510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Giraffe Omnibed

K101788510(k) SpecialCleared Jul 21, 201026-day reviewClass II21 CFR 880.5400Recalls 8 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMZ Incubator, Neonatal

Cleared on Jul 21, 2010 after a 26-day review. The FDA summary names its predicate without a number. 3 of the 5 devices in its family, counting itself, have a Class I recall. 4 later devices descend from it. It has been recalled 8 times, including 3 Class I recalls.

Decided
Jul 21, 2010
Review
26 days
Class
Class II
Product code
FMZ
Recalls
8 (3 Class I)
Predicates
0

4 later devices descend from it, 1 citing it directly. 3 of the 5 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2016: 1 descendant cleared, 1 with a Class I recall12022: 1 descendant cleared, 1 with a Class I recall12025: 1 descendant cleared2026: 1 descendant cleared2010201520202025
  1. This deviceK101788 Giraffe OmnibedOhmeda Medical · 2010
  2. Descendant · cites it directly · 3 of its ownK152814 Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Recalled (Class I)
  3. Descendant · 2 generations on · 2 of its ownK213553 Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · Recalled (Class I)
  4. Descendant · 3 generations on · 1 of its ownK251663 Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025
  5. Descendant · 4 generations on · 0 of its ownK260326 Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation…Datex-Ohmeda, Inc. · 2026

Class I recalls in this family

Most influential descendants

Showing 4 of 4 · See all
  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · cites it directly3descendants
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on2descendants
  • K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2025 · 3 generations on1descendant
  • K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc. · 2026 · 4 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K101788Giraffe OmnibedOhmeda Medical2010—
+1K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company2016Class I
+2K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc.2022Class I
+3K251663Giraffe OmniBed Carestation (CS1); Giraffe Incubator Carestation (CS1)Datex-Ohmeda, Inc.2025—
+4K260326Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)Datex-Ohmeda, Inc.2026—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201611
202211
202510
202610
FIG. 1 — Lineage of K101788, 2010–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

3 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMZ device

  1. #1 K251663 Giraffe OmniBed Carestation (CS1); Giraffe…score 14.4 · 5 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—Giraffe OmniBed
Named without a submission number
—

What descends from it? (4)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K152814Giraffe OmniBed Carestation CS1 Class IOhmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company20163

Has anything in its family been recalled? (3 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 1, 2026Class IIZ-2475-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
May 1, 2026Class IIZ-2477-2026Component design/selectionGE HealthCare has become aware of a potential issue involving the canopy soffit beneath the control panel end of the device. Under certain conditions, for example, impact during transport, one or more of the fasteners that secure the…
Sep 3, 2024Class IZ-0047-2025Use errorGE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors…
May 19, 2023Class IIZ-2101-2023Nonconforming Material/ComponentGE HealthCare has become aware of the potential that an incorrect type of secondary latch could have been installed on the northeast (NE) and/or northwest (NW) bedside panels for certain Giraffe OmniBed and Giraffe OmniBed Carestation…
Apr 28, 2020Class IIZ-2235-2020Device DesignPotential for certain Giraffe Incubators and Giraffe OmniBeds to deliver a different environmental oxygen level than what is displayed by the device, if a certain optional component referred to as the Servo Oxygen Module is installed.
Apr 2, 2020Class IIZ-2111-2020Device DesignDuring Giraffe Shuttle charging, the batteries emit low concentrations of gases (outgassing) producing a strong sulfur odor.
Oct 11, 2019Class IZ-0538-2020Device DesignGiraffe Incubator, Giraffe OmniBed, Giraffe Incubator Carestation, and Giraffe OmniBed Carestation Bedside panels and portholes can appear closed without being latched. Potential patient fall.
Jan 8, 2013Class IZ-1051-2013Process controlNeonatal beds and incubator display boards may randomly overwrite stored information and radiant heater and oxygen set points upon power up or recovery from a power failure.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K152814Giraffe OmniBed Carestation CS1Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company · 2016 · Cites it directly · ◆ in the 2016 descendants bar of FIG. 13 Class I recalls
  • K213553Giraffe Incubator Carestation CS1Datex Ohmeda, Inc. · 2022 · 2 generations on · ◆ in the 2022 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • Of its descendants, K152814 (Ohmeda Medical, A Division of Datex-Ohmeda,Inc. A GE Company, 2016) has gone furthest, with 3 descendants of its own.

How does its category compare?

All FMZ devices

FMZ Incubator, Neonatal: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · FMZ

2in 2025

2007: 32008: 12010: 42011: 12014: 12015: 12016: 12017: 12018: 22019: 22022: 12025: 22026 (partial year): 2
2007high 4 (2010) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMZ

60in 2010

2007: 81 days median2010: 60 days median
2007high 81 (2007) · low 60 (2010)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMZ

2in 2024

2008: 42011: 52013: 1, 1 Class I2016: 12018: 32019: 12023: 8, 8 Class I2024: 2, 2 Class I2026 (partial year): 3
2007high 8 (2023) · low 0 (2007)2026

Class I (11 since 2007) Class II/III not classified

Adverse event reports per year · FMZ

250in 2025

2007: 142008: 102009: 82010: 112011: 102012: 122013: 112014: 32015: 92016: 122017: 42018: 182019: 242020: 82021: 152022: 272023: 192024: 1182025: 250
2007high 250 (2025) · low 3 (2014)2025

MAUDE reports naming product code FMZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007381014
20081—410
20090—08
2010460011
20111—510
20120—012
20130—1 (1)11
20141—03
20151—09
20161—112
20171—04
20182—318
20192—124
20200—08
20210—015
20221—027
20230—8 (8)19
20240—2 (2)118
20252—0250
20262—3—

What is it indicated for?

The OmnniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. it cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and…

Read the full statement (137 words)

The OmnniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. it cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.

Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.

This device may incorporate a Serv o Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K101788: Giraffe Omnibed.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K101788

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026