510(k) Report

510(k) · Cleared

Recalled · 2011

Leksell Gamma Knife Target System, Model 24001

K984328510(k) TraditionalCleared May 21, 1999169-day reviewClass II21 CFR 892.5750Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Elekta Instrument AB · IWB System, Radiation Therapy, Radionuclide

Cleared on May 21, 1999 as substantially equivalent to K924849 (Elekta Radiosurgery, Inc., 1995) after a 169-day review. Its family tree reaches back 1 generation to 1995. 15 later devices descend from it. It has been recalled 4 times.

Decided
May 21, 1999
Review
169 days
Class
Class II
Product code
IWB
Recalls
4
Predicates
1

Its longest line of descent runs back 1 generation to K924849 (1995). 15 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2006: 1 descendant cleared2011: 2 descendants cleared2007: 1 descendant cleared2009: 1 descendant cleared2012: 1 descendant cleared2014: 3 descendants cleared2015: 2 descendants cleared2016: 1 descendant cleared2018: 1 descendant cleared2022: 1 descendant cleared2025: 1 descendant clearedK924849 · Leksell Gamma Unit Model 23004,Type B. · Elekta Radiosurgery, Inc. · 19951995200020052010201520202025
Descendants, by generation 1 generation forward: 1 descendant2 generations forward: 2 descendants3 generations forward: 3 descendants4 generations forward: 2 descendants5 generations forward: 3 descendants6 generations forward: 2 descendants7 generations forward: 1 descendant8 generations forward: 1 descendant K924849 · Leksell Gamma Unit Model 23004,Type B. · Elekta Radiosurgery, Inc. · 19950+5
  1. Root · 1 generation backK924849 Leksell Gamma Unit Model 23004,Type B.Elekta Radiosurgery, Inc. · 1995
  2. This deviceK984328 Leksell Gamma Knife Target System, Model 24001Elekta Instrument AB · 1999
  3. Descendant · cites it directly · 14 of its ownK061941 Leksell Gamma Knife Perfexion, Model 715000Elekta Instrument AB · 2006
  4. Descendant · 2 generations on · 11 of its ownK063512 Leksell Gamma Knife PerfexionElekta Instrument AB · 2007 · Recalled
  5. Descendant · 3 generations on · 7 of its ownK120811 Leksell Gamma Knife PerfexionElekta Instrument AB · 2012 · Recalled
  6. Descendant · 4 generations on · 6 of its ownK133565 Leksell Gamma Knife PerfexionElekta Instrument AB · 2014
  7. Descendant · 5 generations on · 4 of its ownK151159 Leksell Gamma Knife PerfexionElekta Instrument AB · 2015 · Recalled

Showing 5 of 15 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12009: 12011: 22012: 12014: 32015: 22016: 12018: 12022: 12025: 1
2006high 3 (2014) · low 0 (2008)2026

Most influential descendants

Showing 5 of 15 · See all
  • K061941Leksell Gamma Knife Perfexion, Model 715000Elekta Instrument AB · 2006 · cites it directly14descendants
  • K063512Leksell Gamma Knife PerfexionElekta Instrument AB · 2007 · 2 generations on11descendants
  • K120811Leksell Gamma Knife PerfexionElekta Instrument AB · 2012 · 3 generations on7descendants
  • K133565Leksell Gamma Knife PerfexionElekta Instrument AB · 2014 · 4 generations on6descendants
  • K151159Leksell Gamma Knife PerfexionElekta Instrument AB · 2015 · 5 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K924849Leksell Gamma Unit Model 23004,Type B.Elekta Radiosurgery, Inc.1995—
0 (this device)K984328Leksell Gamma Knife Target System, Model 24001Elekta Instrument AB1999—
+1K061941Leksell Gamma Knife Perfexion, Model 715000Elekta Instrument AB2006—
+2K063512Leksell Gamma Knife PerfexionElekta Instrument AB20074
+3K120811Leksell Gamma Knife PerfexionElekta Instrument AB20121
+4K133565Leksell Gamma Knife PerfexionElekta Instrument AB2014—
+5K151159Leksell Gamma Knife PerfexionElekta Instrument AB20151

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
200910
201120
201210
201430
201520
201610
201810
202210
202510
FIG. 1 — Lineage of K984328, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in IWB? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new IWB device

  1. #1 K250391 Leksell Gamma Knife® (Elekta Esprit); Leksell…score 11.2 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K924849Leksell Gamma Unit Model 23004,Type B.
Summary page 1
…ionuclide Radiation Therapy Device 21 CFR 892.5750 19.3 IDENTIFICATION OF PREDICATE DEVICE Company Device 510(k) No. Date Cleared Elekta Instruments Leksell Gamma Unit K924849 November 20, 1995 Model 23004, Type B Leksell® Gamma Knife Target System Model 24001 Original Premarket 510(k) No…
Elekta Radiosurgery, Inc.1995—

What descends from it? (15)

See all 15

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061941Leksell Gamma Knife Perfexion, Model 715000Elekta Instrument AB200614

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 26, 2011Not classifiedZ-0312-2012PendingSeveral of the LMR03 actuators with bronze drive nut have failed unexpectedly, creating a potential safety hazard for the operator and patient.
Oct 1, 2007Not classifiedZ-1236-2010OtherAfter updating the LGK actuator in spare part 810361, the old sleigh became obsolete due to causing insufficient locking of the helmet in combination with the new actuator.
Sep 1, 2006Not classifiedZ-1533-2010OtherLeksell Gamma Knife C 1.2 with MCU version 4.0.0.6 improperly adjusted trolley for 4mm helmet caused jamming when helmet changer attempted to lock helmet. Lock did not fully engage and could still be raised. When the lock does not engage…
Aug 1, 2006Not classifiedZ-1429-2010OtherDue to a material change in a critical component of the linear actuators the quality of said part was found to be inadequate.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K061941 (Elekta Instrument AB, 2006) has gone furthest, with 14 descendants of its own.

How does its category compare?

All IWB devices

IWB System, Radiation Therapy, Radionuclide: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · IWB

1in 2025

2007: 12009: 12011: 12012: 12014: 32015: 32016: 12017: 12018: 32019: 12020: 12021: 22022: 12024: 12025: 1
2007high 3 (2014) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · IWB

57in 2018

2014: 170 days median2015: 79 days median2018: 57 days median
2007high 170 (2014) · low 57 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · IWB

1in 2023

2007: 32008: 32009: 12010: 12011: 22014: 12023: 1
2007high 3 (2007) · low 0 (2012)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · IWB

3in 2025

2007: 82008: 152009: 82010: 92011: 52012: 32013: 22014: 82015: 52016: 72017: 42018: 22019: 32020: 22023: 12025: 3
2007high 15 (2008) · low 0 (2021)2025

MAUDE reports naming product code IWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—38
20080—315
20091—18
20100—19
20111—25
20121—03
20130—02
2014317018
201537905
20161—07
20171—04
201835702
20191—03
20201—02
20212—00
20221—00
20230—11
20241—00
20251—03
20260—0—

What is it indicated for?

The Leksell® Gamma Knife Target System Model 24001 is a teletherapy device indicated for use in the stereotactic irradiation of intracranial structures

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K984328: Leksell Gamma Knife Target System, Model 24001.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K984328

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026