5008X Hemodialysis System

K231534510(k) TraditionalCleared Feb 2, 2024248-day reviewClass II21 CFR 876.5860Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Fresenius Medical Care · KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Cleared on Feb 2, 2024 as substantially equivalent to K222952 (Fresenius Medical Care Renal Therapies Group, LLC, 2022) and 5 other predicates after a 248-day review. Its family tree reaches back 12 generations to 1994. 4 of the 59 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled once.

Decided
Feb 2, 2024
Review
248 days
Class
Class II
Product code
KDI
Recalls
1
Predicates
6

Its longest line of descent runs back 12 generations to K921456 (1994). 56 ancestors in all, the oldest K770691 (1977). 2 later devices descend from it, 1 citing it directly. 4 of the 59 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2020: 1 other ancestor2019: 1 other ancestor2017: 1 other ancestor2012: 2 other ancestors2002: 3 other ancestors2015: 1 other ancestor2009: 1 other ancestor2008: 1 other ancestor, 1 with a Class I recall12007: 1 other ancestor2005: 1 other ancestor2004: 1 other ancestor2001: 2 other ancestors1999: 2 other ancestors1998: 1 other ancestor1997: 3 other ancestors1995: 3 other ancestors1994: 2 other ancestors1990: 1 other ancestor1989: 2 other ancestors1988: 2 other ancestors1987: 2 other ancestors1984: 1 other ancestor1981: 1 other ancestor1977: 1 other ancestor2014: 1 other ancestor2000: 2 other ancestors1996: 1 other ancestor2013: 2 other ancestors2011: 1 other ancestor2025: 2 descendants clearedK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK093902 · Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 · Fresenius Medical Care North America · 2010 · recalledK101715 · Fresenius 2008T Hemodialysis Machine Model 190573 · Fresenius Medical Care North America, Design Cente · 2010 · recalledK120017 · Fresenius 2008 Hemodiaysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK121341 · Fresenius 2008T Hemodialysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK124035 · 2008K@Home Hemodialysis Machine · Fresenius Medical Care North America · 2013×2: 2 generations cleared close together, K150708 (2015), K153449 (2016) K173972 · 2008T BlueStar Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2018 · Class I recallK222952 · 2008T BlueStar Hemodailysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2022 · Class I recall19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 5 other ancestors2 generations back: 23 other ancestors, 1 with a Class I recall13 generations back: 6 other ancestors4 generations back: 2 other ancestors5 generations back: 1 other ancestor6 generations back: 2 other ancestors7 generations back: 2 other ancestors9 generations back: 2 other ancestors10 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 1 descendant K222952 · 2008T BlueStar Hemodailysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2022 · Class I recallK173972 · 2008T BlueStar Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2018 · Class I recallK153449 · 2008K2 Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2016 · recalledK150708 · 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis Machine · Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Class I recallK124035 · 2008K@Home Hemodialysis Machine · Fresenius Medical Care North America · 2013K121341 · Fresenius 2008T Hemodialysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK120017 · Fresenius 2008 Hemodiaysis Machine with Bibag System · Fresenius Medical Care North America · 2012 · recalledK101715 · Fresenius 2008T Hemodialysis Machine Model 190573 · Fresenius Medical Care North America, Design Cente · 2010 · recalledK093902 · Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573 · Fresenius Medical Care North America · 2010 · recalledK080964 · Fresenius Model 2008T Hemodialysis Machine · Fresenius Medical Care North America · 2009 · recalledK994267 · Fresenius 2008K · Fresenius Medical Care · 2000 · recalledK921456 · Fresenius 3008/FDS-08 Dialysate Delivery System · Fresenius USA, Inc. · 1994 · recalled−10−50
  1. Root · 12 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 8 generations between
  3. 3 generations backK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  4. 2 generations backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  5. 1 generation backK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Recalled (Class I)
  6. This deviceK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024
  7. Descendant · cites it directly · 1 of its ownK243505 5008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025 · Recalled
  8. Descendant · 2 generations on · 0 of its ownK251851 DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025
Show all 12 generations
  1. Root · 12 generations backK921456 Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc. · 1994 · Recalled
  2. 11 generations backK994267 Fresenius 2008KFresenius Medical Care · 2000 · Recalled
  3. 10 generations backK080964 Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America · 2009 · Recalled
  4. 9 generations backK093902 Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America · 2010 · Recalled
  5. 8 generations backK101715 Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente · 2010 · Recalled
  6. 7 generations backK120017 Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  7. 6 generations backK121341 Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America · 2012 · Recalled
  8. 5 generations backK124035 2008K@Home Hemodialysis MachineFresenius Medical Care North America · 2013
  9. 4 generations backK150708 2008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC · 2015 · Recalled (Class I)
  10. 3 generations backK153449 2008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2016 · Recalled
  11. 2 generations backK173972 2008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · Recalled (Class I)
  12. 1 generation backK222952 2008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Recalled (Class I)
  13. This deviceK231534 5008X Hemodialysis SystemFresenius Medical Care · 2024

