510(k) · Cleared
Recalled · Class II · 2016Ohmeda Medical Spot PT Lite Phototherapy System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ohmeda Medical · LBI Unit, Neonatal Phototherapy
Cleared on Jul 19, 2001 after a 62-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 2 later devices descend from it. It has been recalled once.
- Decided
- Jul 19, 2001
- Review
- 62 days
- Class
- Class II
- Product code
- LBI
- Recalls
- 1
- Predicates
- 0
2 later devices descend from it, 2 citing it directly.
Most influential descendants
Showing 2 of 2 · See allView FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| 0 (this device) | K011549 | Ohmeda Medical Spot PT Lite Phototherapy System | Ohmeda Medical | 2001 | — |
| +1 | K170585 | Skylife | Neolight, LLC | 2017 | — |
| +1 | K120820 | Giraffe Blue Spot PT Lite | Lumitex, Inc. | 2012 | 1 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2012 | 1 | 0 |
| 2017 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in LBI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new LBI device
- #1 K251308 SnugLit (TM) Wearable Phototherapy System…score 11.6 · 4 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against?
FDA has not posted a summary or statement for it, so its predicates are not known.
What descends from it? (2)
2 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Nov 18, 2016 | Class II | Z-1503-2017 | Under Investigation by firm | Certain power cords could overheat. The affected power cords were manufactured by Electric-Cord and are referred to as Taller Bridge power cords. The overheating of the power cord could result in fire, charring, smoke, or sparking where… |
No ancestor or descendant has a Class I recall.
What does this lineage say?
- Of its descendants, K170585 (Neolight, LLC, 2017) has gone furthest, with 0 descendants of its own.
How does its category compare?
All LBI devicesLBI Unit, Neonatal Phototherapy: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · LBI
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · LBI
134in 2016
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · LBI
1in 2019
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · LBI
3in 2025
MAUDE reports naming product code LBI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 91 | 0 | 1 |
| 2008 | 0 | — | 0 | 4 |
| 2009 | 2 | — | 1 | 4 |
| 2010 | 1 | — | 0 | 9 |
| 2011 | 3 | 157 | 0 | 5 |
| 2012 | 5 | 108 | 0 | 3 |
| 2013 | 0 | — | 0 | 6 |
| 2014 | 1 | — | 1 | 5 |
| 2015 | 1 | — | 1 | 6 |
| 2016 | 3 | 134 | 2 | 22 |
| 2017 | 2 | — | 0 | 577 |
| 2018 | 2 | — | 0 | 37 |
| 2019 | 1 | — | 1 | 6 |
| 2020 | 1 | — | 0 | 6 |
| 2021 | 1 | — | 0 | 2 |
| 2022 | 0 | — | 0 | 8 |
| 2023 | 0 | — | 0 | 0 |
| 2024 | 0 | — | 0 | 6 |
| 2025 | 1 | — | 0 | 3 |
| 2026 | 1 | — | 0 | — |
https://510kreport.com/device/K011549 · printed Nov 6, 2026
Cite this page
510(k) Report. “K011549: Ohmeda Medical Spot PT Lite Phototherapy System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K011549
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026