Oncology Salvage System

K002757510(k) TraditionalCleared Nov 24, 200080-day reviewClass II21 CFR 888.3350Recalls 14

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Biomet, Inc. · JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Cleared on Nov 24, 2000 after an 80-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 62 later devices descend from it. It has been recalled 14 times.

Decided
Nov 24, 2000
Review
80 days
Class
Class II
Product code
JDI
Recalls
14
Predicates
0

62 later devices descend from it, 21 citing it directly.

Descendants, by year cleared 2004: 3 descendants cleared2005: 4 descendants cleared2012: 6 descendants cleared2019: 4 descendants cleared2002: 4 descendants cleared2009: 2 descendants cleared2014: 4 descendants cleared2006: 2 descendants cleared2008: 2 descendants cleared2013: 4 descendants cleared2015: 5 descendants cleared2007: 2 descendants cleared2010: 2 descendants cleared2011: 1 descendant cleared2026: 2 descendants cleared2025: 5 descendants cleared2017: 3 descendants cleared2024: 1 descendant cleared2018: 2 descendants cleared2023: 1 descendant cleared2021: 3 descendants cleared200020052010201520202025
  1. This deviceK002757 Oncology Salvage SystemBiomet, Inc. · 2000
  2. Descendant · cites it directly · 19 of its ownK053505 Regenerex Ultra Porous Construct-Femoral and Tibial Knee AugmentsBiomet Manufacturing, Inc. · 2006
  3. Descendant · 2 generations on · 13 of its ownK080361 Regenerex Tibial ComponentsBiomet, Inc. · 2008 · Recalled
  4. Descendant · cites it directly · 12 of its ownK041352 Compress Distal Femoral Replacement SystemBiomet, Inc. · 2004
  5. Descendant · cites it directly · 11 of its ownK052685 Orthopedic Salvage System- OssBiomet, Inc. · 2005 · Recalled
  6. Descendant · cites it directly · 11 of its ownK043547 Compress Segmental Femoral Replacement SystemBiomet, Inc. · 2005 · Recalled

Showing 5 of 62 descendants. See all descendants

Descendants cleared per year

2002: 42004: 32005: 42006: 22007: 22008: 22009: 22010: 22011: 12012: 62013: 42014: 42015: 52017: 32018: 22019: 42021: 32023: 12024: 12025: 52026: 2
2002high 6 (2012) · low 0 (2003)2026

Most influential descendants

Showing 5 of 62 · See all
  • K053505Regenerex Ultra Porous Construct-Femoral and Tibial Knee AugmentsBiomet Manufacturing, Inc. · 2006 · cites it directly19descendants
  • K080361Regenerex Tibial ComponentsBiomet, Inc. · 2008 · 2 generations on13descendants
  • K041352Compress Distal Femoral Replacement SystemBiomet, Inc. · 2004 · cites it directly12descendants
  • K052685Orthopedic Salvage System- OssBiomet, Inc. · 2005 · cites it directly11descendants
  • K043547Compress Segmental Femoral Replacement SystemBiomet, Inc. · 2005 · cites it directly11descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K002757Oncology Salvage SystemBiomet, Inc.2000—
+1K053505Regenerex Ultra Porous Construct-Femoral and Tibial Knee AugmentsBiomet Manufacturing, Inc.2006—
+2K080361Regenerex Tibial ComponentsBiomet, Inc.20081
+1K041352Compress Distal Femoral Replacement SystemBiomet, Inc.2004—
+1K052685Orthopedic Salvage System- OssBiomet, Inc.20052
+1K043547Compress Segmental Femoral Replacement SystemBiomet, Inc.20056

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200240
200430
200540
200620
200720
200820
200920
201020
201110
201260
201340
201440
201550
201730
201820
201940
202130
202310
202410
202550
202620
FIG. 1 — Lineage of K002757, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JDI device

  1. #1 K253515 Zimmer Biomet OSS™ Orthopedic Salvage System…score 15.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (62)

