510(k) Report

510(k) · Cleared

Recalled · 2011

XR 46 Dxa (Central) Bone Densitometer, Model XR 46

K992125510(k) TraditionalCleared Nov 22, 1999152-day reviewClass II21 CFR 892.1170Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Norland Medical Systems, Inc. · KGI Densitometer, Bone

Cleared on Nov 22, 1999 as substantially equivalent to K881865 (Norland Corp., 1989) and 3 other predicates after a 152-day review. Its family tree reaches back 2 generations to 1995. It has been recalled once.

Decided
Nov 22, 1999
Review
152 days
Class
Class II
Product code
KGI
Recalls
1
Predicates
4

Its longest line of descent runs back 2 generations to K935454 (1995). 6 ancestors in all, the oldest K881865 (1989). No later device cites it as a predicate yet.

Other ancestors, by year cleared 1998: 2 other ancestors1997: 1 other ancestor1989: 1 other ancestorK935454 · DPX Tissue Quantification Output · Lunar Corp. · 1995K973459 · Norland Body Composition Option for Norland Dexa Bone Densitometers · Norland Corp. · 19981985199019952000
Other ancestors, by generation 1 generation back: 3 other ancestors2 generations back: 1 other ancestor K973459 · Norland Body Composition Option for Norland Dexa Bone Densitometers · Norland Corp. · 1998K935454 · DPX Tissue Quantification Output · Lunar Corp. · 19950
  1. Root · 2 generations backK935454 DPX Tissue Quantification OutputLunar Corp. · 1995
  2. 1 generation backK973459 Norland Body Composition Option for Norland Dexa Bone DensitometersNorland Corp. · 1998
  3. This deviceK992125 XR 46 Dxa (Central) Bone Densitometer, Model XR 46Norland Medical Systems, Inc. · 1999
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K935454DPX Tissue Quantification OutputLunar Corp.1995—
−1K973459Norland Body Composition Option for Norland Dexa Bone DensitometersNorland Corp.1998—
0 (this device)K992125XR 46 Dxa (Central) Bone Densitometer, Model XR 46Norland Medical Systems, Inc.1999—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199710
199820
FIG. 1 — Lineage of K992125, 1995–1999Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KGI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KGI device

  1. #1 K252718 Lunar Astrascore 13.3 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (4)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K881865Norland Model XR-26 X-Ray Bone Densitometer
Summary page 1
…Modet XR 46 DXA Bone Densitometer DXA table (central) bone densitometer Bone densitometer, (21 CFR 892.1170), product code 90KGI; Class II XR 26 DXA bone densitometer (K881865), Body Composition Assessment (K973459), Reference Population (K964307), and Fracture Risk Assessment for Norland bone densito…
Norland Corp.1989—
*K973459Norland Body Composition Option for Norland Dexa Bone Densitometers
Summary page 1
…able (central) bone densitometer Bone densitometer, (21 CFR 892.1170), product code 90KGI; Class II XR 26 DXA bone densitometer (K881865), Body Composition Assessment (K973459), Reference Population (K964307), and Fracture Risk Assessment for Norland bone densitometers (K980569). The XR 46 is a full f…
Norland Corp.1998—
*K964307Expanded Reference Values for DPX Bon Densitometers
Summary page 1
…Bone densitometer, (21 CFR 892.1170), product code 90KGI; Class II XR 26 DXA bone densitometer (K881865), Body Composition Assessment (K973459), Reference Population (K964307), and Fracture Risk Assessment for Norland bone densitometers (K980569). The XR 46 is a full featured, pencil beam, DXA, table…
Lunar Corp.1997—
*K980569Fracture Risk Assessment Capability for Norland X-Ray Bone Densitometers
Summary page 1
…6 DXA bone densitometer (K881865), Body Composition Assessment (K973459), Reference Population (K964307), and Fracture Risk Assessment for Norland bone densitometers (K980569). The XR 46 is a full featured, pencil beam, DXA, table bone densitometer that scans the Hip, Spine, Forearm, Whole Body, and ot…
Norland Corp.1998—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 23, 2011Not classifiedZ-0213-2012Software designThe DICOM Software, Norland Illuminatus DXA Software with the DICOM Option (Part Number 434A133) version 4.2.0 through 4.3.1, which is installed on the XR-600, XR-800, XR-46 and EXCELL DXA Bone Densitometers, could lead to improper…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 4 years and passes through 3 companies; a new generation was cleared every 2.3 years on median.
  • None of the 6 ancestors has a recall.

How does its category compare?

All KGI devices

KGI Densitometer, Bone: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KGI

3in 2025

2007: 22008: 52009: 12011: 22012: 52013: 12014: 22016: 22018: 12019: 12022: 22023: 12025: 32026 (partial year): 2
2007high 5 (2008) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KGI

156in 2025

2008: 173 days median2012: 142 days median2025: 156 days median
2007high 173 (2008) · low 142 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KGI

1in 2016

2008: 62009: 92011: 12016: 1
2007high 9 (2009) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KGI

8in 2025

2008: 22009: 32011: 52012: 12013: 12014: 22016: 12017: 162018: 62020: 22021: 12022: 12023: 22024: 22025: 8
2007high 16 (2017) · low 0 (2007)2025

MAUDE reports naming product code KGI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—00
2008517362
20091—93
20100—00
20112—15
2012514201
20131—01
20142—02
20150—00
20162—11
20170—016
20181—06
20191—00
20200—02
20210—01
20222—01
20231—02
20240—02
2025315608
20262—0—

What is it indicated for?

The XR 46 performs DXA scans of the AP Spine, Hip, Forearm, Lateral Spine, and Whole Body, as well as other user selectable sites. It provides, BMD (g/cm2), Area (cm2), and BMC (g) values. It compares these values to gelfier and ethnic matched reference populations and provides T-Score and % Young Reference, Z-Score and % age matched, and long term and short term change values. This includes sBMD (mg/cm2). The XR 46 performs soft tissue assessment and…

Read the full statement (171 words)

The XR 46 performs DXA scans of the AP Spine, Hip, Forearm, Lateral Spine, and Whole Body, as well as other user selectable sites. It provides, BMD (g/cm2), Area (cm2), and BMC (g) values. It compares these values to gelfier and ethnic matched reference populations and provides T-Score and % Young Reference, Z-Score and % age matched, and long term and short term change values. This includes sBMD (mg/cm2).

The XR 46 performs soft tissue assessment and provides lean mass, fat mass, percent fat, and total soft tissue values for all scan sites, including Whole Body.

The XR 46 includes a Report Writer that allows the operator to customize the appearance of their reports and to automate the tedious part of the report. It also allows the user to enter ranges of patient values and the specific statements they want to be printed on their reports for each of these ranges.

The bone density measurements from the XR 46 can be used as an aid to physicians in determining fracture risk.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K992125: XR 46 Dxa (Central) Bone Densitometer, Model XR 46.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K992125

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026