510(k) Report

510(k) · Cleared

Cleared

Belmont Model PBP Intra-Aortic Balloon Pump

K915580510(k) TraditionalCleared Oct 30, 1992339-day reviewClass II21 CFR 870.3535

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Belmont Instrument Corp. · DSP System, Balloon, Intra-Aortic And Control

Cleared on Oct 30, 1992 after a 339-day review. FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known. 8 of the 20 devices in its family, counting itself, have a Class I recall. 19 later devices descend from it.

Decided
Oct 30, 1992
Review
339 days
Class
Class II
Product code
DSP
Recalls
None
Predicates
0

19 later devices descend from it, 2 citing it directly. 8 of the 20 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2001: 1 descendant cleared1998: 1 descendant cleared2002: 1 descendant cleared2006: 1 descendant cleared, 1 with a Class I recall12007: 2 descendants cleared, 1 with a Class I recall12020: 1 descendant cleared, 1 with a Class I recall12017: 3 descendants cleared, 2 with a Class I recall22011: 2 descendants cleared, 1 with a Class I recall12019: 1 descendant cleared2023: 1 descendant cleared2012: 1 descendant cleared2013: 1 descendant cleared2018: 1 descendant cleared, 1 with a Class I recall12015: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared1990200020102020
  1. This deviceK915580 Belmont Model PBP Intra-Aortic Balloon PumpBelmont Instrument Corp. · 1992
  2. Descendant · cites it directly · 17 of its ownK002256 Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc. · 2001
  3. Descendant · 2 generations on · 14 of its ownK060309 Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · Recalled (Class I)
  4. Descendant · 3 generations on · 8 of its ownK063525 Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter…Datascope Corp. · 2007 · Recalled (Class I)
  5. Descendant · 3 generations on · 4 of its ownK162820 AC3 Series IABP SystemArrow International, Inc. · 2017 · Recalled (Class I)
  6. Descendant · 4 generations on · 3 of its ownK192238 AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex · 2019

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

1998: 12001: 12002: 12006: 12007: 22011: 22012: 12013: 12015: 12017: 32018: 12019: 12020: 12023: 12025: 1
1998high 3 (2017) · low 0 (1999)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 19 · See all
  • K002256Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc. · 2001 · cites it directly17descendants
  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · 2 generations on14descendants
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp. · 2007 · 3 generations on8descendants
  • K162820AC3 Series IABP SystemArrow International, Inc. · 2017 · 3 generations on4descendants
  • K192238AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex · 2019 · 4 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K915580Belmont Model PBP Intra-Aortic Balloon PumpBelmont Instrument Corp.1992—
+1K002256Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc.2001—
+2K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc.2006Class I
+3K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300…Datascope Corp.2007Class I
+3K162820AC3 Series IABP SystemArrow International, Inc.2017Class I
+4K192238AC3 Series Intra-Aortic Balloon Pump (IABP)Arrow International, Teleflex2019—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199810
200110
200210
200611
200721
201121
201210
201310
201511
201732
201811
201910
202011
202310
202510
FIG. 1 — Lineage of K915580, 1992–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

8 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DSP device

  1. #1 K250542 AC3™ Range™ Intra-Aortic Balloon Pumpscore 11.2 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA lists a 510(k) statement for it but has not posted the file, so its predicates are not known.

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K002256Arrow Acat 2 Intra-Aortic Balloon PumpArrow Intl., Inc.200117
K982538Belmont Portable Balloon Pump, Bard Transact IabpBelmont Instrument Corp.19980

Has anything in its family been recalled? (8 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K060309Autocat Intra-Aortic Balloon Pump SeriesArrow Intl., Inc. · 2006 · 2 generations on · ◆ in the 2006 descendants bar of FIG. 128 Class I recalls
  • K201112AC3 IABP, AC3 Optimus IABP, AutoCAT2 Wave IABP, AutoCAT2 IABPArrow International, Inc. · 2020 · 3 generations on · ◆ in the 2020 descendants bar of FIG. 111 Class I recalls
  • K063525Datascope Fiber Optic Pressure Sensor Intra-Aortic Balloon Catheter and CS300 Intra-Aortic Balloon Pump…Datascope Corp. · 2007 · 3 generations on · ◆ in the 2007 descendants bar of FIG. 11 Class I recall
  • K162820AC3 Series IABP SystemArrow International, Inc. · 2017 · 3 generations on · ◆ in the 2017 descendants bar of FIG. 12 Class I recalls
  • K112372Cardiosave Hybrid Intra-Aortic Balloon PumpCardiac Assist, Maquet Cardiovascular, LLC · 2011 · 4 generations on · ◆ in the 2011 descendants bar of FIG. 110 Class I recalls
  • K181122CARDIOSAVE Intra-Aortic Balloon PumpDatascope Corp. · 2018 · 5 generations on · ◆ in the 2018 descendants bar of FIG. 127 Class I recalls
  • K151254CARDIOSAVE Hybrid Intra-Aortic Balloon PumpDatascope Corp. · 2015 · 5 generations on · ◆ in the 2015 descendants bar of FIG. 112 Class I recalls
  • K163542CARDIOSAVE Hybrid and Rescue Intra-Aortic Balloon Pump, CARDIOSAVE Li-Ion Battery Transport and Storage CaseDatascope Corp. · 2017 · 6 generations on · ◆ in the 2017 descendants bar of FIG. 17 Class I recalls

What does this lineage say?

  • Of its descendants, K002256 (Arrow Intl., Inc., 2001) has gone furthest, with 17 descendants of its own.

How does its category compare?

All DSP devices

DSP System, Balloon, Intra-Aortic And Control: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DSP

1in 2025

2007: 22009: 22011: 22012: 22013: 12015: 12017: 32018: 12019: 32020: 22023: 12025: 1
2007high 3 (2017) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DSP

140in 2019

2017: 73 days median2019: 140 days median
2007high 140 (2019) · low 73 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DSP

4in 2025

2007: 32009: 142010: 82012: 32014: 12016: 9, 9 Class I2018: 2, 2 Class I2019: 32021: 5, 3 Class I2022: 33, 33 Class I2024: 12025: 4
2007high 33 (2022) · low 0 (2008)2026

Class I (47 since 2007) Class II/III not classified

Adverse event reports per year · DSP

7,422in 2025

2007: 4612008: 6172009: 6562010: 1,0192011: 5902012: 4902013: 4952014: 1,7382015: 4042016: 4192017: 2,1642018: 3,3312019: 3,5522020: 3,3932021: 4,3452022: 5,1132023: 6,9802024: 7,3172025: 7,422
2007high 7,422 (2025) · low 404 (2015)2025

MAUDE reports naming product code DSP, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—3461
20080—0617
20092—14656
20100—81,019
20112—0590
20122—3490
20131—0495
20140—11,738
20151—0404
20160—9 (9)419
201737302,164
20181—2 (2)3,331
2019314033,552
20202—03,393
20210—5 (3)4,345
20220—33 (33)5,113
20231—06,980
20240—17,317
20251—47,422
20260—0—
Cite this page

510(k) Report. “K915580: Belmont Model PBP Intra-Aortic Balloon Pump.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 3, 2026. https://510kreport.com/device/K915580

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026