510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

14 FR Two-Lumen Hemodialysis Catherization Kit with Blue Flextip Arrowg+Ard Blue Catheter for High Volume Infusions, Pro

K993933510(k) TraditionalCleared Sep 14, 2000301-day reviewClass II21 CFR 876.5540Recalls 5 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Arrow Intl., Inc. · MPB Catheter, Hemodialysis, Non-Implanted

Cleared on Sep 14, 2000 as substantially equivalent to K900263 (Arrow Intl., Inc., 1990) and 1 other predicate after a 301-day review. Its family tree reaches back 1 generation to 1990. 5 of the 11 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 5 times, including a Class I recall.

Decided
Sep 14, 2000
Review
301 days
Class
Class II
Product code
MPB
Recalls
5 (1 Class I)
Predicates
2

Its longest line of descent runs back 1 generation to K895417 (1990). 2 ancestors in all. 8 later devices descend from it, 5 citing it directly. 5 of the 11 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1990: 1 other ancestor, 1 with a Class I recall12006: 2 descendants cleared2013: 1 descendant cleared2015: 1 descendant cleared, 1 with a Class I recall12001: 1 descendant cleared, 1 with a Class I recall12003: 1 descendant cleared2007: 1 descendant cleared2017: 1 descendant clearedK895417 · Arrow Large-Bore Dual Lumen Hemodialysis Kit · Arrow Intl., Inc. · 1990 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor, 1 with a Class I recall1 1 generation forward: 5 descendants, 2 with a Class I recall22 generations forward: 3 descendants K895417 · Arrow Large-Bore Dual Lumen Hemodialysis Kit · Arrow Intl., Inc. · 1990 · Class I recall0
  1. Root · 1 generation backK895417 Arrow Large-Bore Dual Lumen Hemodialysis KitArrow Intl., Inc. · 1990 · Recalled (Class I)
  2. This deviceK993933 14 FR Two-Lumen Hemodialysis Catherization Kit with Blue Flextip…Arrow Intl., Inc. · 2000
  3. Descendant · cites it directly · 1 of its ownK122091 Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray Cook Spectrum…Cook, Inc. · 2013
  4. Descendant · cites it directly · 1 of its ownK053589 Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc. · 2006
  5. Descendant · cites it directly · 1 of its ownK011761 Mac Two-Lumen Central Venous Kit with Arrowg Ard Blue Antimicrobial…Arrow Intl., Inc. · 2001 · Recalled (Class I)
  6. Descendant · 2 generations on · 0 of its ownK161504 Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/TrayCook Incorporated · 2017
  7. Descendant · cites it directly · 0 of its ownK143102 Multi-Lumen Acute Hemodialysis Catheter for High Volume InfusionsArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2015 · Recalled (Class I)

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2001: 12003: 12006: 22007: 12013: 12015: 12017: 1
2001high 2 (2006) · low 0 (2002)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K122091Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray Cook Spectrum Turbo-Flo HD Acute…Cook, Inc. · 2013 · cites it directly1descendant
  • K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc. · 2006 · cites it directly1descendant
  • K011761Mac Two-Lumen Central Venous Kit with Arrowg Ard Blue Antimicrobial Surface and Integral…Arrow Intl., Inc. · 2001 · cites it directly1descendant
  • K161504Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/TrayCook Incorporated · 2017 · 2 generations on0descendants
  • K143102Multi-Lumen Acute Hemodialysis Catheter for High Volume InfusionsArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2015 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K895417Arrow Large-Bore Dual Lumen Hemodialysis KitArrow Intl., Inc.1990Class I
0 (this device)K99393314 FR Two-Lumen Hemodialysis Catherization Kit with Blue Flextip Arrowg+Ard…Arrow Intl., Inc.2000—
+1K122091Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray Cook Spectrum Turbo-Flo…Cook, Inc.2013—
+1K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc.2006—
+1K011761Mac Two-Lumen Central Venous Kit with Arrowg Ard Blue Antimicrobial Surface…Arrow Intl., Inc.2001Class I
+2K161504Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/TrayCook Incorporated2017—
+1K143102Multi-Lumen Acute Hemodialysis Catheter for High Volume InfusionsArrow International, Inc. (Subsidiary of Teleflex, Inc.)2015Class I

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199011

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200111
200310
200620
200710
201310
201511
201710
FIG. 1 — Lineage of K993933, 1990–2017Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MPB device

  1. #1 K250836 14F Duo-Flow® Side X Side Double Lumen Catheterscore 10.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K900263Antimicrobial Multi-Lumen Central Venous Catheter
Summary page 1
…0, Catheter, Hemodialysis, non-implanted Legally marketed devices to which the device is substantially equivalent: « Antimicrobial multi-lumen central venous catheter (K900263) * Arrow large-bore dual lumen hemodialysis kit (K895417) Description of device: The materials, construction, and manufacturing…
Arrow Intl., Inc.1990Class I
*K895417Arrow Large-Bore Dual Lumen Hemodialysis Kit
Summary page 1
…d devices to which the device is substantially equivalent: « Antimicrobial multi-lumen central venous catheter (K900263) * Arrow large-bore dual lumen hemodialysis kit (K895417) Description of device: The materials, construction, and manufacturing processes of both the legally marketed 12 Fr Hemodialysi…
Arrow Intl., Inc.1990Class I

