Primary IV Set

K964435510(k) TraditionalCleared Apr 2, 1997148-day reviewClass II21 CFR 880.5440Recalls 188 (5 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Icu Medical, Inc. · FMG Stopcock, I.V. Set

Cleared on Apr 2, 1997 after a 148-day review. The FDA summary names its predicate without a number. 3 of the 18 devices in its family, counting itself, have a Class I recall. 17 later devices descend from it. It has been recalled 188 times, including 5 Class I recalls.

Decided
Apr 2, 1997
Review
148 days
Class
Class II
Product code
FMG
Recalls
188 (5 Class I)
Predicates
0

17 later devices descend from it, 5 citing it directly. 3 of the 18 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2003: 1 descendant cleared2009: 1 descendant cleared2008: 2 descendants cleared, 1 with a Class I recall12010: 1 descendant cleared2018: 2 descendants cleared2013: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared2017: 1 descendant cleared2020: 1 descendant cleared2021: 2 descendants cleared2023: 1 descendant cleared2025: 2 descendants cleared2024: 1 descendant cleared1995200020052010201520202025
  1. This deviceK964435 Primary IV SetIcu Medical, Inc. · 1997
  2. Descendant · cites it directly · 12 of its ownK081361 Chemoclave Cytotoxic Medication Preparation and Delivery System…Icu Medical, Inc. · 2008 · Recalled (Class I)
  3. Descendant · 2 generations on · 10 of its ownK131549 Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · Recalled (Class I)
  4. Descendant · 3 generations on · 9 of its ownK141448 Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015
  5. Descendant · 4 generations on · 3 of its ownK192866 Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · Recalled
  6. Descendant · 4 generations on · 1 of its ownK201422 Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2003: 12008: 22009: 12010: 12013: 12015: 12017: 12018: 22020: 12021: 22023: 12024: 12025: 2
2003high 2 (2008) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 17 · See all
  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · cites it directly12descendants
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on10descendants
  • K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant · 2015 · 3 generations on9descendants
  • K192866Chemfort CSTDSimplivia Healthcare, Ltd. · 2020 · 4 generations on3descendants
  • K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC · 2021 · 4 generations on1descendant
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K964435Primary IV SetIcu Medical, Inc.1997—
+1K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc.2008Class I
+2K131549Chemolock, Chemolock PortIcu Medical, Inc.2013Class I
+3K141448Tevadaptor Closed Drug Reconstitution and Transfer SystemTeva Medical Ltd., Migada Plant2015—
+4K192866Chemfort CSTDSimplivia Healthcare, Ltd.20201
+4K201422Arisure Closed System Drug Transfer Device (CSTD)Yukon Medical, LLC2021—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200821
200910
201010
201311
201510
201710
201820
202010
202120
202310
202410
202520
FIG. 1 — Lineage of K964435, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMG device

  1. #1 K242255 Qitexio® 4-Way Stopcock (QIT014)score 10.5 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Medical Product Specialists I.V./Secondary Sets
Named without a submission number
—
*—McGaw Intravenous Administration Sets
Named without a submission number
—
*—Churchill Medical Systems IV/Extension Sets
Named without a submission number
—

What descends from it? (17)

See all 17

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXX Class IIcu Medical, Inc.200812
K030246Victus I.V. Administration Set, Models 27071 and 27072Victus, Inc.20031
K093271Extension Set (Standardbore Tubing), Extension Set (Minibore Tubing), IV Administration…Victus, Inc.20100
K082836Icu Medical Universal Single-Use SpikesIcu Medical, Inc.20090
K080007Graduated Adapter with ClaveIcu Medical, Inc.20080

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 12, 2026Class IIZ-3163-2026Process controlIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration…
May 12, 2026Class IIZ-3157-2026Process controlIntravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid…
Oct 2, 2025Class IZ-0154-2026Process controlIV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused…
Aug 1, 2022Class IIZ-1721-2022Process controlDue to a manufacturing defect which results in a visible gap between the connector's top and bottom housing.
Dec 2, 2020Class IIZ-0955-2021Process controlDue to the incorrect filter assembly.
Dec 2, 2020Class IIZ-0956-2021Process controlDue to the incorrect filter assembly.
Dec 2, 2020Class IIZ-0957-2021Process controlDue to the incorrect filter assembly.
Dec 2, 2020Class IIZ-0958-2021Process controlDue to the incorrect filter assembly.
Jul 6, 2020Class IIZ-0207-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0209-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0254-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0242-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0208-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0251-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0184-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0196-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0256-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0222-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0228-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0194-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0186-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0202-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0237-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0243-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0205-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0245-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0241-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0203-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0244-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0190-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0210-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0188-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0253-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0187-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0255-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0211-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0258-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0185-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0223-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0198-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0235-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0191-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0239-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0226-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0227-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0195-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0229-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0206-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0199-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Jul 6, 2020Class IIZ-0231-2021Process controlIdentification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K081361Chemoclave Cytotoxic Medication Preparation and Delivery System, Model, CH-XXXXIcu Medical, Inc. · 2008 · Cites it directly · ◆ in the 2008 descendants bar of FIG. 12 Class I recalls
  • K131549Chemolock, Chemolock PortIcu Medical, Inc. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K081361 (Icu Medical, Inc., 2008) has gone furthest, with 12 descendants of its own.

How does its category compare?

All FMG devices

FMG Stopcock, I.V. Set: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FMG

1in 2025

2007: 22008: 22009: 22013: 22014: 12018: 12019: 22022: 22023: 22025: 1
2007high 2 (2007) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMG

Too few FMG 510(k)s in any year since 2007 to give a median.

Recalls per year · FMG

33in 2025

2007: 52008: 42010: 22011: 82016: 12017: 22018: 32020: 52025: 33, 1 Class I2026 (partial year): 3
2007high 33 (2025) · low 0 (2009)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · FMG

868in 2025

2007: 262008: 482009: 272010: 1312011: 952012: 932013: 1172014: 1052015: 962016: 1482017: 1142018: 1642019: 4002020: 2842021: 2452022: 2802023: 6892024: 9352025: 868
2007high 935 (2024) · low 26 (2007)2025

MAUDE reports naming product code FMG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—526
20082—448
20092—027
20100—2131
20110—895
20120—093
20132—0117
20141—0105
20150—096
20160—1148
20170—2114
20181—3164
20192—0400
20200—5284
20210—0245
20222—0280
20232—0689
20240—0935
20251—33 (1)868
20260—3—

What is it indicated for?

The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient's vascular system through the administration set's needle or catheter (which is inserted into a vein).

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K964435: Primary IV Set.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K964435

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026