Combination Oximeter/Carbon Dioxide Gas Analyzer

K964239510(k) TraditionalCleared May 9, 1997198-day reviewClass II21 CFR 868.1400

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Spegas Industries, Ltd. · CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Cleared on May 9, 1997 as substantially equivalent to K950387 (Spegas Industries, Ltd., 1995) after a 198-day review. Its family tree reaches back 1 generation to 1995. 2 of the 44 devices in its family, counting itself, have a Class I recall. 42 later devices descend from it.

Decided
May 9, 1997
Review
198 days
Class
Class II
Product code
CCK
Recalls
None
Predicates
1

Its longest line of descent runs back 1 generation to K950387 (1995). 42 later devices descend from it, 7 citing it directly. 2 of the 44 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 1999: 2 descendants cleared1998: 3 descendants cleared2004: 3 descendants cleared2003: 1 descendant cleared2007: 1 descendant cleared2009: 3 descendants cleared2019: 3 descendants cleared2001: 1 descendant cleared2015: 5 descendants cleared2012: 1 descendant cleared2014: 3 descendants cleared, 1 with a Class I recall12008: 1 descendant cleared2010: 2 descendants cleared2023: 1 descendant cleared, 1 with a Class I recall12018: 2 descendants cleared2011: 2 descendants cleared2026: 2 descendants cleared2016: 1 descendant cleared2020: 1 descendant cleared2021: 1 descendant cleared2025: 1 descendant cleared2024: 2 descendants clearedK950387 · Gemini · Spegas Industries, Ltd. · 19951995200020052010201520202025
Descendants, by generation 1 generation forward: 7 descendants2 generations forward: 9 descendants, 1 with a Class I recall13 generations forward: 9 descendants, 1 with a Class I recall14 generations forward: 7 descendants5 generations forward: 8 descendants6 generations forward: 2 descendants K950387 · Gemini · Spegas Industries, Ltd. · 19950+5
  1. Root · 1 generation backK950387 GeminiSpegas Industries, Ltd. · 1995
  2. This deviceK964239 Combination Oximeter/Carbon Dioxide Gas AnalyzerSpegas Industries, Ltd. · 1997
  3. Descendant · cites it directly · 20 of its ownK024300 Microcap Plus/NPB-75Oridion Medical 1987, Ltd. · 2003
  4. Descendant · 2 generations on · 12 of its ownK063752 Lifesense, Model LS1-9RNonin Medical, Inc. · 2007
  5. Descendant · cites it directly · 11 of its ownK983635 Julian Anesthesia Workstation, Model # JulianDraeger Medical, Inc. · 1999
  6. Descendant · 2 generations on · 10 of its ownK042607 Primus USDraeger Medical, Inc. · 2004 · Recalled
  7. Descendant · 3 generations on · 7 of its ownK092101 Reusable Pulse Oximeter Sensor, Model 8000 SeriesNonin Medical, Inc. · 2009

Showing 5 of 42 descendants. See all descendants

Descendants cleared per year

1998: 31999: 22001: 12003: 12004: 32007: 12008: 12009: 32010: 22011: 22012: 12014: 32015: 52016: 12018: 22019: 32020: 12021: 12023: 12024: 22025: 12026: 2
1998high 5 (2015) · low 0 (2000)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 42 · See all
  • K024300Microcap Plus/NPB-75Oridion Medical 1987, Ltd. · 2003 · cites it directly20descendants
  • K063752Lifesense, Model LS1-9RNonin Medical, Inc. · 2007 · 2 generations on12descendants
  • K983635Julian Anesthesia Workstation, Model # JulianDraeger Medical, Inc. · 1999 · cites it directly11descendants
  • K042607Primus USDraeger Medical, Inc. · 2004 · 2 generations on10descendants
  • K092101Reusable Pulse Oximeter Sensor, Model 8000 SeriesNonin Medical, Inc. · 2009 · 3 generations on7descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K950387GeminiSpegas Industries, Ltd.1995—
0 (this device)K964239Combination Oximeter/Carbon Dioxide Gas AnalyzerSpegas Industries, Ltd.1997—
+1K024300Microcap Plus/NPB-75Oridion Medical 1987, Ltd.2003—
+2K063752Lifesense, Model LS1-9RNonin Medical, Inc.2007—
+1K983635Julian Anesthesia Workstation, Model # JulianDraeger Medical, Inc.1999—
+2K042607Primus USDraeger Medical, Inc.20042
+3K092101Reusable Pulse Oximeter Sensor, Model 8000 SeriesNonin Medical, Inc.2009—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199830
199920
200110
200310
200430
200710
200810
200930
201020
201120
201210
201431
201550
201610
201820
201930
202010
202110
202311
202420
202510
202620
FIG. 1 — Lineage of K964239, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new CCK device

