510(k) · Cleared
Recalled · 2006PHD Personal Hemodialysis Instrument-1M101, PHD Water Pre-Treatment Kit-1W605,PHD Hard Plumbing Kit-1W604, PHD Blood Tub
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Aksys, Ltd. · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
Cleared on Mar 26, 2002 as substantially equivalent to K881270 (Gambro, Inc., 1988) and 10 other predicates after a 434-day review. Its family tree reaches back 1 generation to 1977. 12 later devices descend from it. It has been recalled once.
- Decided
- Mar 26, 2002
- Review
- 434 days
- Class
- Class II
- Product code
- KDI
- Recalls
- 1
- Predicates
- 11
Its longest line of descent runs back 1 generation to K770691 (1977). 11 ancestors in all. 12 later devices descend from it, 5 citing it directly.
- Root · 1 generation backK770691 Gambro Venous Blood LineGambro, Inc. · 1977
- This deviceK010131 PHD Personal Hemodialysis Instrument-1M101, PHD Water Pre-Treatment…Aksys, Ltd. · 2002
- Descendant · cites it directly · 5 of its ownK112314 Nephros Olpur H2H Hemodiafiltration (HDF) Module and Olpur MD 220…Nephros, Inc. · 2012
- Descendant · cites it directly · 5 of its ownK043436 Nxstage Dialysate Preparation Module, Model NX0269Nxstage Medical, Inc. · 2005
- Descendant · 2 generations on · 3 of its ownK060296 Nxstage Dialysate Preparation ModuleNxstage Medical, Inc. · 2006
- Descendant · cites it directly · 2 of its ownK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024 · Recalled
- Descendant · 3 generations on · 2 of its ownK080919 Nxstage Pureflow SLNxstage Medical, Inc. · 2008 · Recalled
Showing 5 of 12 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 12 · See all- K112314Nephros Olpur H2H Hemodiafiltration (HDF) Module and Olpur MD 220 HemodiafilterNephros, Inc. · 2012 · cites it directly5descendants
- K043436Nxstage Dialysate Preparation Module, Model NX0269Nxstage Medical, Inc. · 2005 · cites it directly5descendants
- K060296Nxstage Dialysate Preparation ModuleNxstage Medical, Inc. · 2006 · 2 generations on3descendants
- K2315345008X Hemodialysis SystemFresenius Medical Care · 2024 · cites it directly2descendants
- K080919Nxstage Pureflow SLNxstage Medical, Inc. · 2008 · 3 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K770691 | Gambro Venous Blood Line | Gambro, Inc. | 1977 | — |
| 0 (this device) | K010131 | PHD Personal Hemodialysis Instrument-1M101, PHD Water Pre-Treatment… | Aksys, Ltd. | 2002 | — |
| +1 | K112314 | Nephros Olpur H2H Hemodiafiltration (HDF) Module and Olpur MD 220 Hemodiafilter | Nephros, Inc. | 2012 | — |
| +1 | K043436 | Nxstage Dialysate Preparation Module, Model NX0269 | Nxstage Medical, Inc. | 2005 | — |
| +2 | K060296 | Nxstage Dialysate Preparation Module | Nxstage Medical, Inc. | 2006 | — |
| +1 | K231534 | 5008X Hemodialysis System | Fresenius Medical Care | 2024 | 1 |
| +3 | K080919 | Nxstage Pureflow SL | Nxstage Medical, Inc. | 2008 | 1 |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1981 | 1 | 0 |
| 1984 | 1 | 0 |
| 1987 | 1 | 0 |
| 1988 | 2 | 0 |
| 1989 | 2 | 0 |
| 1990 | 1 | 0 |
| 1994 | 1 | 0 |
| 1997 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2008 | 2 | 0 |
| 2011 | 1 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2022 | 1 | 0 |
| 2024 | 2 | 0 |
| 2025 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in KDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KDI device
