510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Aia Nexia-Automated Enzyme Immunoassay System

K971103510(k) TraditionalCleared Jun 27, 199793-day reviewClass I21 CFR 862.2560Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tosoh Medics, Inc. · KHO Fluorometer, For Clinical Use

Cleared on Jun 27, 1997 after a 93-day review. FDA’s file for it is a 510(k) statement, which names no predicate. It has been recalled 2 times.

Decided
Jun 27, 1997
Review
93 days
Class
Class I
Product code
KHO
Recalls
2
Predicates
0

FDA’s file for it is a 510(k) statement, which names no predicate.

FIG. 1 — Lineage of K971103Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KHO? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 15, 2013Class IIZ-1830-2013Nonconforming Material/ComponentVisual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.
Apr 2, 2003Not classifiedZ-0218-05OtherA software upgrade was released by the firm for its AIA-600 II analyzers, in that washing deficiencies caused by the analyzer were corrected.

No ancestor or descendant has a Class I recall.

How does its category compare?

All KHO devices

KHO Fluorometer, For Clinical Use: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KHO

1in 2012

2012: 1
2007high 1 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KHO

Too few KHO 510(k)s in any year since 2007 to give a median.

Recalls per year · KHO

7in 2018

2011: 22012: 12014: 22017: 42018: 7
2007high 7 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KHO

178in 2025

2007: 52008: 12009: 242010: 492011: 292012: 682013: 532014: 652015: 302016: 352017: 1522018: 2,1232019: 4402020: 2572021: 2132022: 2242023: 2602024: 1682025: 178
2007high 2,123 (2018) · low 1 (2008)2025

MAUDE reports naming product code KHO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—05
20080—01
20090—024
20100—049
20110—229
20121—168
20130—053
20140—265
20150—030
20160—035
20170—4152
20180—72,123
20190—0440
20200—0257
20210—0213
20220—0224
20230—0260
20240—0168
20250—0178
20260—0—

What is it indicated for?

The AIA NexIA Automated Enzyme Immunoassay System is a combination device, utilizing a fluorometer and automated pipettor, intended for clinical use. The NexIA is used in conjunction with the AIA-PACK reagent series to measure a variety of analytes in biological fluids which aid in the diagnosis and treatment of various disorders.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K971103: Aia Nexia-Automated Enzyme Immunoassay System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K971103

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)