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K2435055008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC · 2025 · cites it directly1descendant
  • K251851DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC · 2025 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−12K921456Fresenius 3008/FDS-08 Dialysate Delivery SystemFresenius USA, Inc.19943
−11K994267Fresenius 2008KFresenius Medical Care20005
−10K080964Fresenius Model 2008T Hemodialysis MachineFresenius Medical Care North America20094
−9K093902Modification to:Fresenius 2008T Hemodialysis Machine, Model 190573Fresenius Medical Care North America20102
−8K101715Fresenius 2008T Hemodialysis Machine Model 190573Fresenius Medical Care North America, Design Cente20101
−7K120017Fresenius 2008 Hemodiaysis Machine with Bibag SystemFresenius Medical Care North America20123
−6K121341Fresenius 2008T Hemodialysis Machine with Bibag SystemFresenius Medical Care North America20125
−5K1240352008K@Home Hemodialysis MachineFresenius Medical Care North America2013—
−4K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T…Fresenius Medical Care Renal Therapies Group, LLC2015Class I
−3K1534492008K2 Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC20162
−2K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC2018Class I
−1K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC2022Class I
0 (this device)K2315345008X Hemodialysis SystemFresenius Medical Care2024—
+1K2435055008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC20256
+2K251851DIASAFEplusUS (F00013010)Fresenius Medical Care Renal Therapies Group, LLC2025—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197710
198110
198410
198720
198820
198920
199010
199420
199530
199610
199730
199810
199920
200020
200120
200230
200410
200510
200710
200811
200910
201110
201220
201320
201410
201510
201710
201910
202010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
202520
FIG. 1 — Lineage of K231534, 1994–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KDI device

  1. #1 K243874 FX CorAL 60 (F00012969); FX CorAL 80…score 18.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (6)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K2229522008T BlueStar Hemodailysis Machine
Summary page 7
…ional 510(k) 5.3.1. 5008X Machine Predicate Device The predicate device for the 5008X Hemodialysis System is the 2008T BlueStar™ Hemodialysis system cleared under K222952. The legally marketed secondary predicate device, the Nephros OLpur H2H Module (K112314), is included to support additional thera…
Fresenius Medical Care Renal Therapies Group, LLC2022Class I
*K112314Nephros Olpur H2H Hemodiafiltration (HDF) Module and Olpur MD 220 Hemodiafilter
Summary page 7
…008X Hemodialysis System is the 2008T BlueStar™ Hemodialysis system cleared under K222952. The legally marketed secondary predicate device, the Nephros OLpur H2H Module (K112314), is included to support additional therapy modes. The legally marketed predicate device, Aksys PHD Personal Hemodialysis Instrum…
Nephros, Inc.2012—
*K010131PHD Personal Hemodialysis Instrument-1M101, PHD Water Pre-Treatment Kit-1W605,PHD Hard…
Summary page 7
…e Nephros OLpur H2H Module (K112314), is included to support additional therapy modes. The legally marketed predicate device, Aksys PHD Personal Hemodialysis Instrument (K010131), is included to support the online bolus feature. 5.3.2. 5008X Bloodlines (also referred to as “Bloodline”) Predicate Device…
Aksys, Ltd.20021
*K201207CombiSet SMARTECH Hemodialysis Blood Tubing Set with attached Priming Set and Integrated…
Summary page 7
…5008X Bloodlines (also referred to as “Bloodline”) Predicate Devices The legally marked predicate device is the CombiSet SMARTECH Hemodialysis Bloodlines cleared under K201207. The legally marketed secondary predicate device is the CAREline Airless Hemodialysis Bloodlines cleared under K172238. 5.3.3.…
Fresenius Medical Care Renal Therapies Group, LLC2020—
*K172238CAREline Airless Hemodialysis Blood Tubing Sets
Summary page 7
…Set SMARTECH Hemodialysis Bloodlines cleared under K201207. The legally marketed secondary predicate device is the CAREline Airless Hemodialysis Bloodlines cleared under K172238. 5.3.3. DIASAFE®plusUS filter Predicate Device Two (2) DIASAFE®plusUS filters cleared under K182367 are used on the 5008X…
Fresenius Medical Care Renal Therapies Group, LLC2017—
*K182367DIASAFEplusUS
Summary page 7
…is the CAREline Airless Hemodialysis Bloodlines cleared under K172238. 5.3.3. DIASAFE®plusUS filter Predicate Device Two (2) DIASAFE®plusUS filters cleared under K182367 are used on the 5008X hemodialysis machine to filter dialysate. 5.4. Device Description 5.4.1. Device Identificat…
Fresenius Medical Care Renal Therapies Group, LLC2019—

