HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K with Easi-Ecg Option

K990476510(k) TraditionalCleared Jul 22, 1999156-day reviewClass II21 CFR 870.1025

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hewlett-Packard GmbH · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Jul 22, 1999 as substantially equivalent to K872781 (Totemite Medical Electronics, 1987) and 1 other predicate after a 156-day review. Its family tree reaches back 1 generation to 1987. 437 later devices descend from it.

Decided
Jul 22, 1999
Review
156 days
Class
Class II
Product code
MHX
Recalls
None
Predicates
2

Its longest line of descent runs back 1 generation to K872781 (1987). 2 ancestors in all. 437 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1999: 1 descendant cleared2001: 1 descendant cleared2004: 10 descendants cleared2003: 8 descendants cleared2002: 4 descendants cleared2006: 17 descendants cleared2010: 21 descendants cleared2011: 21 descendants cleared2007: 7 descendants cleared2012: 19 descendants cleared2009: 19 descendants cleared2008: 14 descendants cleared2005: 10 descendants cleared2013: 17 descendants cleared2015: 14 descendants cleared2018: 25 descendants cleared2016: 17 descendants cleared2014: 16 descendants cleared2017: 17 descendants cleared2021: 19 descendants cleared2025: 29 descendants cleared2026: 20 descendants cleared2022: 28 descendants cleared2023: 27 descendants cleared2020: 14 descendants cleared2024: 25 descendants cleared2019: 17 descendants clearedReference device: K882609 · Models M1175A & M1176A Component Monitoring System · Waltham Div. Hewlett Packard · 1989Reference device: K973437 · Hewlett-Packard Models M1001B Ecg M1002B Ecg/Respiration Plug-in Modules · Hewlett-Packard GmbH · 1997K872781 · Easi Lead System Cable Assembly · Totemite Medical Electronics · 19871990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 29 descendants3 generations forward: 62 descendants4 generations forward: 59 descendants5 generations forward: 62 descendants6 generations forward: 73 descendants7 generations forward: 61 descendants8 generations forward: 34 descendants9 generations forward: 21 descendants10 generations forward: 25 descendants11 generations forward: 7 descendants12 generations forward: 1 descendant K872781 · Easi Lead System Cable Assembly · Totemite Medical Electronics · 19870+5+10
  1. Root · 1 generation backK872781 Easi Lead System Cable AssemblyTotemite Medical Electronics · 1987
  2. This deviceK990476 HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model…Hewlett-Packard GmbH · 1999
  3. Descendant · cites it directly · 436 of its ownK992595 HP Viridia Component Monitoring System, Model Rev. K with Easi-ST…Hewlett-Packard GmbH · 1999
  4. Descendant · cites it directly · 431 of its ownK013199 Philips Component Compact Monitor,Model M1275B, Philips Compenent…Philips Medical Systems, Inc. · 2001
  5. Descendant · 2 generations on · 421 of its ownK021300 Modification of Philips Component Compact Monitor, Release a.03…Philips Medical Systems · 2002
  6. Descendant · 2 generations on · 416 of its ownK020531 Philips Multi-Parameter Module, Model M3001A; Philips Component…Philips Medical Systems · 2002
  7. Descendant · 2 generations on · 407 of its ownK021778 Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems · 2002

Showing 5 of 437 descendants. See all descendants

Descendants cleared per year

1999: 12001: 12002: 42003: 82004: 102005: 102006: 172007: 72008: 142009: 192010: 212011: 212012: 192013: 172014: 162015: 142016: 172017: 172018: 252019: 172020: 142021: 192022: 282023: 272024: 252025: 292026: 20
1999high 29 (2025) · low 0 (2000)2026

