O-Arm Imaging System

K060344510(k) TraditionalCleared Mar 29, 200647-day reviewClass II21 CFR 892.1650Recalls 1

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Breakaway Imaging, LLC · OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Cleared on Mar 29, 2006 as substantially equivalent to K050996 (Breakaway Imaging, LLC, 2005) after a 47-day review. Its family tree reaches back 3 generations to 2000. 2 of the 57 devices in its family, counting itself, have a Class I recall. 51 later devices descend from it. It has been recalled once.

Decided
Mar 29, 2006
Review
47 days
Class
Class II
Product code
OXO
Recalls
1
Predicates
1

Its longest line of descent runs back 3 generations to K003266 (2000). 5 ancestors in all. 51 later devices descend from it, 2 citing it directly. 2 of the 57 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2001: 2 other ancestors, 1 with a Class I recall12008: 1 descendant cleared2010: 2 descendants cleared2026: 8 descendants cleared2015: 2 descendants cleared2021: 3 descendants cleared, 1 with a Class I recall12011: 3 descendants cleared2017: 3 descendants cleared2018: 5 descendants cleared2012: 1 descendant cleared2020: 4 descendants cleared2025: 6 descendants cleared2024: 7 descendants cleared2014: 1 descendant cleared2023: 4 descendants cleared2022: 1 descendant clearedK003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 2000K032280 · Siremobil Iso-C 3D · Siemens AG · 2003K050996 · O-Arm Imaging System · Breakaway Imaging, LLC · 2005 · recalled200020052010201520202025
Other ancestors, by generationDescendants, by generation 3 generations back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 2 descendants2 generations forward: 4 descendants3 generations forward: 4 descendants, 1 with a Class I recall14 generations forward: 12 descendants5 generations forward: 7 descendants6 generations forward: 3 descendants7 generations forward: 2 descendants8 generations forward: 1 descendant9 generations forward: 1 descendant10 generations forward: 1 descendant11 generations forward: 2 descendants12 generations forward: 5 descendants13 generations forward: 5 descendants14 generations forward: 2 descendants K050996 · O-Arm Imaging System · Breakaway Imaging, LLC · 2005 · recalledK032280 · Siremobil Iso-C 3D · Siemens AG · 2003K003266 · Siremobil Iso-C 3D · Siemens Medical Solutions USA, Inc. · 20000+5+10
  1. Root · 3 generations backK003266 Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc. · 2000
  2. 2 generations backK032280 Siremobil Iso-C 3DSiemens AG · 2003
  3. 1 generation backK050996 O-Arm Imaging SystemBreakaway Imaging, LLC · 2005 · Recalled
  4. This deviceK060344 O-Arm Imaging SystemBreakaway Imaging, LLC · 2006
  5. Descendant · cites it directly · 33 of its ownK081672 C-InsightMazor Surgical Technologies, Ltd. · 2008
  6. Descendant · 2 generations on · 31 of its ownK102130 Tenzing SystemMazor Surgical Technologies, Ltd. · 2010
  7. Descendant · 3 generations on · 24 of its ownK110911 Renaissance SystemMazor Robotics, Ltd. · 2011
  8. Descendant · 4 generations on · 23 of its ownK113228 Renaissance SystemMazor Robotics, Ltd. · 2011
  9. Descendant · 5 generations on · 21 of its ownK140167 Renaissance X SystemMazor Robotics, Ltd. · 2014

