510(k) Report

510(k) · Cleared

Recalled · Class II · 2013

Synchron LX I 725 Clinical System

K023049510(k) SpecialCleared Oct 4, 200222-day reviewClass II21 CFR 862.1665Recalls 18

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · JGS Electrode, Ion Specific, Sodium

Cleared on Oct 4, 2002 as substantially equivalent to K011213 (Beckman Coulter, Inc., 2001) after a 22-day review. Its family tree reaches back 1 generation to 2001. 4 later devices descend from it. It has been recalled 18 times.

Decided
Oct 4, 2002
Review
22 days
Class
Class II
Product code
JGS
Recalls
18
Predicates
1

Its longest line of descent runs back 1 generation to K011213 (2001). 4 later devices descend from it, 3 citing it directly.

Descendants, by year cleared 2004: 2 descendants cleared2006: 1 descendant cleared2011: 1 descendant clearedK011213 · Modification to Synchron LX 20 Clinical Chemistry System · Beckman Coulter, Inc. · 2001 · recalled200020052010201520202025
Descendants, by generation 1 generation forward: 3 descendants2 generations forward: 1 descendant K011213 · Modification to Synchron LX 20 Clinical Chemistry System · Beckman Coulter, Inc. · 2001 · recalled0
  1. Root · 1 generation backK011213 Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc. · 2001 · Recalled
  2. This deviceK023049 Synchron LX I 725 Clinical SystemBeckman Coulter, Inc. · 2002
  3. Descendant · cites it directly · 1 of its ownK060256 Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc. · 2006 · Recalled
  4. Descendant · 2 generations on · 0 of its ownK103842 Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC…Beckman Coulter, Inc. · 2011 · Recalled
  5. Descendant · cites it directly · 0 of its ownK042191 Diazyme Potassium Enzymatic Assay KitDiazyme Laboratories · 2004
  6. Descendant · cites it directly · 0 of its ownK042189 Diazyme Sodium Enzymatic AssayDiazyme Laboratories · 2004

Most influential descendants

Showing 4 of 4 · See all
  • K060256Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc. · 2006 · cites it directly1descendant
  • K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800 Synchron System…Beckman Coulter, Inc. · 2011 · 2 generations on0descendants
  • K042191Diazyme Potassium Enzymatic Assay KitDiazyme Laboratories · 2004 · cites it directly0descendants
  • K042189Diazyme Sodium Enzymatic AssayDiazyme Laboratories · 2004 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K011213Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc.200111
0 (this device)K023049Synchron LX I 725 Clinical SystemBeckman Coulter, Inc.2002—
+1K060256Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc.200624
+2K103842Unicel DXC 600 Synchron System Software Version 5.0 and Unicel DXC 800…Beckman Coulter, Inc.20113
+1K042191Diazyme Potassium Enzymatic Assay KitDiazyme Laboratories2004—
+1K042189Diazyme Sodium Enzymatic AssayDiazyme Laboratories2004—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200420
200610
201110
FIG. 1 — Lineage of K023049, 2001–2011Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JGS? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JGS device

  1. #1 K252475 ABL90 FLEX PLUS Systemscore 12.2 · 4 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K011213Modification to Synchron LX 20 Clinical Chemistry SystemBeckman Coulter, Inc.200111

Also cleared against K011213: 4 other devices →

What descends from it? (4)

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060256Unicel DXC 600I Synchron Access Clinical SystemBeckman Coulter, Inc.20061
K042191Diazyme Potassium Enzymatic Assay KitDiazyme Laboratories20040
K042189Diazyme Sodium Enzymatic AssayDiazyme Laboratories20040

