Cadd-Prizm PCS II Model 6101 Ambulatory Infusion System

K000842510(k) TraditionalCleared Jun 8, 200085-day reviewClass II21 CFR 880.5725Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Sims Deltec, Inc. · FRN Pump, Infusion

Cleared on Jun 8, 2000 after an 85-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 of the 17 devices in its family, counting itself, have a Class I recall. 16 later devices descend from it. It has been recalled 2 times.

Decided
Jun 8, 2000
Review
85 days
Class
Class II
Product code
FRN
Recalls
2
Predicates
0

16 later devices descend from it, 2 citing it directly. 2 of the 17 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2013: 1 descendant cleared2004: 1 descendant cleared2005: 2 descendants cleared2020: 1 descendant cleared, 1 with a Class I recall12014: 2 descendants cleared2017: 2 descendants cleared2016: 2 descendants cleared2008: 1 descendant cleared, 1 with a Class I recall12023: 1 descendant cleared2024: 2 descendants cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK000842 Cadd-Prizm PCS II Model 6101 Ambulatory Infusion SystemSims Deltec, Inc. · 2000
  2. Descendant · cites it directly · 7 of its ownK123049 Sapphire Infusion PumpQ Core Medical, Ltd. · 2013
  3. Descendant · cites it directly · 7 of its ownK042228 Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004
  4. Descendant · 2 generations on · 6 of its ownK050770 Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005
  5. Descendant · 3 generations on · 5 of its ownK052451 MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005
  6. Descendant · 3 generations on · 4 of its ownK081905 Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · Recalled (Class I)

Showing 5 of 16 descendants. See all descendants

Descendants cleared per year

2004: 12005: 22008: 12013: 12014: 22016: 22017: 22020: 12023: 12024: 22026: 1
2004high 2 (2005) · low 0 (2006)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 16 · See all
  • K123049Sapphire Infusion PumpQ Core Medical, Ltd. · 2013 · cites it directly7descendants
  • K042228Beeline Motiv, Propolis, PcaMckinley, Inc. · 2004 · cites it directly7descendants
  • K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC · 2005 · 2 generations on6descendants
  • K052451MPS Acacia Medflo Elastomeric PumpMps Acacia · 2005 · 3 generations on5descendants
  • K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · 3 generations on4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K000842Cadd-Prizm PCS II Model 6101 Ambulatory Infusion SystemSims Deltec, Inc.2000—
+1K123049Sapphire Infusion PumpQ Core Medical, Ltd.2013—
+1K042228Beeline Motiv, Propolis, PcaMckinley, Inc.2004—
+2K050770Modification to Accufuser, Accufuser Plus Standard Procedure KitMckinley Medical, LLC2005—
+3K052451MPS Acacia Medflo Elastomeric PumpMps Acacia2005—
+3K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc.2008Class I

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200520
200811
201310
201420
201620
201720
202011
202310
202420
202610
FIG. 1 — Lineage of K000842, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FRN device

  1. #1 K253703 RemunityPRO+ Pump for Remodulin®…score 12.6 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (16)

See all 16

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K123049Sapphire Infusion PumpQ Core Medical, Ltd.20137
K042228Beeline Motiv, Propolis, PcaMckinley, Inc.20047

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 19, 2007Not classifiedZ-0718-2007OtherThe year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into an immediate 'Clock battery is Low/Service Immediately' or 'Clock Battery needs service…
Jun 23, 2003Not classifiedZ-0980-03OtherPump keyboard entries by the patient could have resulted in the patient having unintended access to programming screens and led to in improper drug dosage.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K192860Sapphire Infusion Pump, Sapphire administration setsQ Core Medical, Ltd. · 2020 · 2 generations on · ◆ in the 2020 descendants bar of FIG. 11 Class I recall
  • K081905Accuflow and Accuflux Elastomeric Infusion PumpWestmed, Inc. · 2008 · 3 generations on · ◆ in the 2008 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K123049 (Q Core Medical, Ltd., 2013) has gone furthest, with 7 descendants of its own.

How does its category compare?

All FRN devices

FRN Pump, Infusion: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · FRN

8in 2025

2007: 122008: 82009: 92010: 32011: 32012: 42013: 62014: 62015: 32016: 32017: 32018: 32019: 22020: 12021: 22022: 62023: 72024: 52025: 82026 (partial year): 2
2007high 12 (2007) · low 1 (2020)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FRN

172in 2025

2007: 102 days median2008: 89 days median2009: 66 days median2010: 93 days median2011: 103 days median2012: 217 days median2013: 312 days median2014: 172 days median2015: 232 days median2016: 301 days median2017: 305 days median2018: 267 days median2022: 268 days median2023: 265 days median2024: 103 days median2025: 172 days median
2007high 312 (2013) · low 66 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FRN

7in 2025

2007: 152008: 242009: 1022010: 372011: 692012: 42, 2 Class I2013: 22, 3 Class I2014: 4, 3 Class I2015: 52016: 7, 1 Class I2017: 132018: 12019: 18, 1 Class I2020: 16, 5 Class I2021: 3, 2 Class I2022: 22023: 16, 10 Class I2024: 9, 4 Class I2025: 7, 4 Class I2026 (partial year): 11, 2 Class I
2007high 102 (2009) · low 1 (2018)2026

Class I (37 since 2007) Class II/III not classified

Adverse event reports per year · FRN

52,792in 2025

2007: 1,0862008: 8,2612009: 1,9522010: 6,1332011: 41,3672012: 15,4822013: 40,0822014: 94,1802015: 26,7422016: 16,5602017: 12,0602018: 16,5912019: 15,8502020: 94,1862021: 199,3422022: 888,0202023: 165,8382024: 57,7302025: 52,792
2007high 888,020 (2022) · low 1,086 (2007)2025

MAUDE reports naming product code FRN, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200712102151,086
2008889248,261
20099661021,952
2010393376,133
201131036941,367
2012421742 (2)15,482
2013631222 (3)40,082
201461724 (3)94,180
20153232526,742
201633017 (1)16,560
201733051312,060
20183267116,591
20192—18 (1)15,850
20201—16 (5)94,186
20212—3 (2)199,342
202262682888,020
2023726516 (10)165,838
202451039 (4)57,730
202581727 (4)52,792
20262—11 (2)—
Cite this page

510(k) Report. “K000842: Cadd-Prizm PCS II Model 6101 Ambulatory Infusion System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K000842

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026