See all 62

21 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K053505Regenerex Ultra Porous Construct-Femoral and Tibial Knee AugmentsBiomet Manufacturing, Inc.200619
K041352Compress Distal Femoral Replacement SystemBiomet, Inc.200412
K052685Orthopedic Salvage System- OssBiomet, Inc.200511
K043547Compress Segmental Femoral Replacement SystemBiomet, Inc.200511
K021260Reduced Size Oncology Salvage SystemBiomet, Inc.20027
K123501Oss Modular Stems, Oss Diaphyseal SegmentsBiomet, Inc.20136
K051479RS (Reduced Size) Oss (Orthopedic Salvage System) Additional ComponentsBiomet Manufacturing, Inc.20056
K051570Avl Hinge Knee SystemBiomet, Inc.20054
K141331Biomet Orthopaedic Salvage System (Oss)Biomet, Inc.20143
K121029Mets Modular Distal FemurStanmore Implants Worldwide, Ltd.20123

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Nov 25, 2020Class IIZ-0623-2021Process controlSpecific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.
Apr 17, 2020Class IIZ-2147-2020Process controlPotential presence of elevated endotoxin levels that exceed the specification limit
Oct 10, 2018Class IIZ-0590-2019Under Investigation by firmZimmer Biomet is conducting a lot specific medical device recall for two lots of the Orthopedic Salvage System (OSS) and the Comprehensive Shoulder System. The investigation determined that the Mini Humeral Stem is potentially labeled as…
Jun 5, 2017Class IIZ-0303-2018Labeling False and MisleadingPossibility that the patient label inside the product box was incorrectly labeled. The outer box label and actual product are correctly identified but the internal patient label could be incorrect.
Nov 1, 2011Not classifiedZ-0417-2012Mixed-up of materials/componentsBiomet has initiated this action following an investigation which identified that Part 150367, (OSS Cemented 1M Stem 13x150) was incorrectly labeled as part 150368 (OSS Cemented 1M Stem 14x150).
May 19, 2009Not classifiedZ-1528-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 19, 2009Not classifiedZ-1530-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 19, 2009Not classifiedZ-1527-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 19, 2009Not classifiedZ-1529-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 19, 2009Not classifiedZ-1531-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 19, 2009Not classifiedZ-1526-2009Process controlThe retaining ring may fall out of the instrument during surgery.
May 15, 2008Not classifiedZ-1881-2008Labeling False and MisleadingMislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while clarifying package…
May 15, 2008Not classifiedZ-1880-2008Labeling False and MisleadingMislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while clarifying package…
May 15, 2008Not classifiedZ-1879-2008Labeling False and MisleadingMislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while clarifying package…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K053505 (Biomet Manufacturing, Inc., 2006) has gone furthest, with 19 descendants of its own.

How does its category compare?

All JDI devices

JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JDI

1in 2025

2007: 32008: 52009: 32011: 22012: 42013: 32014: 32015: 32016: 42017: 12018: 12019: 22020: 72021: 32022: 32023: 22024: 32025: 12026 (partial year): 2
2007high 7 (2020) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JDI

59in 2024

2007: 81 days median2008: 96 days median2009: 65 days median2012: 167 days median2013: 176 days median2014: 98 days median2015: 190 days median2016: 166 days median2020: 247 days median2021: 162 days median2022: 69 days median2024: 59 days median
2007high 247 (2020) · low 59 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JDI

1in 2025

2007: 262008: 62009: 32010: 32011: 52012: 42013: 112014: 32015: 12016: 52017: 52018: 12019: 12020: 772025: 1
2007high 77 (2020) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JDI

1,644in 2025

2007: 6962008: 8102009: 9642010: 2,0402011: 3,1602012: 5,1222013: 5,2062014: 4,9682015: 4,9062016: 3,9952017: 6,1352018: 6,7282019: 7,4072020: 4,0152021: 3,1702022: 1,9782023: 1,6352024: 1,8212025: 1,644
2007high 7,407 (2019) · low 696 (2007)2025

MAUDE reports naming product code JDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200738126696
20085966810
20093653964
20100—32,040
20112—53,160
2012416745,122
20133176115,206
201439834,968
2015319014,906
2016416653,995
20171—56,135
20181—16,728
20192—17,407
20207247774,015
2021316203,170
202236901,978
20232—01,635
202435901,821
20251—11,644
20262—0—
Cite this page

510(k) Report. “K002757: Oncology Salvage System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K002757

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026