Also cleared against K900263: 7 other devices →

What descends from it? (8)

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K122091Cook Turbo-Flo HD Acute Hemodialysis Catheter Set/Tray Cook Spectrum Turbo-Flo HD Acute…Cook, Inc.20131
K053589Long-Term Hemodialysis Catheters with Biobloc CoatingBard Access Systems, Inc.20061
K011761Mac Two-Lumen Central Venous Kit with Arrowg Ard Blue Antimicrobial Surface and Integral… Class IArrow Intl., Inc.20011
K143102Multi-Lumen Acute Hemodialysis Catheter for High Volume Infusions Class IArrow International, Inc. (Subsidiary of Teleflex, Inc.)20150
K060288Retro Silver and Xpresso Silver Chronic Hemodialysis CathetersSpire Biomedical, Inc.20060

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 21, 2026Class IZ-2373-2026Nonconforming Material/ComponentLidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a…
Dec 22, 2017Class IIZ-1336-2018Process controlCertain lots of cardiovascular catheters may not be completely sealed and so sterility of the product cannot be guaranteed.
Oct 17, 2017Class IIZ-0658-2018Process controlThese finished good kits may contain the incorrect Springwire Guide (SWG).
May 9, 2017Class IIZ-2334-2017PackagingArrow International is recalling the affected product because the packaging may not be completely sealed, which may compromise sterility.
Feb 18, 2004Not classifiedZ-0639-04Othercatheter slips out of the suture wing during use.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K900263Antimicrobial Multi-Lumen Central Venous CatheterArrow Intl., Inc. · 1990 · Direct predicate · ◆ in the 1990 ancestors bar of FIG. 12 Class I recalls
  • K895417Arrow Large-Bore Dual Lumen Hemodialysis KitArrow Intl., Inc. · 1990 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K143102Multi-Lumen Acute Hemodialysis Catheter for High Volume InfusionsArrow International, Inc. (Subsidiary of Teleflex, Inc.) · 2015 · Cites it directly · ◆ in the 2015 descendants bar of FIG. 11 Class I recall
  • K011761Mac Two-Lumen Central Venous Kit with Arrowg Ard Blue Antimicrobial Surface and Integral Hemostasis…Arrow Intl., Inc. · 2001 · Cites it directly · ◆ in the 2001 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • 2 of the 2 ancestors have a recall of any class; 2 have a Class I recall.
  • Of its descendants, K122091 (Cook, Inc., 2013) has gone furthest, with 1 descendant of its own.
  • 7 other devices also cite its predicate K900263.

How does its category compare?

All MPB devices

MPB Catheter, Hemodialysis, Non-Implanted: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MPB

1in 2025

2007: 32009: 12010: 12011: 22012: 12013: 12016: 12017: 22020: 32025: 1
2007high 3 (2007) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MPB

205in 2020

2007: 136 days median2020: 205 days median
2007high 205 (2020) · low 136 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MPB

4so far in 2026, the most since 2007

2007: 12017: 22026 (partial year): 4, 3 Class I
2007high 4 (2026) · low 0 (2008)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · MPB

187in 2025

2007: 412008: 1472009: 1662010: 602011: 742012: 782013: 1242014: 822015: 512016: 702017: 862018: 1202019: 1192020: 1792021: 1922022: 1662023: 2032024: 1812025: 187
2007high 203 (2023) · low 41 (2007)2025

MAUDE reports naming product code MPB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20073136141
20080—0147
20091—0166
20101—060
20112—074
20121—078
20131—0124
20140—082
20150—051
20161—070
20172—286
20180—0120
20190—0119
202032050179
20210—0192
20220—0166
20230—0203
20240—0181
20251—0187
20260—4 (3)—

What is it indicated for?

The large-bore two-lumen catheter permits venous access to the central circulation for rapid fluid administration, temporary or acute hemodialysis, apheresis and hemofiltration. It may be inserted into the jugular, subclavian, or femoral veins The ARROWg+ard Blue® antimicrobial surface catheter helps provide protection against catheter-related infections resulting from microorganisms migrating the subcutaneous tract along the exterior surface of the…

Read the full statement (107 words)

The large-bore two-lumen catheter permits venous access to the central circulation for rapid fluid administration, temporary or acute hemodialysis, apheresis and hemofiltration. It may be inserted into the jugular, subclavian, or femoral veins

The ARROWg+ard Blue® antimicrobial surface catheter helps provide protection against catheter-related infections resulting from microorganisms migrating the subcutaneous tract along the exterior surface of the catheter when used for infusion. Clinical data have not been collected that demonstrate the use of the ARROWg+ard Blue® antimicrobial surface in decreasing catheter-related infections in hemodialysis patients. The catheter is not intended to be used as a treatment for existing mfectlons, nor is it indicated for long-term use.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K993933: 14 FR Two-Lumen Hemodialysis Catherization Kit with Blue Flextip Arrowg+Ard Blue Catheter for High Volume Infusions, Pro.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K993933

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026