Most influential descendants

Showing 5 of 437 · See all
  • K992595HP Viridia Component Monitoring System, Model Rev. K with Easi-ST Segment Measurement…Hewlett-Packard GmbH · 1999 · cites it directly436descendants
  • K013199Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model M1276BPhilips Medical Systems, Inc. · 2001 · cites it directly431descendants
  • K021300Modification of Philips Component Compact Monitor, Release a.03, Model M1275BPhilips Medical Systems · 2002 · 2 generations on421descendants
  • K020531Philips Multi-Parameter Module, Model M3001A; Philips Component Compact Monitor, Release…Philips Medical Systems · 2002 · 2 generations on416descendants
  • K021778Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems · 2002 · 2 generations on407descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K872781Easi Lead System Cable AssemblyTotemite Medical Electronics1987—
0 (this device)K990476HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K…Hewlett-Packard GmbH1999—
+1K992595HP Viridia Component Monitoring System, Model Rev. K with Easi-ST Segment…Hewlett-Packard GmbH1999—
+1K013199Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model…Philips Medical Systems, Inc.2001—
+2K021300Modification of Philips Component Compact Monitor, Release a.03, Model M1275BPhilips Medical Systems2002—
+2K020531Philips Multi-Parameter Module, Model M3001A; Philips Component Compact…Philips Medical Systems2002—
+2K021778Philips Component Compact Monitor, Release a.04, Model #M1275BPhilips Medical Systems2002—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199910
200110
200240
200380
2004100
2005100
2006170
200770
2008140
2009190
2010210
2011210
2012190
2013170
2014160
2015140
2016170
2017170
2018250
2019170
2020140
2021190
2022280
2023270
2024250
2025290
2026200
FIG. 1 — Lineage of K990476, 1987–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MHX? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K872781Easi Lead System Cable Assembly
Summary page 1
…ctor and Alarm, Arrythmia 3. The new device with the EASI-ECG Option is substantially equivalent to the following predicate devices: the Totemite EASI Lead System Cable K872781B (12/29/87), and the Zymed T8010 Telemetry Central Station Monitor K951370 (10/6/95). The accessories and materials are optional…
Totemite Medical Electronics1987—
*K951370Zymed Telemetry System; Model T8010
Summary page 1
…bstantially equivalent to the following predicate devices: the Totemite EASI Lead System Cable K872781B (12/29/87), and the Zymed T8010 Telemetry Central Station Monitor K951370 (10/6/95). The accessories and materials are optional and are the same accessories originally cleared for use with HP M1175A/76A…
Zymed Medical Instrumentation1995—
†K882609Models M1175A & M1176A Component Monitoring SystemWaltham Div. Hewlett Packard1989—
†K973437Hewlett-Packard Models M1001B Ecg M1002B Ecg/Respiration Plug-in ModulesHewlett-Packard GmbH1997—

Also cleared against K872781: 1 other device →

What descends from it? (437)

See all 437

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K992595HP Viridia Component Monitoring System, Model Rev. K with Easi-ST Segment Measurement…Hewlett-Packard GmbH1999436
K013199Philips Component Compact Monitor,Model M1275B, Philips Compenent Rack, Model M1276BPhilips Medical Systems, Inc.2001431
K033513Easi Ecg AlgorithmPhilips Medical Systems2004220

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • None of the 2 ancestors has a recall.
  • Of its descendants, K992595 (Hewlett-Packard GmbH, 1999) has gone furthest, with 436 descendants of its own.
  • 1 other device also cites its predicate K872781.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Hewlett-Packard Viridia Component Monitoring System, (M1175A/76A/77RA), and Viridia 24/26 (M1205A) Rev.K with EASI-ECG Option is indicated for:

Assessment of symptoms that may be related to rhythm disturbances of the heart: patients with palpitations, the evaluation of arrhythmia’s in adult and pediatric patients.

Assessment of risk in patients with or without symptoms of arrhythmia.

Assessment of efficacy of antiarrhythmic therapy.

Assessment of pacemaker function.

Assessment of symptomatic or asymptomatic patients to evaluate for ischemic heart disease.

Assessment is indicated for single-hospital environment.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K990476: HP Viridia Component Monitoring System, HP Viridia CMS 24/26, Model Rev. K with Easi-Ecg Option.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K990476

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026