  1. #1 K253030 Capnostream™35 Portable Respiratory Monitor…score 15.8 · 16 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K950387Gemini
Summary page 2
…ISTING FDA FORM 2892: A 733250 DOC: 2.1 MICROCAP/NPB-75 Sectiion 2.1.1 DEVICE DESCRIPTION The Spegas MicroCap/NPB-75 and the comparison device Gemini+ (K950387) are integrated instruments that measure end tidal CO2 (EtCO») and saturated arterial oxygen (SpO2). As a derivative of the EtCO…
Spegas Industries, Ltd.1995—

What descends from it? (42)

See all 42

7 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K024300Microcap Plus/NPB-75Oridion Medical 1987, Ltd.200320
K983635Julian Anesthesia Workstation, Model # JulianDraeger Medical, Inc.199911
K980325Microstream Nasal Cannula FilterlineOridion Medical, Ltd.19982
K980327Microstream Filterline IcuOridion Medical, Ltd.19982
K980324Microstream Fillerline or/EmsOridion Medical, Ltd.19981
K031741Medley ETCO2 Module (Includes Accessories), Model 8300Alaris Medical Systems, Inc.20040
K982969Nonin Pulse Oximeter and Carbon Dioxide Detector, Models 9843, 9845 and 9847Nonin Medical, Inc.19990

Has anything in its family been recalled? (2 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K133886Perseus A500Draeger Medical GmbH · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 12 Class I recalls
  • K230931AtlanDrägerwerk AG Co. Kgaa · 2023 · 3 generations on · ◆ in the 2023 descendants bar of FIG. 18 Class I recalls

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K024300 (Oridion Medical 1987, Ltd., 2003) has gone furthest, with 20 descendants of its own.

How does its category compare?

All CCK devices

CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · CCK

3in 2025

2007: 52008: 32009: 32010: 32011: 22012: 72013: 42014: 92015: 52016: 22017: 52018: 62019: 72020: 42021: 42022: 32023: 42024: 32025: 32026 (partial year): 3
2007high 9 (2014) · low 2 (2011)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · CCK

169in 2025

2007: 56 days median2008: 248 days median2009: 101 days median2010: 125 days median2012: 150 days median2013: 132 days median2014: 166 days median2015: 105 days median2017: 134 days median2018: 296 days median2019: 261 days median2020: 188 days median2021: 227 days median2022: 226 days median2023: 555 days median2024: 592 days median2025: 169 days median2026 (partial year): 224 days median
2007high 592 (2024) · low 56 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · CCK

1in 2022

2008: 12009: 192010: 42011: 142012: 32013: 2, 1 Class I2015: 12018: 52019: 7, 7 Class I2020: 12022: 1
2007high 19 (2009) · low 0 (2007)2026

Class I (8 since 2007) Class II/III not classified

Adverse event reports per year · CCK

1,259in 2025

2007: 72008: 202009: 212010: 122011: 182012: 132013: 282014: 512015: 1142016: 882017: 582018: 632019: 1412020: 2,3022021: 2,2722022: 2,4752023: 1,6372024: 1,2162025: 1,259
2007high 2,475 (2022) · low 7 (2007)2025

MAUDE reports naming product code CCK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200755607
20083248120
200931011921
20103125412
20112—1418
20127150313
201341322 (1)28
20149166051
201551051114
20162—088
20175134058
20186296563
201972617 (7)141
2020418812,302
2021422702,272
2022322612,475
2023455501,637
2024359201,216
2025316901,259
202632240—
Cite this page

510(k) Report. “K964239: Combination Oximeter/Carbon Dioxide Gas Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K964239

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026