510(k) · Cleared
ClearedBelmont Portable Balloon Pump, Bard Transact Iabp
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Belmont Instrument Corp. · DSP System, Balloon, Intra-Aortic And Control
Cleared on Oct 2, 1998 as substantially equivalent to K915580 (Belmont Instrument Corp., 1992) after a 73-day review. Its family tree reaches back 1 generation to 1992.
- Decided
- Oct 2, 1998
- Review
- 73 days
- Class
- Class II
- Product code
- DSP
- Recalls
- None
- Predicates
- 1
Its longest line of descent runs back 1 generation to K915580 (1992). No later device cites it as a predicate yet.
- Root · 1 generation backK915580 Belmont Model PBP Intra-Aortic Balloon PumpBelmont Instrument Corp. · 1992
- This deviceK982538 Belmont Portable Balloon Pump, Bard Transact IabpBelmont Instrument Corp. · 1998
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K915580 | Belmont Model PBP Intra-Aortic Balloon Pump | Belmont Instrument Corp. | 1992 | — |
| 0 (this device) | K982538 | Belmont Portable Balloon Pump, Bard Transact Iabp | Belmont Instrument Corp. | 1998 | — |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DSP? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DSP device
- #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K915580 | Belmont Model PBP Intra-Aortic Balloon PumpSummary page 7…oftware to be used in the IABP Indications For Use: There are no changes to the intended use of the IABP for which a 510(k) was approved in October 1992 [510(k) number K915580]. The balloon pump is an electromechanical system used to drive intra-aortic balloons. It provides temporary support to the left… | 1992 |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Its one ancestor has no recall.
- 1 other device also cites its predicate K915580.
How does its category compare?
All DSP devicesDSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DSP
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DSP
140in 2019
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DSP
4in 2025
Class I (47 since 2007) Class II/III not classified
Adverse event reports per year · DSP
7,422in 2025
MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 3 | 461 |
| 2008 | 0 | — | 0 | 617 |
| 2009 | 2 | — | 14 | 656 |
| 2010 | 0 | — | 8 | 1,019 |
| 2011 | 2 | — | 0 | 590 |
| 2012 | 2 | — | 3 | 490 |
| 2013 | 1 | — | 0 | 495 |
| 2014 | 0 | — | 1 | 1,738 |
| 2015 | 1 | — | 0 | 404 |
| 2016 | 0 | — | 9 (9) | 419 |
| 2017 | 3 | 73 | 0 | 2,164 |
| 2018 | 1 | — | 2 (2) | 3,331 |
| 2019 | 3 | 140 | 3 | 3,552 |
| 2020 | 2 | — | 0 | 3,393 |
| 2021 | 0 | — | 5 (3) | 4,345 |
| 2022 | 0 | — | 33 (33) | 5,113 |
| 2023 | 1 | — | 0 | 6,980 |
| 2024 | 0 | — | 1 | 7,317 |
| 2025 | 1 | — | 4 | 7,422 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
There are no changes to the intended use of the IABP for which a 510(k) was approved in October 1992 [510(k) number K915580]. The balloon pump is an electromechanical system used to drive intra-aortic balloons. It provides temporary support to the left ventricle via the principle of counterpulsation. The intra-aortic balloon is placed in the descending aorta, just distal to the left subclavian artery. Once the balloon is positioned, the pump is adjusted to trigger in synchrony with the ECG or arterial pressure waveform to ensure that inflation and deflation occur at the appropriate points during the cardiac cycle.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K982538 · printed Nov 6, 2026
Cite this page
510(k) Report. “K982538: Belmont Portable Balloon Pump, Bard Transact Iabp.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K982538
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026