510(k) · Cleared
Recalled · Class I · 2026Primary IV Set
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Icu Medical, Inc. · FMG Stopcock, I.V. Set
Cleared on Apr 2, 1997 after a 148-day review. The FDA summary names its predicate without a number. 3 of the 18 devices in its family, counting itself, have a Class I recall. 17 later devices descend from it. It has been recalled 188 times, including 5 Class I recalls.
- Decided
- Apr 2, 1997
- Review
- 148 days
- Class
- Class II
- Product code
- FMG
- Recalls
- 188 (5 Class I)
- Predicates
- 0
17 later devices descend from it, 5 citing it directly. 3 of the 18 devices in this family, counting this one, have a Class I recall.
- This deviceK964435 Primary IV SetIcu Medical, Inc. · 1997
- Descendant · cites it directly · 12 of its ownK081361 Chemoclave Cytotoxic Medication Preparation and Delivery System…Icu Medical, Inc. · 2008 · Recalled (Class I)
- Descendant · 2 generations on · 10 of its ownK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Recalled (Class I)
- Descendant · 3 generations on · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
- Descendant · 4 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
- Descendant · 4 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021
Showing 5 of 17 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 17 · See all- K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · cites it directly12descendants
- K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on10descendants
- K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · 3 generations on9descendants
- K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 4 generations on3descendants
- K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 4 generations on1descendant
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K964435 | Primary IV Set | Icu Medical, Inc. | 1997 | — |
| +1 | K081361 | Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX | Icu Medical, Inc. | 2008 | Class I |
| +2 | K131549 | Chemolock, Chemolock Port | Icu Medical, Inc. | 2013 | Class I |
| +3 | K141448 | Tevadaptor Closed Drug Reconstitution and Transfer System | Teva Medical Ltd., Migada Plant | 2015 | — |
| +4 | K192866 | Chemfort CSTD | Simplivia Healthcare, Ltd. | 2020 | 1 |
| +4 | K201422 | Arisure Closed System Drug Transfer Device (CSTD) | Yukon Medical, LLC | 2021 | — |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 1 | 0 |
| 2008 | 2 | 1 |
| 2009 | 1 | 0 |
| 2010 | 1 | 0 |
| 2013 | 1 | 1 |
| 2015 | 1 | 0 |
| 2017 | 1 | 0 |
| 2018 | 2 | 0 |
| 2020 | 1 | 0 |
| 2021 | 2 | 0 |
| 2023 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new FMG device
- #1 K242255 Qitexio® 4-Way Stopcock (QIT014)score 10.5 · 5 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | Medical Product Specialists I.V./Secondary Sets Named without a submission number | |
| * | — | McGaw Intravenous Administration Sets Named without a submission number | |
| * | — | Churchill Medical Systems IV/Extension Sets Named without a submission number |
What descends from it? (17)
See all 175 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K081361 | Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX Class I | 2008 | 12 |
| K030246 | Victus I.V. Administration Set, Models 27071 and 27072 | 2003 | 1 |
| K093271 | Extension Set (Standardbore Tubing), Extension Set (Minibore Tubing), IV Administration… | 2010 | 0 |
| K082836 | Icu Medical Universal Single-Use Spikes | 2009 | 0 |
| K080007 | Graduated Adapter with Clave | 2008 | 0 |
Has anything in its family been recalled? (3 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| May 12, 2026 | Class II | Z-3163-2026 | Process control | Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration… |
| May 12, 2026 | Class II | Z-3157-2026 | Process control | Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid… |
| Oct 2, 2025 | Class I | Z-0154-2026 | Process control | IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused… |
| Aug 1, 2022 | Class II | Z-1721-2022 | Process control | Due to a manufacturing defect which results in a visible gap between the connector's top and bottom housing. |
| Dec 2, 2020 | Class II | Z-0955-2021 | Process control | Due to the incorrect filter assembly. |
| Dec 2, 2020 | Class II | Z-0956-2021 | Process control | Due to the incorrect filter assembly. |
| Dec 2, 2020 | Class II | Z-0957-2021 | Process control | Due to the incorrect filter assembly. |
| Dec 2, 2020 | Class II | Z-0958-2021 | Process control | Due to the incorrect filter assembly. |
| Jul 6, 2020 | Class II | Z-0207-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0209-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0254-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0242-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0208-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0251-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0184-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0196-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0256-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0222-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0228-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0194-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0186-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0202-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0237-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0243-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0205-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0245-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0241-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0203-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0244-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0190-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0210-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0188-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0253-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0187-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0255-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0211-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0258-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0185-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0223-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0198-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0235-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0191-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0239-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0226-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0227-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0195-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0229-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0206-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0199-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
| Jul 6, 2020 | Class II | Z-0231-2021 | Process control | Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike. |
The 5 latest of 50 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
- K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
- Of its descendants, K081361 (Icu Medical, Inc., 2008) has gone furthest, with 12 descendants of its own.
How does its category compare?
All FMG devicesFMG Stopcock, I.V. Set: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · FMG
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · FMG
Too few FMG 510(k)s in any year since 2007 to give a median.
Recalls per year · FMG
33in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · FMG
868in 2025
MAUDE reports naming product code FMG, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 5 | 26 |
| 2008 | 2 | — | 4 | 48 |
| 2009 | 2 | — | 0 | 27 |
| 2010 | 0 | — | 2 | 131 |
| 2011 | 0 | — | 8 | 95 |
| 2012 | 0 | — | 0 | 93 |
| 2013 | 2 | — | 0 | 117 |
| 2014 | 1 | — | 0 | 105 |
| 2015 | 0 | — | 0 | 96 |
| 2016 | 0 | — | 1 | 148 |
| 2017 | 0 | — | 2 | 114 |
| 2018 | 1 | — | 3 | 164 |
| 2019 | 2 | — | 0 | 400 |
| 2020 | 0 | — | 5 | 284 |
| 2021 | 0 | — | 0 | 245 |
| 2022 | 2 | — | 0 | 280 |
| 2023 | 2 | — | 0 | 689 |
| 2024 | 0 | — | 0 | 935 |
| 2025 | 1 | — | 33 (1) | 868 |
| 2026 | 0 | — | 3 | — |
What is it indicated for?
The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K964435 · printed Nov 4, 2026
Cite this page
510(k) Report. “K964435: Primary IV Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K964435
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026