510(k) Report

510(k) · Cleared

Recalled · 2009

Cross-Checks P

K972300510(k) TraditionalCleared Mar 25, 1998279-day reviewClass II21 CFR 880.2800Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Steritec Products Mfg Co, Inc. · JOJ Indicator, Physical/Chemical Sterilization Process

Cleared on Mar 25, 1998 as substantially equivalent to K921910 (Johnson & Johnson Medical, Inc., 1993) after a 279-day review. Its family tree reaches back 1 generation to 1993. It has been recalled once.

Decided
Mar 25, 1998
Review
279 days
Class
Class II
Product code
JOJ
Recalls
1
Predicates
1

Its longest line of descent runs back 1 generation to K921910 (1993). No later device cites it as a predicate yet.

K921910 · Sterrad Chemical Indicator · Johnson & Johnson Medical, Inc. · 1993199019921994199619982000
  1. Root · 1 generation backK921910 Sterrad Chemical IndicatorJohnson & Johnson Medical, Inc. · 1993
  2. This deviceK972300 Cross-Checks PSteritec Products Mfg Co, Inc. · 1998
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K921910Sterrad Chemical IndicatorJohnson & Johnson Medical, Inc.1993—
0 (this device)K972300Cross-Checks PSteritec Products Mfg Co, Inc.1998—
FIG. 1 — Lineage of K972300, 1993–1998Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JOJ? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JOJ device

  1. #1 K261992 3M™ Attest™ Steam Chemical Integrator (1243A…score 13.3 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K921910Sterrad Chemical Indicator
Summary page 2
…The Steritec Cross-Checks P Process Indicator Strips arc similas in design, and function to the: Johnson & Johnson Sterrad Chemical Indicator Sirips Ref 14100 510(k) K921910) Description of Steritec Cross-Checks P process indicator strips: \ j SteriTec Cross-Checks P strips are 4" x §/8” proces- indi…
Johnson & Johnson Medical, Inc.1993—

Also cleared against K921910: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 15, 2009Not classifiedZ-1532-2010OtherThis action notifies affected SteriTec customers as well as private label distributors of the necessity to modify the shelf life of their products. There is insufficient data on file to support the shelf life dating on the product labeling.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 4 other devices also cite its predicate K921910.

How does its category compare?

All JOJ devices

JOJ Indicator, Physical/Chemical Sterilization Process: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JOJ

6in 2025

2007: 82008: 92009: 72010: 62011: 132012: 72013: 62014: 62015: 82016: 42017: 72018: 52019: 122020: 72021: 42022: 82023: 42024: 122025: 62026 (partial year): 3
2007high 13 (2011) · low 4 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JOJ

90in 2025

2007: 140 days median2008: 57 days median2009: 188 days median2010: 115 days median2011: 151 days median2012: 224 days median2013: 212 days median2014: 198 days median2015: 188 days median2016: 150 days median2017: 172 days median2018: 150 days median2019: 38 days median2020: 108 days median2021: 87 days median2022: 90 days median2023: 53 days median2024: 71 days median2025: 90 days median2026 (partial year): 30 days median
2007high 224 (2012) · low 38 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JOJ

1in 2023

2008: 22009: 22011: 22021: 12023: 1
2007high 2 (2008) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JOJ

12in 2025

2007: 12008: 12009: 92010: 32011: 12012: 82013: 62014: 112015: 72016: 112017: 352018: 122019: 82020: 62021: 292022: 232023: 392024: 252025: 12
2007high 39 (2023) · low 1 (2007)2025

MAUDE reports naming product code JOJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007814001
200895721
2009718829
2010611503
20111315121
2012722408
2013621206
20146198011
2015818807
20164150011
20177172035
20185150012
2019123808
2020710806
2021487129
2022890023
2023453139
20241271025
2025690012
20263300—

What is it indicated for?

Cross-Checks P chemical process indicator strips are designed to be used in the Sterrad sterilizer. Cross-Checks P Process Indicator Strips will exhibit a blue to red color change when exposed to hydrogen peroxide during the diffusion phase of the Sterrad cycle. When used as directed, the Cross-Checks P strips give visible confirmation that the item has been exposed to hydrogen peroxide during the Sterrad cycle.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K972300: Cross-Checks P.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K972300

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026