- #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (12)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K881270 | Gambro Multipurpose System MPS-10Summary page 1…tion system for hemodialysis (per 21 CRF 876.5665) SUBSTANTIALLY EQUIVALENT DEVICES: Hemaodialysis Delivery Systems 1. Gambro Multipurpose System MPS-10 Gambro, Inc. (K881270) 2. Fresenius 2008 Touch Panel Contirol Dialysis System Fresenius USA, Inc. (K890824) 3. RSP® Hemodialysis Machine Travenol Lab… | 1988 |
| * | K890824 | Fresenius 2008 Touch Panel Control Dialysis SystemSummary page 1…Hemaodialysis Delivery Systems 1. Gambro Multipurpose System MPS-10 Gambro, Inc. (K881270) 2. Fresenius 2008 Touch Panel Contirol Dialysis System Fresenius USA, Inc. (K890824) 3. RSP® Hemodialysis Machine Travenol Laboratories, Inc, (Pre-Amendment) Reuse Equipment for Hemeodialysis 4. HR 3000 Home Pat… | 1989 |
| * | K841153 | Adaptor HR-3000 ReuseSummary page 1…® Hemodialysis Machine Travenol Laboratories, Inc, (Pre-Amendment) Reuse Equipment for Hemeodialysis 4. HR 3000 Home Patient Reprocessing Device Colorado Medical, Inc. (K841153) S, K‘egatron II Dialyzer Reprocessing System Minntech Corp. (K904210) 510(K) SUMMARY: PHD™ PERSONAL HEMODIALYSIS SYSTEM PAGE |… | 1984 |
| * | K904210 | Renatron II Dialyzer Reprocessing System RS 8330Summary page 1…e Equipment for Hemeodialysis 4. HR 3000 Home Patient Reprocessing Device Colorado Medical, Inc. (K841153) S, K‘egatron II Dialyzer Reprocessing System Minntech Corp. (K904210) 510(K) SUMMARY: PHD™ PERSONAL HEMODIALYSIS SYSTEM PAGE | PREMARKET NOTIFICATION FOR THE AKSYS PHD SYSTEM Water… | 1990 |
| * | K811678 | Gambro Water Purification Monitor WRDSummary page 2…PREMARKET NOTIFICATION FOR THE AKSYS PHD SYSTEM Water Purification Systems for Hemodialysis 6. Gambro WRO 10-01 Water Purification Monitor Gambro, Inc. (K811678) 7. ZyzaTech Series Reverse Osmosis Systems ZyzaTech Water Systems, Inc. (K964539) Arterial-Venous Blood Tubing Sets for Hemodi… | 1981 |
| * | K964539 | Portable Series Reverse Osmosis Systems 700, 701, 702, 703, 704, 750, Mre-NF, F800-F802…Summary page 2…Systems for Hemodialysis 6. Gambro WRO 10-01 Water Purification Monitor Gambro, Inc. (K811678) 7. ZyzaTech Series Reverse Osmosis Systems ZyzaTech Water Systems, Inc. (K964539) Arterial-Venous Blood Tubing Sets for Hemodialysis 8. Gambro Venous Blood Line Gambro, Inc. (K770691) 9. Erika Mode] 9608 Art… | 1997 |
| * | K770691 | Gambro Venous Blood LineSummary page 2…Tech Series Reverse Osmosis Systems ZyzaTech Water Systems, Inc. (K964539) Arterial-Venous Blood Tubing Sets for Hemodialysis 8. Gambro Venous Blood Line Gambro, Inc. (K770691) 9. Erika Mode] 9608 Arterial Blood Line National Medical Care, Inc. (K870722) Dialysate Chemical Concentrates 10. Hemodialysi… | 1977 |
| * | K870722 | Label CHG for Erika Model 9608 Arterial Blood LineSummary page 2…erial-Venous Blood Tubing Sets for Hemodialysis 8. Gambro Venous Blood Line Gambro, Inc. (K770691) 9. Erika Mode] 9608 Arterial Blood Line National Medical Care, Inc. (K870722) Dialysate Chemical Concentrates 10. Hemodialysis Bath Concentrate Solutions for Hemodialysis. Dial Medical of Florida, Inc. (K… | 1988 |
| * | K864265 | Hemodialysis Bath Concentrate Solutions for HemodiSummary page 2…Line National Medical Care, Inc. (K870722) Dialysate Chemical Concentrates 10. Hemodialysis Bath Concentrate Solutions for Hemodialysis. Dial Medical of Florida, Inc. (K864265) 11. Liquiflo™ and Biocarb™ Dry Bicarbonate Concentrates Fresenivs USA, Inc. (K896111) Transducer Protectors 12. Dualex® Trans… | 1987 |
