510(k) · Cleared
ClearedHP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K with Easi-Ecg Option
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Hewlett-Packard GmbH · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)
Cleared on Jul 22, 1999 as substantially equivalent to K872781 (Totemite Medical Electronics, 1987) and 1 other predicate after a 156-day review. Its family tree reaches back 1 generation to 1987. 437 later devices descend from it.
- Decided
- Jul 22, 1999
- Review
- 156 days
- Class
- Class II
- Product code
- MHX
- Recalls
- None
- Predicates
- 2
Its longest line of descent runs back 1 generation to K872781 (1987). 2 ancestors in all. 437 later devices descend from it, 3 citing it directly.
- Root · 1 generation backK872781 Easi Lead System Cable AssemblyTotemite Medical Electronics · 1987
- This deviceK990476 HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model…Hewlett-Packard GmbH · 1999
- Descendant · cites it directly · 436 of its ownK992595 HP Viridia Component Monitoring System, Model Rev. K with Easi-ST…Hewlett-Packard GmbH · 1999
- Descendant · cites it directly · 431 of its ownK013199 Philips Component Compact Monitor,Model M1275B, Philips Compenent…Philips Medical Systems, Inc. · 2001
- Descendant · 2 generations on · 421 of its ownK021300 Modification of Philips Component Compact Monitor, Release a.03…Philips Medical Systems · 2002
- Descendant · 2 generations on · 416 of its ownK020531 Philips Multi-Parameter Module, Model M3001A; Philips Component…Philips Medical Systems · 2002
- Descendant · 2 generations on · 407 of its ownK021778 Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems · 2002
Showing 5 of 437 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 437 · See all- K992595HP Viridia Component Monitoring System, Model Rev. K with Easi-ST Segment Measurement…Hewlett-Packard GmbH · 1999 · cites it directly436descendants
- K013199Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model M1276BPhilips Medical Systems, Inc. · 2001 · cites it directly431descendants
- K021300Modification of Philips Component Compact Monitor, Release a.03, Model M1275BPhilips Medical Systems · 2002 · 2 generations on421descendants
- K020531Philips Multi-Parameter Module, Model M3001A; Philips Component Compact Monitor, Release…Philips Medical Systems · 2002 · 2 generations on416descendants
- K021778Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems · 2002 · 2 generations on407descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K872781 | Easi Lead System Cable Assembly | Totemite Medical Electronics | 1987 | — |
| 0 (this device) | K990476 | HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K… | Hewlett-Packard GmbH | 1999 | — |
| +1 | K992595 | HP Viridia Component Monitoring System, Model Rev. K with Easi-ST Segment… | Hewlett-Packard GmbH | 1999 | — |
| +1 | K013199 | Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model… | Philips Medical Systems, Inc. | 2001 | — |
| +2 | K021300 | Modification of Philips Component Compact Monitor, Release a.03, Model M1275B | Philips Medical Systems | 2002 | — |
| +2 | K020531 | Philips Multi-Parameter Module, Model M3001A; Philips Component Compact… | Philips Medical Systems | 2002 | — |
| +2 | K021778 | Philips Component Compact Monitor, Release a.04, Model #M1275B | Philips Medical Systems | 2002 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1995 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2001 | 1 | 0 |
| 2002 | 4 | 0 |
| 2003 | 8 | 0 |
| 2004 | 10 | 0 |
| 2005 | 10 | 0 |
| 2006 | 17 | 0 |
| 2007 | 7 | 0 |
| 2008 | 14 | 0 |
| 2009 | 19 | 0 |
| 2010 | 21 | 0 |
| 2011 | 21 | 0 |
| 2012 | 19 | 0 |
| 2013 | 17 | 0 |
| 2014 | 16 | 0 |
| 2015 | 14 | 0 |
| 2016 | 17 | 0 |
| 2017 | 17 | 0 |
| 2018 | 25 | 0 |
| 2019 | 17 | 0 |
| 2020 | 14 | 0 |
| 2021 | 19 | 0 |
| 2022 | 28 | 0 |
| 2023 | 27 | 0 |
| 2024 | 25 | 0 |
| 2025 | 29 | 0 |
| 2026 | 20 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new MHX device
- #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K872781 | Easi Lead System Cable AssemblySummary page 1…ctor and Alarm, Arrythmia 3. The new device with the EASI-ECG Option is substantially equivalent to the following predicate devices: the Totemite EASI Lead System Cable K872781B (12/29/87), and the Zymed T8010 Telemetry Central Station Monitor K951370 (10/6/95). The accessories and materials are optional… | 1987 |
| * | K951370 | Zymed Telemetry System; Model T8010Summary page 1…bstantially equivalent to the following predicate devices: the Totemite EASI Lead System Cable K872781B (12/29/87), and the Zymed T8010 Telemetry Central Station Monitor K951370 (10/6/95). The accessories and materials are optional and are the same accessories originally cleared for use with HP M1175A/76A… | 1995 |
| † | K882609 | Models M1175A & M1176A Component Monitoring System | 1989 |
| † | K973437 | Hewlett-Packard Models M1001B Ecg M1002B Ecg/Respiration Plug-in Modules | 1997 |
What descends from it? (437)
See all 4373 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
No FDA recall names this device.
No ancestor or descendant has a Class I recall.
What does this lineage say?
How does its category compare?
All MHX devicesMHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · MHX
8in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · MHX
192in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · MHX
21so far in 2026, the most since 2007
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · MHX
3,761in 2025
MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 15 | 96 | 3 | 131 |
| 2008 | 16 | 58 | 2 | 232 |
| 2009 | 13 | 57 | 7 | 382 |
| 2010 | 20 | 90 | 13 | 955 |
| 2011 | 12 | 81 | 16 | 949 |
| 2012 | 16 | 88 | 11 | 676 |
| 2013 | 21 | 120 | 14 | 1,997 |
| 2014 | 13 | 248 | 6 | 1,609 |
| 2015 | 14 | 203 | 5 | 1,284 |
| 2016 | 13 | 149 | 7 | 1,105 |
| 2017 | 14 | 134 | 6 | 1,115 |
| 2018 | 11 | 203 | 12 | 2,157 |
| 2019 | 7 | 243 | 6 | 1,462 |
| 2020 | 11 | 196 | 0 | 1,714 |
| 2021 | 5 | 195 | 0 | 2,423 |
| 2022 | 15 | 189 | 6 | 1,470 |
| 2023 | 8 | 229 | 3 | 2,254 |
| 2024 | 5 | 181 | 3 | 2,814 |
| 2025 | 8 | 192 | 15 | 3,761 |
| 2026 | 14 | 205 | 21 | — |
What is it indicated for?
The Hewlett-Packard Viridia Component Monitoring System, (M1175A/76A/77RA), and Viridia 24/26 (M1205A) Rev.K with EASI-ECG Option is indicated for:
Assessment of symptoms that may be related to rhythm disturbances of the heart: patients with palpitations, the evaluation of arrhythmia’s in adult and pediatric patients.
Assessment of risk in patients with or without symptoms of arrhythmia.
Assessment of efficacy of antiarrhythmic therapy.
Assessment of pacemaker function.
Assessment of symptomatic or asymptomatic patients to evaluate for ischemic heart disease.
Assessment is indicated for single-hospital environment.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K990476 · printed Nov 4, 2026
Cite this page
510(k) Report. “K990476: HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K with Easi-Ecg Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K990476
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026