Immulite 2000 Automated Immunoassay Analyzer

K970227510(k) TraditionalCleared Apr 8, 199781-day reviewClass I21 CFR 862.2160Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Dpc Cirrus · JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Cleared on Apr 8, 1997 after an 81-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 4 later devices descend from it. It has been recalled 13 times.

Decided
Apr 8, 1997
Review
81 days
Class
Class I
Product code
JJE
Recalls
13
Predicates
0

4 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2004: 1 descendant cleared2003: 1 descendant cleared2016: 1 descendant cleared2009: 1 descendant cleared1995200020052010201520202025
  1. This deviceK970227 Immulite 2000 Automated Immunoassay AnalyzerDpc Cirrus · 1997
  2. Descendant · cites it directly · 1 of its ownK041658 Bioplex 200 Ana Screen on the Bioplex 2200 Multi-Analyte Detection…Bio-Rad Laboratories, Inc. · 2004 · Recalled
  3. Descendant · cites it directly · 0 of its ownK160647 IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4Siemens Healthcare Diagnostics, Inc. · 2016
  4. Descendant · 2 generations on · 0 of its ownK083080 Igx Plex Rheumatoid Arthritis (RA) Assay and Sqidworks Diagnostics…Sqi Diagnostics Systems · 2009
  5. Descendant · cites it directly · 0 of its ownK033234 Immulite 2500 Automated Immunoassay AnalyzerDiagnostic Products Corp. · 2003 · Recalled

Most influential descendants

Showing 4 of 4 · See all
  • K041658Bioplex 200 Ana Screen on the Bioplex 2200 Multi-Analyte Detection System, Model Bioplex…Bio-Rad Laboratories, Inc. · 2004 · cites it directly1descendant
  • K160647IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4Siemens Healthcare Diagnostics, Inc. · 2016 · cites it directly0descendants
  • K083080Igx Plex Rheumatoid Arthritis (RA) Assay and Sqidworks Diagnostics PlatformSqi Diagnostics Systems · 2009 · 2 generations on0descendants
  • K033234Immulite 2500 Automated Immunoassay AnalyzerDiagnostic Products Corp. · 2003 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K970227Immulite 2000 Automated Immunoassay AnalyzerDpc Cirrus1997—
+1K041658Bioplex 200 Ana Screen on the Bioplex 2200 Multi-Analyte Detection System…Bio-Rad Laboratories, Inc.20041
+1K160647IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4Siemens Healthcare Diagnostics, Inc.2016—
+2K083080Igx Plex Rheumatoid Arthritis (RA) Assay and Sqidworks Diagnostics PlatformSqi Diagnostics Systems2009—
+1K033234Immulite 2500 Automated Immunoassay AnalyzerDiagnostic Products Corp.20033

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200410
200910
201610
FIG. 1 — Lineage of K970227, 1997–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JJE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041658Bioplex 200 Ana Screen on the Bioplex 2200 Multi-Analyte Detection System, Model Bioplex…Bio-Rad Laboratories, Inc.20041
K160647IMMULITE 2000 Third Generation TSH, IMMULITE 2000 Free T4Siemens Healthcare Diagnostics, Inc.20160
K033234Immulite 2500 Automated Immunoassay AnalyzerDiagnostic Products Corp.20030

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 15, 2020Class IIZ-2469-2020Nonconforming Material/ComponentFlexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
May 15, 2020Class IIZ-2470-2020Nonconforming Material/ComponentFlexible tubing connected to the liquid waste bottle can crack during routine customer maintenance potentially causing liquid waste can leak onto the floor, creating a slip and fall hazard
Mar 5, 2015Class IIZ-1658-2015Nonconforming Material/ComponentThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the…
Mar 5, 2015Class IIZ-1662-2015Nonconforming Material/ComponentThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the…
Mar 5, 2015Class IIZ-1663-2015Nonconforming Material/ComponentThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the…
Mar 5, 2015Class IIZ-1660-2015Nonconforming Material/ComponentThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the…
Mar 5, 2015Class IIZ-1661-2015Nonconforming Material/ComponentThe bottle assemblies were manufactured with a quality issue where in some cases the smaller diameter opening on both IMMULITE 2000 and IMMULITE 2000XPi Water Bottles and Liquid Waste Bottles is undersized and/or deformed, preventing the…
Apr 25, 2014Class IIIZ-1832-2014Device DesignThe keyboard and monitor may not be properly secured to the monitor arm.
Feb 3, 2014Class IIZ-1316-2014Device DesignThere is a potential for incorrect results to be reported from processing of a sample tube.
Oct 21, 2009Not classifiedZ-1033-2010Nonconforming Material/ComponentThe recall was initiated because Siemens Healthcare Diagnostics has identified there is the potential for patient results to be biased low (-0.3 Sample to Cutoff [S/CO] ratio units) due to incompatibility of the Toxoplasma IgM kits with…
May 15, 2009Not classifiedZ-1891-2010Device DesignThese substrate lots have shown a tendency to introduce a shift in both control and patient results with multiple methods.
Jan 5, 2009Not classifiedZ-1399-2009Process designDesign Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.
Apr 22, 2005Not classifiedZ-0994-05OtherThe Transport Chain is used to move a Reaction Tube within the Immulite 2000. It goes on to state the error condition that occurs due to a false trip of the home sensor.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041658 (Bio-Rad Laboratories, Inc., 2004) has gone furthest, with 1 descendant of its own.

How does its category compare?

All JJE devices

JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JJE

1in 2019

2008: 12011: 22012: 12013: 12014: 32019: 1
2007high 3 (2014) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JJE

224in 2014

2014: 224 days median
2007high 224 (2014)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JJE

6in 2025

2007: 202008: 202009: 142010: 322011: 252012: 162013: 212014: 142015: 152016: 92017: 182018: 172019: 72020: 72021: 62023: 122024: 62025: 62026 (partial year): 5
2007high 32 (2010) · low 0 (2022)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JJE

4,400in 2025

2007: 2,1812008: 2,1592009: 3,4402010: 3,2732011: 11,6062012: 3,7072013: 2,4892014: 2,8412015: 2,7932016: 2,5802017: 1,8692018: 4,5902019: 3,9482020: 3,9862021: 3,9202022: 3,9982023: 3,7832024: 4,2212025: 4,400
2007high 11,606 (2011) · low 1,869 (2017)2025

MAUDE reports naming product code JJE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—202,181
20081—202,159
20090—143,440
20100—323,273
20112—2511,606
20121—163,707
20131—212,489
20143224142,841
20150—152,793
20160—92,580
20170—181,869
20180—174,590
20191—73,948
20200—73,986
20210—63,920
20220—03,998
20230—123,783
20240—64,221
20250—64,400
20260—5—
Cite this page

510(k) Report. “K970227: Immulite 2000 Automated Immunoassay Analyzer.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K970227

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history