510(k) Report

510(k) · Cleared

Recalled · Class II · 2023

Philips BV Pulsera/Endura

K010435510(k) TraditionalCleared Mar 12, 200127-day reviewClass II21 CFR 892.1650Recalls 12

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Philips Medical Systems, Inc. · OWB Interventional Fluoroscopic X-Ray System

Cleared on Mar 12, 2001 after a 27-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 1 of the 91 devices in its family, counting itself, has a Class I recall. 90 later devices descend from it. It has been recalled 12 times.

Decided
Mar 12, 2001
Review
27 days
Class
Class II
Product code
OWB
Recalls
12
Predicates
0

90 later devices descend from it, 3 citing it directly. 1 of the 91 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2009: 3 descendants cleared2018: 7 descendants cleared2003: 1 descendant cleared2008: 2 descendants cleared2005: 1 descendant cleared2006: 2 descendants cleared2022: 3 descendants cleared2023: 8 descendants cleared2010: 2 descendants cleared2026: 12 descendants cleared2016: 1 descendant cleared2015: 3 descendants cleared2021: 6 descendants cleared, 1 with a Class I recall12011: 3 descendants cleared2013: 1 descendant cleared2012: 1 descendant cleared2020: 7 descendants cleared2019: 2 descendants cleared2017: 3 descendants cleared2014: 1 descendant cleared2025: 12 descendants cleared2024: 9 descendants cleared200020052010201520202025
  1. This deviceK010435 Philips BV Pulsera/EnduraPhilips Medical Systems, Inc. · 2001
  2. Descendant · cites it directly · 83 of its ownK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. Descendant · 2 generations on · 52 of its ownK050996 O-Arm Imaging SystemBreakaway Imaging, LLC · 2005 · Recalled
  4. Descendant · 3 generations on · 51 of its ownK060344 O-Arm Imaging SystemBreakaway Imaging, LLC · 2006 · Recalled
  5. Descendant · 2 generations on · 33 of its ownK081672 C-InsightMazor Surgical Technologies, Ltd. · 2008
  6. Descendant · 3 generations on · 31 of its ownK102130 Tenzing SystemMazor Surgical Technologies, Ltd. · 2010

Showing 5 of 90 descendants. See all descendants

Descendants cleared per year

2003: 12005: 12006: 22008: 22009: 32010: 22011: 32012: 12013: 12014: 12015: 32016: 12017: 32018: 72019: 22020: 72021: 62022: 32023: 82024: 92025: 122026: 12
2003high 12 (2025) · low 0 (2004)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 90 · See all
  • K032280Siremobil Iso-C 3DSiemens AG · 2003 · cites it directly83descendants
  • K050996O-Arm Imaging SystemBreakaway Imaging, LLC · 2005 · 2 generations on52descendants
  • K060344O-Arm Imaging SystemBreakaway Imaging, LLC · 2006 · 3 generations on51descendants
  • K081672C-InsightMazor Surgical Technologies, Ltd. · 2008 · 2 generations on33descendants
  • K102130Tenzing SystemMazor Surgical Technologies, Ltd. · 2010 · 3 generations on31descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K010435Philips BV Pulsera/EnduraPhilips Medical Systems, Inc.2001—
+1K032280Siremobil Iso-C 3DSiemens AG2003—
+2K050996O-Arm Imaging SystemBreakaway Imaging, LLC20054
+3K060344O-Arm Imaging SystemBreakaway Imaging, LLC20061
+2K081672C-InsightMazor Surgical Technologies, Ltd.2008—
+3K102130Tenzing SystemMazor Surgical Technologies, Ltd.2010—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200510
200620
200820
200930
201020
201130
201210
201310
201410
201530
201610
201730
201870
201920
202070
202161
202230
202380
202490
2025120
2026120
FIG. 1 — Lineage of K010435, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (90)

See all 90

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K032280Siremobil Iso-C 3DSiemens AG200383
K091918Zen-2090 Pro, Fluoroscopic X-Ray System, Model Zen-2090 ProGenoray Co., Ltd.20094
K183101Zenition 50Philips Medical Systems Nederland B.V.20180

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 4, 2023Class IIZ-0224-2024Device DesignPhilips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
Nov 12, 2021Class IIZ-0355-2022Device DesignInstructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns
Nov 12, 2021Class IIZ-0356-2022Device DesignInstructions for Use do not specify the maximum surface temperature of the X-ray tank (assembly housing the X-ray tube, beam filter & cooling oil) and the image intensifier/detector of the system as required an result in burns
Oct 4, 2011Not classifiedZ-0118-2012Nonconforming Material/ComponentFront plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Oct 4, 2011Not classifiedZ-0116-2012Nonconforming Material/ComponentFront plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Oct 4, 2011Not classifiedZ-0117-2012Nonconforming Material/ComponentFront plate broke in to two pieces causing the C-Arm to suspend on the single central bolt
Dec 14, 2010Not classifiedZ-1604-2011Device DesignThe fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
Dec 14, 2010Not classifiedZ-1605-2011Device DesignThe fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
Sep 10, 2008Not classifiedZ-2463-2010OtherThe C-arm rotation brake does not always function properly, and may not hold the C-arm position if the C-arm is not positioned in its maximum rotation and if it is accidentally bumped by medical staff or patients. Unwanted C-arm movement…
Sep 10, 2008Not classifiedZ-2464-2010OtherThe C-arm rotation brake does not always function properly, and may not hold the C-arm position if the C-arm is not positioned in its maximum rotation and if it is accidentally bumped by medical staff or patients. Unwanted C-arm movement…
Sep 5, 2008Not classifiedZ-0140-2011Component design/selectionReliability of the Stand-Trolley cable may result in loss of images on the monitor during an examination. This intermittent problem can lead to System lock-up, where no radiation is possible and images are not visible on the monitor
Sep 5, 2008Not classifiedZ-0141-2011Component design/selectionReliability of the Stand-Trolley cable may result in loss of images on the monitor during an examination. This intermittent problem can lead to System lock-up, where no radiation is possible and images are not visible on the monitor

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 5 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K032280 (Siemens AG, 2003) has gone furthest, with 83 descendants of its own.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—
Cite this page

510(k) Report. “K010435: Philips BV Pulsera/Endura.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K010435

Report an extraction error