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 12, 2013Class IIZ-1034-2013Device DesignBeckman Coulter is initiating a recall due to a defective newly released aluminum incubator belt pulley that may rub against the light shield, which causes the formation of debris. Debris that is caused by the incubator belt pulley…
Nov 1, 2012Class IIZ-0612-2013Device DesignIt has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at…
Apr 5, 2011Not classifiedZ-2132-2011Equipment maintenanceThe recall was initiated because Beckman Coulter has received reports of incorrect electrolyte results on Synchron LX systems due to a variety of maintenance-related hardware issues.
Oct 22, 2010Not classifiedZ-2070-2011Environmental controlThis recall was initiated because Beckman Coulter has confirmed that for User-Defined Chemistries (UDRs), the polychrome correction is not automatically updated when wavelengths are modified.
Oct 5, 2010Not classifiedZ-1937-2011PendingThe recall was initiated because Beckman Coulter confirmed Access waste bottles that ruptured and expelled liquid waste, exposing the operator to the system liquid waste. Over time, waste bottles can degrade, become brittle and…
Aug 10, 2010Not classifiedZ-1619-2011OtherThe recall was initiated because Cartridge Glucose (GLU) on UniCel DxC and Synchron LX instruments may give falsely high results for moderately hemolysed samples. The Impact: (1) Neonatal samples drawn by lancets are more likely to be…
Apr 15, 2010Not classifiedZ-2388-2010OtherUniCel DxC and Synchron LX instruments produced inaccurate (low) GLUCm results. False low results could affect or delay diagnosis or treatment.
Jan 18, 2010Not classifiedZ-2441-2010OtherThe recall was initiated due to " A loose Aliquot Probe tube fitting in the Closed Tube Aliquotter (CTA), may result in erroneous Access immunoassay results or Synchron chemistry results". This may result in dilution of the sample with…
Sep 11, 2008Not classifiedZ-1374-2012Software designThe recall was initiated because Beckman confirmed a customer report that a timing conflict can occur while aspirating samples for the Access Total BhCG assay.
Apr 3, 2008Not classifiedZ-2385-2008Nonconforming Material/ComponentThe field action was initiated after the firm confirmed reports of premature failure of the waste pump tubing in the Access and Access 2 Immunoassay Systems. Premature failure of the waste pump tubing may negatively affect precision.
Oct 17, 2007Not classifiedZ-1183-2008Process controlElectrical grounding failure: During manufacturing, internal testing found one instrument wlth an intermittent failure of the electrical grounding. The source of the problem was traced to inadequate soldering.
Apr 5, 2007Not classifiedZ-1096-2008Software designIntermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags…
Mar 15, 2006Not classifiedZ-0901-06OtherIt was confirmed that the ORDAC (Over Range Detection And Correction) feature (auto & manual) does not perform correctly for the glucose cup chemistry (GLUCm) when sample type of 'Other' is selected w/operating software version 4.5.…
Sep 6, 2005Not classifiedZ-0009-06Software designSynchron LX Systems Operating Software Version 4.5/v4.6 CBD (Chemistry Database) introduced an adjustment to the urine Uric Acid database calculation scheme which for some patients, depending on the volume of 24 hour urine sample, results…
Oct 10, 2003Not classifiedZ-1098-04OtherNumeric results are erroneously accepted by the in-house information system without qualifying symbols "<" or ">". In addition, the in-house system may interpret the results incorrectly.
Oct 7, 2003Not classifiedZ-1116-04Other'Initial Rate High' parameter for User Defined Reagents does not perform as indicated in the Synchron LX Clinical Systems Operations Manual.
Sep 18, 2003Not classifiedZ-0096-04OtherSample wheel may home to the wrong position.
Jan 30, 2003Not classifiedZ-0693-03OtherFalse results from secondary pour off tubes

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has a recall.
  • Of its descendants, K060256 (Beckman Coulter, Inc., 2006) has gone furthest, with 1 descendant of its own.
  • 4 other devices also cite its predicate K011213.

How does its category compare?

All JGS devices

JGS Electrode, Ion Specific, Sodium: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JGS

3in 2023

2008: 22009: 22010: 12011: 32012: 12013: 22014: 12015: 12016: 42017: 12018: 32019: 32020: 22022: 12023: 32026 (partial year): 1
2007high 4 (2016) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JGS

347in 2023

2011: 191 days median2016: 198 days median2018: 180 days median2019: 58 days median2023: 347 days median
2007high 347 (2023) · low 58 (2019)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JGS

4in 2023

2007: 72008: 12010: 12011: 12012: 22013: 52015: 12019: 52020: 12021: 12023: 4
2007high 7 (2007) · low 0 (2009)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JGS

64in 2025

2007: 332008: 32009: 572010: 3612011: 9122012: 4122013: 5102014: 2512015: 1582016: 842017: 2922018: 1582019: 1052020: 892021: 912022: 602023: 982024: 962025: 64
2007high 912 (2011) · low 3 (2008)2025

MAUDE reports naming product code JGS, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—733
20082—13
20092—057
20101—1361
201131911912
20121—2412
20132—5510
20141—0251
20151—1158
20164198084
20171—0292
201831800158
20193585105
20202—189
20210—191
20221—060
20233347498
20240—096
20250—064
20261—0—

What is it indicated for?

The SYNCHRON LXi 725 System combines the SYNCHRON LX20 PRO analyzer and the Access 2 analyzer into a single instrument presentation. Samples are loaded from a single point of entry through a Closed Tube Aliquotter (CTA) unit. The CTA functions as a sample processing manager by aliquotting and routing samples to the Access 2 and/or LX20 PRO analyzer according to programming requirements. The SYNCHRON LX20 PRO is a fully automated, computer-controlled…

Read the full statement (171 words)

The SYNCHRON LXi 725 System combines the SYNCHRON LX20 PRO analyzer and the Access 2 analyzer into a single instrument presentation. Samples are loaded from a single point of entry through a Closed Tube Aliquotter (CTA) unit. The CTA functions as a sample processing manager by aliquotting and routing samples to the Access 2 and/or LX20 PRO analyzer according to programming requirements.

The SYNCHRON LX20 PRO is a fully automated, computer-controlled clinical chemistry analyzer intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid, (sample type is chemistry dependent).

The Access 2 Immunoassay Analyzer is a microcomputer controlled, random access instrument. The analyzer performs enzyme immunoassays utilizing paramagnetic particle solid phase and chemiluminescent detection. The Access 2 Analyzer is intended for the in vitro determination of a variety of analytes of clinical interest in biological fluids such as serum, plasma, urine, and cerebral spinal fluid, (sample type is chemistry dependent).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023049: Synchron LX I 725 Clinical System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023049

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026