| * | K896111 | Fresenius Biocarb Hemodialysis Grade Sodium BicarbSummary page 2…alysis Bath Concentrate Solutions for Hemodialysis. Dial Medical of Florida, Inc. (K864265) 11. Liquiflo™ and Biocarb™ Dry Bicarbonate Concentrates Fresenivs USA, Inc. (K896111) Transducer Protectors 12. Dualex® Transducer Protector Millipore Corporation (K934069) DEVICE DESCRIPTION: The PHD® Personal… | 1989 |
| * | K934069 | Millex, Dualex (Plus, Ultra)Summary page 2…5) 11. Liquiflo™ and Biocarb™ Dry Bicarbonate Concentrates Fresenivs USA, Inc. (K896111) Transducer Protectors 12. Dualex® Transducer Protector Millipore Corporation (K934069) DEVICE DESCRIPTION: The PHD® Personal Hemodialysis System is an automated high flux hemodialysis systetn using & double needls… | 1994 |
| * | — | RSP® Hemodialysis Machine Named without a submission number |
What descends from it? (12)
See all 125 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K112314 | Nephros Olpur H2H Hemodiafiltration (HDF) Module and Olpur MD 220 Hemodiafilter | 2012 | 5 |
| K043436 | Nxstage Dialysate Preparation Module, Model NX0269 | 2005 | 5 |
| K231534 | 5008X Hemodialysis System | 2024 | 2 |
| K233557 | HemoCare Hemodialysis System | 2024 | 0 |
| K081043 | Leak Detection Sensor and Tray Accessory Kit | 2008 | 0 |
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 26, 2006 | Not classified | Z-1239-06 | Other | The hemodialysis device was being marketed with a modified treatment length that exceeds the treatment length for which the device was originally designed. |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- None of the 11 ancestors has a recall.
- Of its descendants, K112314 (Nephros, Inc., 2012) has gone furthest, with 5 descendants of its own.
How does its category compare?
All KDI devicesKDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KDI
9in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KDI
199in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KDI
2in 2025
Class I (10 since 2007) Class II/III not classified
Adverse event reports per year · KDI
2,786in 2025
MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 11 | 91 | 4 | 916 |
| 2008 | 6 | 148 | 0 | 1,042 |
| 2009 | 4 | 236 | 3 | 522 |
| 2010 | 9 | 186 | 1 | 823 |
| 2011 | 5 | 85 | 4 | 2,270 |
| 2012 | 6 | 153 | 1 | 1,765 |
| 2013 | 11 | 247 | 2 | 2,360 |
| 2014 | 8 | 164 | 3 | 6,154 |
| 2015 | 5 | 69 | 5 | 4,182 |
| 2016 | 5 | 230 | 4 | 766 |
| 2017 | 8 | 188 | 1 | 963 |
| 2018 | 2 | — | 1 | 1,716 |
| 2019 | 4 | 263 | 6 | 1,587 |
| 2020 | 9 | 266 | 0 | 1,633 |
| 2021 | 2 | — | 2 | 1,944 |
| 2022 | 7 | 259 | 17 | 2,180 |
| 2023 | 7 | 157 | 8 (8) | 2,692 |
| 2024 | 7 | 235 | 4 (2) | 2,441 |
| 2025 | 9 | 199 | 2 | 2,786 |
| 2026 | 6 | 154 | 2 | — |
What is it indicated for?
Indicated for the treatment of the chronic or acute uremic patient where hemodialysis, including high flux dialysis, is prescribed by the physician.
Indicated for hemodialysis in a variety of environments, to include acute care facility, chronic dialysis facility, self care facility, or home setting, where the operator has been trained and certified to be competent in the use of this device by the attending physician.
Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.
https://510kreport.com/device/K010131 · printed Nov 4, 2026
Cite this page
510(k) Report. “K010131: PHD Personal Hemodialysis Instrument-1M101, PHD Water Pre-Treatment Kit-1W605,PHD Hard Plumbing Kit-1W604, PHD Blood Tub.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K010131
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026