510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Biliary Stent, Biliary Stent System, Biliary Endoprosthesis

K965147510(k) TraditionalCleared Jun 30, 1997189-day reviewClass II21 CFR 876.5010Recalls 23

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Boston Scientific Corp · FGE Stents, Drains And Dilators For The Biliary Ducts

Cleared on Jun 30, 1997 as substantially equivalent to K834468 (Microvasive, 1984) and 1 other predicate after a 189-day review. Its family tree reaches back 1 generation to 1984. 17 later devices descend from it. It has been recalled 23 times.

Decided
Jun 30, 1997
Review
189 days
Class
Class II
Product code
FGE
Recalls
23
Predicates
2

Its longest line of descent runs back 1 generation to K834468 (1984). 2 ancestors in all. 17 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1994: 1 other ancestor2010: 2 descendants cleared2020: 3 descendants cleared2014: 2 descendants cleared2019: 1 descendant cleared2022: 2 descendants cleared2025: 2 descendants cleared2024: 2 descendants cleared2023: 2 descendants cleared2026: 1 descendant clearedK834468 · Drainage Stent · Microvasive · 1984 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 5 descendants3 generations forward: 7 descendants4 generations forward: 3 descendants K834468 · Drainage Stent · Microvasive · 1984 · recalled0
  1. Root · 1 generation backK834468 Drainage StentMicrovasive · 1984 · Recalled
  2. This deviceK965147 Biliary Stent, Biliary Stent System, Biliary EndoprosthesisBoston Scientific Corp · 1997
  3. Descendant · cites it directly · 14 of its ownK101314 Advanix Biliary Stent with Naviflex RX Delivery SystemBoston Scientific Corp · 2010
  4. Descendant · 2 generations on · 4 of its ownK133700 Advanix Pancreatic Stent and Naviflex Rapid Exchange Pancreatic…Boston Scientific Corporation · 2014 · Recalled
  5. Descendant · 2 generations on · 3 of its ownK193600 EndoGI Biliary Stent SystemEndogi Medical, Ltd. · 2020
  6. Descendant · 2 generations on · 3 of its ownK140630 Wallflex Biliary RX Stent SystemBoston Scientific · 2014 · Recalled
  7. Descendant · 3 generations on · 2 of its ownK202477 EndoGI S-Path Biliary Stent SystemEndo GI Medical · 2020

Showing 5 of 17 descendants. See all descendants

Descendants cleared per year

2010: 22014: 22019: 12020: 32022: 22023: 22024: 22025: 22026: 1
2010high 3 (2020) · low 0 (2011)2026

Most influential descendants

Showing 5 of 17 · See all
  • K101314Advanix Biliary Stent with Naviflex RX Delivery SystemBoston Scientific Corp · 2010 · cites it directly14descendants
  • K133700Advanix Pancreatic Stent and Naviflex Rapid Exchange Pancreatic Delivery System and…Boston Scientific Corporation · 2014 · 2 generations on4descendants
  • K193600EndoGI Biliary Stent SystemEndogi Medical, Ltd. · 2020 · 2 generations on3descendants
  • K140630Wallflex Biliary RX Stent SystemBoston Scientific · 2014 · 2 generations on3descendants
  • K202477EndoGI S-Path Biliary Stent SystemEndo GI Medical · 2020 · 3 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K834468Drainage StentMicrovasive19842
0 (this device)K965147Biliary Stent, Biliary Stent System, Biliary EndoprosthesisBoston Scientific Corp1997—
+1K101314Advanix Biliary Stent with Naviflex RX Delivery SystemBoston Scientific Corp2010—
+2K133700Advanix Pancreatic Stent and Naviflex Rapid Exchange Pancreatic Delivery…Boston Scientific Corporation201413
+2K193600EndoGI Biliary Stent SystemEndogi Medical, Ltd.2020—
+2K140630Wallflex Biliary RX Stent SystemBoston Scientific20141
+3K202477EndoGI S-Path Biliary Stent SystemEndo GI Medical2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199410

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201020
201420
201910
202030
202220
202320
202420
202520
202610
FIG. 1 — Lineage of K965147, 1984–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in FGE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FGE device

