | 29 | 0 | 9 |
| 2026 | 1 | — | 8 | — |
What is it indicated for?
EZTest® - Steam self contained biological indicators are for monitoring the efficacy of saturated steam sterilization processors. Performance characteristics are established in accordance with USP XXiil for the 121 C - steam process. Additional saturated steam sterilization conditions are also included in the Certificate of Performance. EZTest® - Steam Bls are also appropriate for use in the high temperature saturated steam processes of 132C, 134 C and 135C.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K963841 · printed Nov 4, 2026
Cite this page
510(k) Report. “K963841: Eztest.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K963841
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026