510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Silhouette FC

K993090510(k) TraditionalCleared Oct 1, 199915-day reviewClass II21 CFR 892.1680Recalls 13

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems, Inc. · KPR System, X-Ray, Stationary

Cleared on Oct 1, 1999 after a 15-day review. The FDA summary names no predicate by number. It has been recalled 13 times.

Decided
Oct 1, 1999
Review
15 days
Class
Class II
Product code
KPR
Recalls
13
Predicates
0

The FDA summary names no predicate by number.

FIG. 1 — Lineage of K993090Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

The FDA summary names no predicate by number.

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 24, 2024Class IIZ-0550-2025Equipment maintenanceSteel cables in the Wall Stand for certain Proteus XR/a systems serviced by GE HealthCare have not been replaced at the intervals detailed in the Preventative Maintenance Action List.
Jul 12, 2024Class IIZ-2733-2024Error in labelingThe XR system, Proteus XR/a, does not have a de-installation manual describing process steps for de-installing the devices. De-installation instruction process steps must be followed when de-installing these devices to avoid personnel…
Aug 9, 2019Class IIZ-2449-2019Device DesignIntermittently not receiving an audible exposure indication upon completion of an exposure on the Proteus stationary radiographic system.
Mar 29, 2019Class IIZ-1752-2019Device DesignThere has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.
Jan 30, 2015Class IIZ-1248-2015Device DesignPotential breakage of the steel cable which supports the bucky device in certain wall stands of Proteus XR/a X-ray imaging systems. A fall of a bucky while the system is in use could result in an injury to a patient or operator.
Dec 8, 2014Class IIZ-0954-2015Nonconforming Material/ComponentGE Healthcare has recently become aware of a potential safety issue involving the breakage of the two steel cables which support the bucky device in the wall stand of Proteus XR/a X-ray imaging systems. No injuries reported.
Jun 4, 2014Class IIZ-1988-2014Equipment maintenanceGE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident…
Aug 13, 2009Not classifiedZ-0526-2010Nonconforming Material/ComponentCertain GE Radiographic fixed X-ray tables (Proteus XR/a, Revolution XR/d, and Definium 8000) were found to have tabletops that move unexpectedly in a bi-directional mode. In rare cases, a malfunction may occur, which could result in the…
Nov 1, 2007Not classifiedZ-0842-2008Process controlThe warning label that is required by 21 CFR 1020.30(j) was not on the control console of the Proteus XR/a Radiographic X-ray System, Model 2259976.
Mar 29, 2007Not classifiedZ-1713-2008Radiation Control for Health and Safety ActThe actual average illuminance for the collimators are approximately 140-lux. This does not meet the 160-lux requirement. GE will be correcting the units by replacing all collimators.
May 31, 2006Class IIZ-1790-2012Device DesignAn incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
May 31, 2006Not classifiedZ-1087-2007OtherLabeling; failure to properly label per 21 CFR 1010.2 and 1010.3 -The Operator Console of the Proteus XR/a General Purpose Radiographic Systems were not properly labeled.
Dec 30, 2004Not classifiedZ-0219-05OtherUnits have a defect in accordance with 21 CFR 1003.2 because an incorrect exposure can occur.

No ancestor or descendant has a Class I recall.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

The Silhouette FC is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993090: Silhouette FC.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993090

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026