510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Aia-Pack TSH 3RD-Gen Assay

K972586510(k) TraditionalCleared Aug 29, 199749-day reviewClass II21 CFR 862.1690Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Tosoh Medics, Inc. · JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Cleared on Aug 29, 1997 after a 49-day review. FDA’s file for it is a 510(k) statement, which names no predicate. 1 later device descends from it. It has been recalled once.

Decided
Aug 29, 1997
Review
49 days
Class
Class II
Product code
JLW
Recalls
1
Predicates
0

1 later device descends from it, 1 citing it directly.

Descendants, by year cleared 2014: 1 descendant cleared1995200020052010201520202025
  1. This deviceK972586 Aia-Pack TSH 3RD-Gen AssayTosoh Medics, Inc. · 1997
  2. Descendant · cites it directly · 0 of its ownK131928 Frend TSH (Reagent Cartridge)Nanoentek USA, Inc. · 2014

Most influential descendants

Showing 1 of 1 · See all
  • K131928Frend TSH (Reagent Cartridge)Nanoentek USA, Inc. · 2014 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K972586Aia-Pack TSH 3RD-Gen AssayTosoh Medics, Inc.1997—
+1K131928Frend TSH (Reagent Cartridge)Nanoentek USA, Inc.2014—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201410
FIG. 1 — Lineage of K972586, 1997–2014Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA 510(k) statement (indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JLW? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JLW device

  1. #1 K251543 Atellica® IM TSH3-Ultra II (TSH3ULII)score 11.4 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA’s file for it is a 510(k) statement, which names no predicate.

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K131928Frend TSH (Reagent Cartridge)Nanoentek USA, Inc.20140

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 5, 2018Class IIZ-2077-2018Device DesignAsfotase Alfa (Strensiq) interferes with certain Tosoh assays which potentially causes falsely increased or falsely decreased immunoassay test results, depending upon the assay.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K131928 (Nanoentek USA, Inc., 2014) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JLW devices

JLW Radioimmunoassay, Thyroid-Stimulating Hormone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JLW

1in 2025

2008: 52009: 22010: 12012: 12013: 12014: 12015: 32016: 32017: 42019: 12022: 12023: 12024: 22025: 12026 (partial year): 1
2007high 5 (2008) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JLW

194in 2017

2008: 143 days median2015: 142 days median2016: 238 days median2017: 194 days median
2007high 238 (2016) · low 142 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JLW

2in 2025

2007: 12010: 12011: 12012: 32016: 12018: 42019: 12020: 22024: 22025: 2
2007high 4 (2018) · low 0 (2008)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JLW

103in 2025

2007: 192008: 102009: 412010: 342011: 1672012: 1442013: 1682014: 1552015: 2032016: 1832017: 2352018: 2492019: 2162020: 1452021: 1462022: 1372023: 1622024: 1692025: 103
2007high 249 (2018) · low 10 (2008)2025

MAUDE reports naming product code JLW, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—119
20085143010
20092—041
20101—134
20110—1167
20121—3144
20131—0168
20141—0155
201531420203
201632381183
201741940235
20180—4249
20191—1216
20200—2145
20210—0146
20221—0137
20231—0162
20242—2169
20251—2103
20261—0—

What is it indicated for?

AIA-PACK TSH 3rd-Gen is intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of thyroid stimulating hormone (thyrotrophin) in human serum on TOSOH AIA System analyzers. Measurements obtained by this device are used in the diagnosis of thyroid or pituitary disorders.

Indications for use read by OCR from a scanned FDA 510(k) statement, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K972586: Aia-Pack TSH 3RD-Gen Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K972586

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026