Modification to Synchron LX 20 Clinical Chemistry System

K011213510(k) SpecialCleared May 16, 200126-day reviewClass II21 CFR 862.1665Recalls 11

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · JGS Electrode, Ion Specific, Sodium

Cleared on May 16, 2001 after a 26-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 20 later devices descend from it. It has been recalled 11 times.

Decided
May 16, 2001
Review
26 days
Class
Class II
Product code
JGS
Recalls
11
Predicates
0

20 later devices descend from it, 5 citing it directly.

Descendants, by year cleared 2002: 1 descendant cleared2004: 5 descendants cleared2001: 1 descendant cleared2005: 1 descendant cleared2006: 2 descendants cleared2020: 3 descendants cleared2011: 1 descendant cleared2007: 1 descendant cleared2021: 1 descendant cleared2023: 3 descendants cleared2026: 1 descendant cleared200020052010201520202025
  1. This deviceK011213 Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc. · 2001
  2. Descendant · cites it directly · 11 of its ownK042291 Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · Recalled
  3. Descendant · 2 generations on · 5 of its ownK183688 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  4. Descendant · 2 generations on · 4 of its ownK183678 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  5. Descendant · 2 generations on · 4 of its ownK191298 i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020
  6. Descendant · cites it directly · 4 of its ownK023049 Synchron LX I 725 Clinical SystemBeckman Coulter, Inc. · 2002 · Recalled

Showing 5 of 20 descendants. See all descendants

Descendants cleared per year

2001: 12002: 12004: 52005: 12006: 22007: 12011: 12020: 32021: 12023: 32026: 1
2001high 5 (2004) · low 0 (2003)2026

Most influential descendants

Showing 5 of 20 · See all
  • K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc. · 2004 · cites it directly11descendants
  • K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 2 generations on5descendants
  • K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 2 generations on4descendants
  • K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc. · 2020 · 2 generations on4descendants
  • K023049Synchron LX I 725 Clinical SystemBeckman Coulter, Inc. · 2002 · cites it directly4descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K011213Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc.2001—
+1K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc.200434
+2K183688i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+2K183678i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+2K191298i-STAT CHEM8+ cartridge with the i-STAT 1 SystemAbbott Point of Care, Inc.2020—
+1K023049Synchron LX I 725 Clinical SystemBeckman Coulter, Inc.200218

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200110
200210
200450
200510
200620
200710
201110
202030
202110
202330
202610
FIG. 1 — Lineage of K011213, 2001–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (20)

See all 20

5 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K042291Unicel DXC 600 and 800 Synchron SystemsBeckman Coulter, Inc.200411
K023049Synchron LX I 725 Clinical SystemBeckman Coulter, Inc.20024
K014034Modification to Synchron LX Clinical Chemistry Systems (LX20 and LX20 Pro)Beckman Coulter, Inc.20011
K051108Piccolo Lactate Dephydrogenase Test SystemAbaxis, Inc.20050
K040174Bilirubin Assay on the Omni S AnalyzerRoche Diagnostics Corp.20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 5, 2011Not classifiedZ-2131-2011Equipment maintenanceThe recall was initiated because Beckman Coulter has received reports of incorrect electrolyte results on Synchron LX systems due to a variety of maintenance-related hardware issues.
Oct 22, 2010Not classifiedZ-2072-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Aug 10, 2010Not classifiedZ-1619-2011OtherThe recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be…
Apr 15, 2010Not classifiedZ-2388-2010OtherUniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
Apr 5, 2007Not classifiedZ-1095-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Mar 15, 2006Not classifiedZ-0900-06OtherIt was confirmed that the ORDAC (Over Range Detection And Correction) feature (auto & manual) does not perform correctly for the glucose cup chemistry (GLUCm) when sample type of 'Other' is selected w/operating software version 4.5.…
Sep 6, 2005Not classifiedZ-0009-06Software designSynchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results…
Oct 10, 2003Not classifiedZ-1098-04OtherNumeric results are erroneously accepted by the in-house information system without qualifying symbols "<" or ">". In addition, the in-house system may interpret the results incorrectly.
Oct 7, 2003Not classifiedZ-1116-04Other'Initial Rate High' parameter for User Defined Reagents does not perform as indicated in the Synchron LX Clinical Systems Operations Manual.
Sep 18, 2003Not classifiedZ-0096-04OtherSample wheel may home to the wrong position.
Dec 16, 2002Not classifiedZ-0545-03OtherThe disk that controls the home position of the sample wheel comes loose preventing a proper home sequence.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K042291 (Beckman Coulter, Inc., 2004) has gone furthest, with 11 descendants of its own.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—
Cite this page

510(k) Report. “K011213: Modification to Synchron LX 20 Clinical Chemistry System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K011213

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026