  1. #1 K240464 WallFlex Biliary RX Stent System; Epic Biliary…score 14.5 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K834468Drainage Stent
Summary page 3
…IMA™ BILIARY DRAINAGE CATHETER New Microvasive® Biliary i Solopass™ Percuflex i Fexima™ Biliary Stent and Delivery System Biliary Stent Drainage Catheter (This 510(k)) (K834468) (K944290) USE Indication Biliary Drainage «— Same «— Same Route of Administration Endoscopic «— Same ?:;fi::g;fi:; s STENT Type Am…
Microvasive19842
*K944290Catheters, Biliary Catheters, Urological
Summary page 3
…RY DRAINAGE CATHETER New Microvasive® Biliary i Solopass™ Percuflex i Fexima™ Biliary Stent and Delivery System Biliary Stent Drainage Catheter (This 510(k)) (K834468) (K944290) USE Indication Biliary Drainage «— Same «— Same Route of Administration Endoscopic «— Same ?:;fi::g;fi:; s STENT Type Amsterdam «—…
Boston Scientific Corp19945

Also cleared against K834468: 1 other device →

What descends from it? (17)

See all 17

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K101314Advanix Biliary Stent with Naviflex RX Delivery SystemBoston Scientific Corp201014
K101786Advanix Double Pigtail Biliary StentBoston Scientific Corp20101

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 26, 2018Class IIZ-1012-2019OtherThe sterile barrier may contain packaging seal defects.
Dec 26, 2018Class IIZ-1013-2019OtherThe sterile barrier may contain packaging seal defects.
Apr 27, 2010Not classifiedZ-1979-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1994-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1984-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1977-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1975-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1991-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1983-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1981-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1985-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1993-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1988-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-197-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1976-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1980-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1992-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1990-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1982-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1986-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1989-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 27, 2010Not classifiedZ-1987-2010PackagingFlexima Biliary Stent System Devices were packaged with Tyvek tray lids that had incorrect adhesive, resulting in sterile barrier seal defects. The breach of the sterility barrier could lead to the device being non-sterile.
Apr 2, 2008Not classifiedZ-2105-2008Packaging process controlRapid Exchange Biliary Stent may be mislabeled as Flexima biliary Stent, which has a different delivery system included in the package.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class.
  • Of its descendants, K101314 (Boston Scientific Corp, 2010) has gone furthest, with 14 descendants of its own.
  • 1 other device also cites its predicate K834468.

How does its category compare?

All FGE devices

FGE Stents, Drains And Dilators For The Biliary Ducts: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FGE

7in 2025

2007: 42008: 112009: 72010: 92011: 92012: 82013: 52014: 82015: 82016: 42017: 92018: 72019: 22020: 42021: 12022: 62023: 62024: 62025: 72026 (partial year): 8
2007high 11 (2008) · low 1 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FGE

257in 2025

2007: 316 days median2008: 214 days median2009: 160 days median2010: 81 days median2011: 114 days median2012: 96 days median2013: 161 days median2014: 156 days median2015: 79 days median2016: 49 days median2017: 59 days median2018: 215 days median2020: 121 days median2022: 101 days median2023: 199 days median2024: 188 days median2025: 257 days median2026 (partial year): 153 days median
2007high 316 (2007) · low 49 (2016)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FGE

1in 2024

2007: 542008: 932009: 182010: 322011: 22012: 42013: 32017: 12018: 432019: 132020: 12021: 272022: 212023: 192024: 12026 (partial year): 11
2007high 93 (2008) · low 0 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · FGE

1,923in 2025

2007: 1,0062008: 1,4652009: 1,6552010: 1,4342011: 1,3212012: 1,1832013: 1,0602014: 1,2702015: 1,5432016: 1,4532017: 1,5152018: 1,6812019: 1,8962020: 2,1822021: 2,2972022: 1,7492023: 1,6152024: 1,9482025: 1,923
2007high 2,297 (2021) · low 1,006 (2007)2025

MAUDE reports naming product code FGE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074316541,006
200811214931,465
20097160181,655
2010981321,434
2011911421,321
201289641,183
2013516131,060
2014815601,270
201587901,543
201644901,453
201795911,515
20187215431,681
20192—131,896
2020412112,182
20211—272,297
20226101211,749
20236199191,615
2024618811,948
2025725701,923
2026815311—

What is it indicated for?

The New Microvasive® Biliary Stent and Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K965147: Biliary Stent, Biliary Stent System, Biliary Endoprosthesis.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K965147

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026