Vital Signs Monitor; Advisor Model # Bci 9200

K982279510(k) TraditionalCleared Sep 25, 199887-day reviewClass II21 CFR 870.1025Recalls 40

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bci Intl., Inc. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Sep 25, 1998 as substantially equivalent to K970801 (Bci Intl., Inc., 1997) after an 87-day review. Its family tree reaches back 1 generation to 1997. 2 later devices descend from it. It has been recalled 40 times.

Decided
Sep 25, 1998
Review
87 days
Class
Class II
Product code
MHX
Recalls
40
Predicates
1

Its longest line of descent runs back 1 generation to K970801 (1997). 2 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2007: 1 descendant clearedReference device: K970209 · Capnocheck Plus · Bci Intl., Inc. · 1997 · Class I recallK970801 · Bci Model 6004 Nibp Monitor (6004) · Bci Intl., Inc. · 1997 · recalled1995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 1 descendant K970801 · Bci Model 6004 Nibp Monitor (6004) · Bci Intl., Inc. · 1997 · recalled0
  1. Root · 1 generation backK970801 Bci Model 6004 Nibp Monitor (6004)Bci Intl., Inc. · 1997 · Recalled
  2. This deviceK982279 Vital Signs Monitor; Advisor Model # Bci 9200Bci Intl., Inc. · 1998
  3. Descendant · cites it directly · 1 of its ownK030407 Advisor, Model 9200Bci, Inc. · 2003 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK061931 Solaris Compatible Reusable Adult SPO2 Finger Sensors, Models…Solaris Medical Technology, Inc. · 2007

Most influential descendants

Showing 2 of 2 · See all
  • K030407Advisor, Model 9200Bci, Inc. · 2003 · cites it directly1descendant
  • K061931Solaris Compatible Reusable Adult SPO2 Finger Sensors, Models S100A-090103 and 300103…Solaris Medical Technology, Inc. · 2007 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K970801Bci Model 6004 Nibp Monitor (6004)Bci Intl., Inc.19971
0 (this device)K982279Vital Signs Monitor; Advisor Model # Bci 9200Bci Intl., Inc.1998—
+1K030407Advisor, Model 9200Bci, Inc.200343
+2K061931Solaris Compatible Reusable Adult SPO2 Finger Sensors, Models S100A-090103 and…Solaris Medical Technology, Inc.2007—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200710
FIG. 1 — Lineage of K982279, 1997–2007Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K970801Bci Model 6004 Nibp Monitor (6004)
Summary page 3
…This test was to determine the accuracy of the NIBP technology using the MicroNIBP module. The next test showed that the MicroNIBP module and the BCI 6004 NIBP monitor (K970801) performed the same. This is to support the accuracy of the MAP readings. The SP-10 testing was conducted under an approved IRB…
Bci Intl., Inc.19971
*—BCI 6200 Vital Signs Monitor
Named without a submission number
—
*—BCI 9100 Multigas Monitor
Named without a submission number
—
*—Siemens SC6000 Bedside Monitor
Named without a submission number
—
†K970209Capnocheck PlusBci Intl., Inc.1997Class I

Also cleared against K970801: 2 other devices →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K030407Advisor, Model 9200Bci, Inc.20031

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Dec 22, 2008Not classifiedZ-1326-2009Device DesignRecent improvements were made to the Advisor case plastic resulting in the inserts being deeply embedded into the plastic. This affects the ability of the monitor to be adequately secured on the wall mount bracket or roll stand bracket…
Jun 5, 2007Not classifiedZ-1089-2007OtherTurn off unexpectedly/ Not turn on : monitors turned off unexpectedly sounding the critical failure alarm. Advisor Vital Signs monitors that were off, would not turn back on.
Sep 1, 2005Not classifiedZ-0227-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0228-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0237-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0239-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0221-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0224-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0226-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0229-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0232-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0234-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0236-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0238-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0240-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0222-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0220-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0223-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0225-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0230-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0231-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0233-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Sep 1, 2005Not classifiedZ-0235-06OtherPotential Safety related problem associated with the pace detect function on some Advisor monitors that have impedance respiration capabilities in that random pace markers may appear when the pace detect function is turned on and the…
Mar 16, 2005Not classifiedZ-0825-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0826-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0827-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0828-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0831-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0834-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0824-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0829-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0830-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0833-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0832-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Mar 16, 2005Not classifiedZ-0835-05OtherAltitude testing of the Advisor with Capnography module found the CO2 (End Tidal and Inspired) values are incorrect when the Advisor is configured to display in the relative units of percent (%) at elevations above 4,100 feet (at standard…
Jan 5, 2005Not classifiedZ-0531-05OtherAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Jan 5, 2005Not classifiedZ-0532-05OtherAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Jan 5, 2005Not classifiedZ-0533-05OtherAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Jan 5, 2005Not classifiedZ-0534-05OtherAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.
Jan 5, 2005Not classifiedZ-0535-05OtherAn issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K030407 (Bci, Inc., 2003) has gone furthest, with 1 descendant of its own.
  • 2 other devices also cite its predicate K970801.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The 9200 vital signs monitor is intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of the hospital or clinic where low end monitoring systems are needed. The basic monitor package includes ECG (3 lead / 5 lead), impedance respiration (RSP), non-invasive blood pressure (NIBP), pulse oximetry (SpO,), two invasive blood pressures (P1 and P2), and two temperature channels (T1 and T2). A 2-inch…

Read the full statement (162 words)

The 9200 vital signs monitor is intended to be used in the ICU, CCU, OR, ER, RR, Labor and Delivery rooms, special procedure labs and other areas of the hospital or clinic where low end monitoring systems are needed. The basic monitor package includes ECG (3 lead / 5 lead), impedance respiration (RSP), non-invasive blood pressure (NIBP), pulse oximetry (SpO,), two invasive blood pressures (P1 and P2), and two temperature channels (T1 and T2). A 2-inch internal, graphical/alphanumeric printer and a battery are provided as options. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. The device will provide fast, reliable measurements on patients ranging from children (pediatric) to adults when using the appropriate BCI accessories.

The monitor is not intended for home use. The monitor is not intended to be an apnea monitor. It was not designed or validated for use as an apnea monitor. The monitor is not intended for neonatal use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982279: Vital Signs Monitor; Advisor Model # Bci 9200.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K982279

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026