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K2435055008X Hemodialysis SystemFresenius Medical Care Renal Therapies Group, LLC20251

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 29, 2025Class IIZ-1992-2025Process controlPotential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K2229522008T BlueStar Hemodailysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2022 · Direct predicate · ◆ on the line of descent in FIG. 18 Class I recalls
  • K1739722008T BlueStar Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2018 · 2 generations back · ◆ on the line of descent in FIG. 18 Class I recalls
  • K080807Blood Tubing Sets (Sterile Fluid Path)Nxstage Medical, Inc. · 2008 · 2 generations back · ◆ in the 2008 ancestors bar of FIG. 17 Class I recalls
  • K1507082008K Hemodialysis Machine, 2008K@home Hemodialysis Machine, 2008T Hemodialysis MachineFresenius Medical Care Renal Therapies Group, LLC · 2015 · 4 generations back · ◆ on the line of descent in FIG. 18 Class I recalls

What does this lineage say?

  • The line of descent spans 30 years and passes through 5 companies; a new generation was cleared every 1.4 years on median.
  • 21 of the 56 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K243505 (Fresenius Medical Care Renal Therapies Group, LLC, 2025) has gone furthest, with 1 descendant of its own.

How does its category compare?

All KDI devices

KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System: FDA decisions, review times, recalls and adverse event reports over 20 years. 2024, this device’s year, is highlighted.

FDA decisions per year · KDI

9in 2025

2007: 112008: 62009: 42010: 92011: 52012: 62013: 112014: 82015: 52016: 52017: 82018: 22019: 42020: 92021: 22022: 72023: 72024: 72025: 92026 (partial year): 6
2007high 11 (2007) · low 2 (2018)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KDI

199in 2025

2007: 91 days median2008: 148 days median2009: 236 days median2010: 186 days median2011: 85 days median2012: 153 days median2013: 247 days median2014: 164 days median2015: 69 days median2016: 230 days median2017: 188 days median2019: 263 days median2020: 266 days median2022: 259 days median2023: 157 days median2024: 235 days median2025: 199 days median2026 (partial year): 154 days median
2007high 266 (2020) · low 69 (2015)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KDI

2in 2025

2007: 42009: 32010: 12011: 42012: 12013: 22014: 32015: 52016: 42017: 12018: 12019: 62021: 22022: 172023: 8, 8 Class I2024: 4, 2 Class I2025: 22026 (partial year): 2
2007high 17 (2022) · low 0 (2008)2026

Class I (10 since 2007) Class II/III not classified

Adverse event reports per year · KDI

2,786in 2025

2007: 9162008: 1,0422009: 5222010: 8232011: 2,2702012: 1,7652013: 2,3602014: 6,1542015: 4,1822016: 7662017: 9632018: 1,7162019: 1,5872020: 1,6332021: 1,9442022: 2,1802023: 2,6922024: 2,4412025: 2,786
2007high 6,154 (2014) · low 522 (2009)2025

MAUDE reports naming product code KDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200711914916
2008614801,042
200942363522
201091861823
201158542,270
2012615311,765
20131124722,360
2014816436,154
201556954,182
201652304766
201781881963
20182—11,716
2019426361,587
2020926601,633
20212—21,944
20227259172,180
202371578 (8)2,692
202472354 (2)2,441
2025919922,786
202661542—

What is it indicated for?

Machine: The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration. The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine.…

Read the full statement (618 words)

Machine:

The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.

The 5008X Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine.

The bibag System is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag System is intended to be used as one component in the preparation of dialysate according to a physician’s prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.

The 5008X Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.

The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

DIASAFEPlusUS Filter:

The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, non- pyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).

The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.

The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.

5008X Tubing Set:

The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.

The blood tubing sets are indicated for use with a prescribed hemodialyzer. For use with a compatible hemodialyzer, as per the labeling.

Additional Indications for Use for the blood tubing set equipped with Twister AFRC

The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.

Additional Indications for Use for the blood tubing set equipped with CLiC

The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.

The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

CitraSure Disinfectant:

CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K231534: 5008X Hemodialysis System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K231534

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026