Showing 5 of 51 descendants. See all descendants

Descendants cleared per year

2008: 12010: 22011: 32012: 12014: 12015: 22017: 32018: 52020: 42021: 32022: 12023: 42024: 72025: 62026: 8
2008high 8 (2026) · low 0 (2009)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 51 · See all
  • K081672C-InsightMazor Surgical Technologies, Ltd. · 2008 · cites it directly33descendants
  • K102130Tenzing SystemMazor Surgical Technologies, Ltd. · 2010 · 2 generations on31descendants
  • K110911Renaissance SystemMazor Robotics, Ltd. · 2011 · 3 generations on24descendants
  • K113228Renaissance SystemMazor Robotics, Ltd. · 2011 · 4 generations on23descendants
  • K140167Renaissance X SystemMazor Robotics, Ltd. · 2014 · 5 generations on21descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K003266Siremobil Iso-C 3DSiemens Medical Solutions USA, Inc.2000—
−2K032280Siremobil Iso-C 3DSiemens AG2003—
−1K050996O-Arm Imaging SystemBreakaway Imaging, LLC20054
0 (this device)K060344O-Arm Imaging SystemBreakaway Imaging, LLC2006—
+1K081672C-InsightMazor Surgical Technologies, Ltd.2008—
+2K102130Tenzing SystemMazor Surgical Technologies, Ltd.2010—
+3K110911Renaissance SystemMazor Robotics, Ltd.2011—
+4K113228Renaissance SystemMazor Robotics, Ltd.2011—
+5K140167Renaissance X SystemMazor Robotics, Ltd.2014—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200121

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200810
201020
201130
201210
201410
201520
201730
201850
202040
202131
202210
202340
202470
202560
202680
FIG. 1 — Lineage of K060344, 2000–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OXO device

  1. #1 K253296 Digital C-ARM with 3Dscore 15.1 · 13 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K050996O-Arm Imaging System
Summary page 1
…bstantial Equivalence The O-arm™ Imaging System with the 1-D Display Control Mouse is substantially equivalent to the O-arm™ Imaging System, which was cleared in 510(k) K050996. The display control functions provided by the optional sterile disposable wireless 1-D Control Mouse are also provided by conso…
Breakaway Imaging, LLC20054

Also cleared against K050996: 2 other devices →

What descends from it? (51)

See all 51

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K081672C-InsightMazor Surgical Technologies, Ltd.200833
K092564O-Arm Imaging SystemMedtronic Navigation, Inc. (Littleton)201016

Has anything in its family been recalled? (2 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 10, 2010Not classifiedZ-1998-2010Device DesignThe possibility that the real time image display can freeze up during use of a 2D continuous fluoroscopy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K003837Flexiview 8800 Digital Mobile Imaging SystemGe Oec Medical Systems · 2001 · 3 generations back · ◆ in the 2001 ancestors bar of FIG. 11 Class I recall
  • K211616ExcelsiusHubGlobus Medical, Inc. · 2021 · 3 generations on · ◆ in the 2021 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 6 years and passes through 3 companies; a new generation was cleared every 1.7 years on median.
  • 3 of the 5 ancestors have a recall of any class; 1 has a Class I recall.
  • Of its descendants, K081672 (Mazor Surgical Technologies, Ltd., 2008) has gone furthest, with 33 descendants of its own.
  • 2 other devices also cite its predicate K050996.

How does its category compare?

All OXO devices

OXO Image-Intensified Fluoroscopic X-Ray System, Mobile: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OXO

4in 2025

2010: 12012: 42013: 22014: 22015: 12016: 32017: 42018: 12019: 32021: 22024: 22025: 42026 (partial year): 2
2007high 4 (2012) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OXO

97in 2025

2012: 48 days median2016: 128 days median2017: 57 days median2019: 213 days median2025: 97 days median
2007high 213 (2019) · low 48 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OXO

4in 2024

2012: 42013: 32014: 42016: 32017: 42018: 112024: 42026 (partial year): 1
2007high 11 (2018) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OXO

118in 2025

2012: 52013: 412014: 892015: 972016: 3,8202017: 1,6972018: 1,6472019: 1,2972020: 8162021: 5542022: 2952023: 2752024: 1922025: 118
2007high 3,820 (2016) · low 0 (2007)2025

MAUDE reports naming product code OXO, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—00
20080—00
20090—00
20101—00
20110—00
201244845
20132—341
20142—489
20151—097
2016312833,820
201745741,697
20181—111,647
2019321301,297
20200—0816
20212—0554
20220—0295
20230—0275
20242—4192
20254970118
20262—1—

What is it indicated for?

The O-arm™ Imaging System is designed for 2D fluoroscopic and 3D imaging for intraoperative applications in surgical theaters, particularly for orthopedic applications. The O-arm™ Imaging System is compatible with certain Image Guided Surgical Systems.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K060344: O-Arm Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K060344

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026