510(k) · Cleared
Recalled · Class I · 2025Plastic Port Implantable Venous Access System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Sims Deltec, Inc. · LJT Port & Catheter, Implanted, Subcutaneous, Intravascular
Cleared on Jan 27, 2000 as substantially equivalent to K830730 (Pharmacia, Inc., 1983) and 3 other predicates after a 43-day review. Its family tree reaches back 1 generation to 1983. 1 of the 5 devices in its family, counting itself, has a Class I recall. It has been recalled 11 times, including 6 Class I recalls.
- Decided
- Jan 27, 2000
- Review
- 43 days
- Class
- Class II
- Product code
- LJT
- Recalls
- 11 (6 Class I)
- Predicates
- 4
Its longest line of descent runs back 1 generation to K830730 (1983). 4 ancestors in all. No later device cites it as a predicate yet. 1 of the 5 devices in this family, counting this one, has a Class I recall.
- Root · 1 generation backK830730 Port-a-Cath Implantable Cath. SysPharmacia, Inc. · 1983 · Recalled
- This deviceK994216 Plastic Port Implantable Venous Access SystemSims Deltec, Inc. · 2000
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K830730 | Port-a-Cath Implantable Cath. Sys | Pharmacia, Inc. | 1983 | 1 |
| 0 (this device) | K994216 | Plastic Port Implantable Venous Access System | Sims Deltec, Inc. | 2000 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1988 | 1 | 0 |
| 1994 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LJT device
- #1 K251253 PowerPort™ isp Implantable Port; PowerPort™…score 14.9 · 9 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (5)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K830730 | Port-a-Cath Implantable Cath. SysSummary page 2…PARISON Plastic Port Systems PORT-A-CATH® M.R.L.® Implanted Systems Port MANUFACTURER SIMS Deltec, Inc. SIMS Deltec, Inc. Bard Access System 510(K) NUMBER Subject Device K830730B Unknown K875276 K932840 K942024 INDICATION FOR USE A system is indicated A system is indicated The BardPort Implanted when patie… | 1983 |
| * | K875276 | Port-a-Cath (R) Implantable Drug Delivery SystemSummary page 2…ort Systems PORT-A-CATH® M.R.L.® Implanted Systems Port MANUFACTURER SIMS Deltec, Inc. SIMS Deltec, Inc. Bard Access System 510(K) NUMBER Subject Device K830730B Unknown K875276 K932840 K942024 INDICATION FOR USE A system is indicated A system is indicated The BardPort Implanted when patient therapy when p… | 1988 |
| * | K932840 | Port-a-Cath Implantable Vascular Access SystemSummary page 2…ems PORT-A-CATH® M.R.L.® Implanted Systems Port MANUFACTURER SIMS Deltec, Inc. SIMS Deltec, Inc. Bard Access System 510(K) NUMBER Subject Device K830730B Unknown K875276 K932840 K942024 INDICATION FOR USE A system is indicated A system is indicated The BardPort Implanted when patient therapy when patient t… | 1994 |
| * | K942024 | Port-a-Cath Low Profile Implantable Venous Access SystemSummary page 2…-A-CATH® M.R.L.® Implanted Systems Port MANUFACTURER SIMS Deltec, Inc. SIMS Deltec, Inc. Bard Access System 510(K) NUMBER Subject Device K830730B Unknown K875276 K932840 K942024 INDICATION FOR USE A system is indicated A system is indicated The BardPort Implanted when patient therapy when patient therapy P… | 1994 |
| * | — | M.R.L® Implanted Ports Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (1 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Feb 13, 2025 | Class I | Z-1287-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Feb 13, 2025 | Class I | Z-1281-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Feb 13, 2025 | Class I | Z-1282-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Feb 13, 2025 | Class I | Z-1280-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Feb 13, 2025 | Class I | Z-1283-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Feb 13, 2025 | Class I | Z-1288-2025 | Process control | The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect. |
| Mar 29, 2004 | Not classified | Z-0874-04 | Other | Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization. |
| Mar 29, 2004 | Not classified | Z-0871-04 | Other | Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization. |
| Mar 29, 2004 | Not classified | Z-0875-04 | Other | Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization. |
| Mar 29, 2004 | Not classified | Z-0872-04 | Other | Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization. |
| Mar 29, 2004 | Not classified | Z-0873-04 | Other | Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization. |
The 5 latest of 11 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- 3 of the 4 ancestors have a recall of any class.
- 1 other device also cites its predicate K830730.
How does its category compare?
All LJT devicesLJT Port & Catheter, Implanted, Subcutaneous, Intravascular: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LJT
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LJT
182in 2016
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LJT
1in 2022
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · LJT
4,308in 2025
MAUDE reports naming product code LJT, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 85 | 1 | 208 |
| 2008 | 7 | 119 | 2 | 161 |
| 2009 | 2 | — | 2 | 172 |
| 2010 | 2 | — | 19 | 207 |
| 2011 | 5 | 144 | 1 | 202 |
| 2012 | 2 | — | 2 | 153 |
| 2013 | 7 | 150 | 0 | 141 |
| 2014 | 2 | — | 0 | 143 |
| 2015 | 4 | 164 | 1 | 149 |
| 2016 | 3 | 182 | 0 | 135 |
| 2017 | 0 | — | 1 | 262 |
| 2018 | 0 | — | 1 | 878 |
| 2019 | 1 | — | 7 | 765 |
| 2020 | 1 | — | 0 | 959 |
| 2021 | 0 | — | 0 | 1,285 |
| 2022 | 0 | — | 1 | 1,334 |
| 2023 | 1 | — | 0 | 1,221 |
| 2024 | 1 | — | 0 | 1,896 |
| 2025 | 2 | — | 0 | 4,308 |
| 2026 | 1 | — | 1 (1) | — |
What is it indicated for?
A system is indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K994216 · printed Nov 4, 2026
Cite this page
510(k) Report. “K994216: Plastic Port Implantable Venous Access System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K994216
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026