510(k) Report

PMA · Approved

Recalled · Class II · 2026

Lorad Digital Breast Imager

P010025PMAApproved Mar 15, 2002319-day reviewClass II21 CFR 892.1715Recalls 4

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Hologic, Inc. · MUE Full Field Digital, System, X-Ray, Mammographic

Approved on Mar 15, 2002 through premarket approval, which rests on the device’s own clinical evidence after a 319-day review. 5 later devices descend from it. It has been recalled 4 times.

Decided
Mar 15, 2002
Review
319 days
Class
Class II
Product code
MUE
Recalls
4

5 later devices descend from it, 1 citing it directly.

Descendants, by year cleared 2017: 2 descendants cleared2021: 1 descendant cleared2025: 2 descendants cleared200020052010201520202025
  1. This deviceP010025 Lorad Digital Breast ImagerHologic, Inc. · 2002
  2. Descendant · cites it directly · 4 of its ownK162670 RSM 1824C with RConsole1DRTECH Corporation · 2017
  3. Descendant · 2 generations on · 1 of its ownK202902 2430MCA with Xmaru WRayence Co., Ltd. · 2021
  4. Descendant · 2 generations on · 1 of its ownK170930 RSM 2430CDRTECH Corporation · 2017
  5. Descendant · 3 generations on · 0 of its ownK243849 2430TCA with Xmaru WRayence Co., Ltd. · 2025
  6. Descendant · 3 generations on · 0 of its ownK241113 Vivix-MVieworks Co., Ltd. · 2025

Most influential descendants

Showing 5 of 5 · See all
  • K162670RSM 1824C with RConsole1DRTECH Corporation · 2017 · cites it directly4descendants
  • K2029022430MCA with Xmaru WRayence Co., Ltd. · 2021 · 2 generations on1descendant
  • K170930RSM 2430CDRTECH Corporation · 2017 · 2 generations on1descendant
  • K2438492430TCA with Xmaru WRayence Co., Ltd. · 2025 · 3 generations on0descendants
  • K241113Vivix-MVieworks Co., Ltd. · 2025 · 3 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)P010025Lorad Digital Breast ImagerHologic, Inc.2002—
+1K162670RSM 1824C with RConsole1DRTECH Corporation2017—
+2K2029022430MCA with Xmaru WRayence Co., Ltd.2021—
+2K170930RSM 2430CDRTECH Corporation2017—
+3K2438492430TCA with Xmaru WRayence Co., Ltd.2025—
+3K241113Vivix-MVieworks Co., Ltd.2025—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201720
202110
202520
FIG. 1 — Lineage of P010025, 2002–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA PMA database (decision fields and supplements), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MUE? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MUE device

  1. #1 K253678 MAMMOMAT B.brilliant (11372950)score 13.6 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

Does it have a predicate?

PMA decisions do not claim a predicate device.

What descends from it? (5)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K162670RSM 1824C with RConsole1DRTECH Corporation20174

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 8, 2026Class IIZ-2167-2026Device DesignComplaints have been received of systems developing loose, missing, or broken internal bolts over time.
May 13, 2014Class IIZ-1700-2014Component change controlSelenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
Sep 6, 2013Class IIZ-2301-2013Software designSoftware: When the Selenia upon meeting the hard drive threshold of 4,000 images, may delete the most recently dated images
Feb 28, 2008Not classifiedZ-1546-2008Software designMarker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K162670 (DRTECH Corporation, 2017) has gone furthest, with 4 descendants of its own.

How does its category compare?

All MUE devices

MUE Full Field Digital, System, X-Ray, Mammographic: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MUE

3in 2025

2007: 62008: 52009: 82010: 102011: 102012: 42013: 72014: 12015: 12016: 12017: 82018: 22019: 22020: 42021: 42022: 32023: 12024: 12025: 32026 (partial year): 4
2007high 10 (2010) · low 1 (2014)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MUE

255in 2025

2011: 192 days median2012: 123 days median2013: 207 days median2017: 95 days median2020: 188 days median2021: 164 days median2022: 404 days median2025: 255 days median2026 (partial year): 185 days median
2007high 404 (2022) · low 95 (2017)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MUE

4in 2025

2008: 12009: 12010: 22011: 52014: 22017: 62018: 12019: 12023: 12025: 42026 (partial year): 1
2007high 6 (2017) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MUE

40in 2025

2008: 42009: 32010: 22011: 102012: 62013: 112014: 72015: 62016: 192017: 1152018: 1452019: 802020: 492021: 252022: 362023: 292024: 332025: 40
2007high 145 (2018) · low 0 (2007)2025

MAUDE reports naming product code MUE, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20076—00
20085—14
20098—13
201010—22
201110192510
2012412306
20137207011
20141—27
20151—06
20161—019
20178956115
20182—1145
20192—180
20204188049
20214164025
20223404036
20231—129
20241—033
20253255440
202641851—
Cite this page

510(k) Report. “P010025: Lorad Digital Breast Imager.” Lineage record from FDA PMA data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/P010025

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026