510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

Philips Omnidiagnost

K982993510(k) TraditionalCleared Sep 29, 199833-day reviewClass II21 CFR 892.1650Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Sep 29, 1998 as substantially equivalent to K912470 (Philips Medical Systems, Inc., 1991) after a 33-day review. Its family tree reaches back 1 generation to 1991. 3 later devices descend from it. It has been recalled 4 times.

Decided
Sep 29, 1998
Review
33 days
Class
Class II
Product code
OWB
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K912470 (1991). 3 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2003: 1 descendant cleared2010: 1 descendant cleared2005: 1 descendant clearedK912470 · Philips Diagnost 96 · Philips Medical Systems, Inc. · 1991 · recalled1990200020102020
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants K912470 · Philips Diagnost 96 · Philips Medical Systems, Inc. · 1991 · recalled0
  1. Root · 1 generation backK912470 Philips Diagnost 96Philips Medical Systems, Inc. · 1991 · Recalled
  2. This deviceK982993 Philips OmnidiagnostPhilips Medical Systems, Inc. · 1998
  3. Descendant · cites it directly · 2 of its ownK032046 Phillips Omnidiagnost ElevaPhilips Medical Systems, Inc. · 2003 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK092713 Spectre Wireless Encrypted Footswitch and Hand Switch SystemTransamerican Medical Imaging · 2010 · Recalled
  5. Descendant · 2 generations on · 0 of its ownK050692 FLXISPhilips Medical Systems North America Co. · 2005

Most influential descendants

Showing 3 of 3 · See all
  • K032046Phillips Omnidiagnost ElevaPhilips Medical Systems, Inc. · 2003 · cites it directly2descendants
  • K092713Spectre Wireless Encrypted Footswitch and Hand Switch SystemTransamerican Medical Imaging · 2010 · 2 generations on0descendants
  • K050692FLXISPhilips Medical Systems North America Co. · 2005 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K912470Philips Diagnost 96Philips Medical Systems, Inc.19919
0 (this device)K982993Philips OmnidiagnostPhilips Medical Systems, Inc.1998—
+1K032046Phillips Omnidiagnost ElevaPhilips Medical Systems, Inc.20038
+2K092713Spectre Wireless Encrypted Footswitch and Hand Switch SystemTransamerican Medical Imaging20101
+2K050692FLXISPhilips Medical Systems North America Co.2005—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
201010
FIG. 1 — Lineage of K982993, 1991–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in OWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K912470Philips Diagnost 96
Summary page 1
…uced under K920950). Substantial equivalence Information The OmniDiagnost is a modification of, and substantially equivalent to the Philips Diagnost 96 system (FDA ref. K912470). Safety Information The Philips OmniDiagnost complies with the applicable portions of 21 CFR parts 1020.30/.31/.32 and voluntar…
Philips Medical Systems, Inc.19919

Also cleared against K912470: 2 other devices →

What descends from it? (3)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032046Phillips Omnidiagnost ElevaPhilips Medical Systems, Inc.20032

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 28, 2019Class IIZ-1054-2019Component design/selectionThe locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.
May 1, 2018Class IIZ-2301-2018Device DesignThe fixation of the upper and lower tilt actuator, of the Omnidiagnost system might break off and the table will start to rotate from 0 to + 90 /- 90 degree with high speed. This rotating movement can not be stopped by the user. This…
Jul 6, 2009Not classifiedZ-1665-2009Radiation Control for Health and Safety ActThe x-ray system fail to meet the requirements of 21 CFR 1020.32(d)(2)(ii) when a specific tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch
Jul 17, 2003Not classifiedZ-0874-03OtherLabeling of push button caps located on the control panel may not clearly identify the applied Source to Image distance (SID).

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K032046 (Philips Medical Systems, Inc., 2003) has gone furthest, with 2 descendants of its own.
  • 2 other devices also cite its predicate K912470.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The Philips OmniDiagnost is a diagnostic imaging device intended for radiographic,

fluoroscopic, angiographic, and interventional applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K982993: Philips Omnidiagnost.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K982993

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026