510(k) Report

510(k) · Cleared

Recalled · Class I · 2026

Giraffe and Panda Warmers

K123309510(k) TraditionalCleared Feb 5, 2013104-day reviewClass II21 CFR 880.5130Recalls 13 (3 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ohmeda Medical · FMT Warmer, Infant Radiant

Cleared on Feb 5, 2013 as substantially equivalent to K122267 (Ohmeda Medical, A Division of Datex-Ohmeda, Inc., 2012) and 4 other predicates after a 104-day review. Its family tree reaches back 8 generations to 1997. 7 of the 26 devices in its family, counting itself, have a Class I recall. It has been recalled 13 times, including 3 Class I recalls.

Decided
Feb 5, 2013
Review
104 days
Class
Class II
Product code
FMT
Recalls
13 (3 Class I)
Predicates
5

Its longest line of descent runs back 8 generations to K962603 (1997). 25 ancestors in all, the oldest K760102 (1976). No later device cites it as a predicate yet. 7 of the 26 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year cleared 2012: 1 other ancestor, 1 with a Class I recall12010: 1 other ancestor, 1 with a Class I recall12007: 2 other ancestors, 2 with a Class I recall22000: 4 other ancestors1996: 2 other ancestors2003: 1 other ancestor1997: 1 other ancestor1989: 1 other ancestor1986: 1 other ancestor1988: 1 other ancestor1985: 1 other ancestor1976: 1 other ancestorK962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K033298 · SPO2.Com Oximetry Sensors, Multiple Models · Masimo Corp. · 2003×2: 2 generations cleared close together, K070377 (2007), K072157 (2007) K090697 · Modification to: Giraffe and Panda Warmer · Ohmeda Medical · 2009 · Class I recall1980199020002010
  1. Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 4 generations backK033298 SPO2.Com Oximetry Sensors, Multiple ModelsMasimo Corp. · 2003
  6. 3 generations backK070377 Giraffe and Panda WarmerOhmeda Medical · 2007 · Recalled (Class I)
  7. 2 generations backK072157 Modification to Giraffe and Panda Warmers (Giraffe and Panda…Ohmeda Medical · 2007 · Recalled
  8. 1 generation backK090697 Modification to: Giraffe and Panda WarmerOhmeda Medical · 2009 · Recalled (Class I)
  9. This deviceK123309 Giraffe and Panda WarmersOhmeda Medical · 2013

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−7K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−6K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−5K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−4K033298SPO2.Com Oximetry Sensors, Multiple ModelsMasimo Corp.2003—
−3K070377Giraffe and Panda WarmerOhmeda Medical2007Class I
−2K072157Modification to Giraffe and Panda Warmers (Giraffe and Panda Uninterruptible…Ohmeda Medical20071
−1K090697Modification to: Giraffe and Panda WarmerOhmeda Medical2009Class I
0 (this device)K123309Giraffe and Panda WarmersOhmeda Medical2013—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
197610
198510
198610
198810
198910
199620
199710
200040
200310
200722
201011
201211
FIG. 1 — Lineage of K123309, 1997–2013Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

7 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new FMT device

  1. #1 K243606 Babyroo TN300score 9.3 · 3 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (5)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K122267Giraffe and Panda Warmers
Summary page 1
…Infant Warmer Classification Names: Warmer, Infant Radiant Product Code: FMT 880.5130 Predicate Device(s): Giraffe and Panda Warmers; K 122267, K 10 1804, K(090697, K(072157, K070377 Device Description: The Giraffe and Panda Warme…
Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2012Class I
*K101804Giraffe and Panda Warmers
Summary page 1
…mer Classification Names: Warmer, Infant Radiant Product Code: FMT 880.5130 Predicate Device(s): Giraffe and Panda Warmers; K 122267, K 10 1804, K(090697, K(072157, K070377 Device Description: The Giraffe and Panda Warmers are devi…
Ohmeda Medical, A Division of Datex-Ohmeda, Inc.2010Class I
*K090697Modification to: Giraffe and Panda Warmer
Summary page 2
…w configurations provide heat to a patient using the same patient temperature control algorithm previously cleared per 5 10(k) K090697. The Maximum heater output has been reduced to support use of tall bassinets. The pre…
Ohmeda Medical2009Class I
*K072157Modification to Giraffe and Panda Warmers (Giraffe and Panda Uninterruptible Power Supply)
Summary page 1
…ant Product Code: FMT 880.5130 Predicate Device(s): Giraffe and Panda Warmers; K 122267, K 10 1804, K(090697, K(072157, K070377 Device Description: The Giraffe and Panda Warmers are devices with a radiant heating…
Ohmeda Medical20071
*K070377Giraffe and Panda Warmer
Summary page 1
…Product Code: FMT 880.5130 Predicate Device(s): Giraffe and Panda Warmers; K 122267, K 10 1804, K(090697, K(072157, K070377 Device Description: The Giraffe and Panda Warmers are devices with a radiant heating so…
Ohmeda Medical2007Class I

Also cleared against K122267: 1 other device →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (7 Class I)

Recalls of this device

DateClassRecallRoot causeReason
May 8, 2026Class IZ-2359-2026Process controlBlender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or…
Mar 15, 2019Class IZ-1847-2019Device DesignThe bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
Mar 15, 2019Class IZ-1846-2019Device DesignThe bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
Sep 25, 2018Class IIZ-0482-2019Process designPotential electric leakage affecting the user.
Sep 25, 2018Class IIZ-0481-2019Process designPotential electric leakage affecting the user.
Sep 25, 2018Class IIZ-0483-2019Process designPotential electric leakage affecting the user.
Sep 25, 2018Class IIZ-0484-2019Process designPotential electric leakage affecting the user.
Apr 2, 2014Class IIZ-1547-2014Software designMedical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Apr 2, 2014Class IIZ-1550-2014Software designMedical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Apr 2, 2014Class IIZ-1549-2014Software designMedical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Apr 2, 2014Class IIZ-1548-2014Software designMedical device software may be defective and may cause user-set alarm limits to result in false positive or false negative Oxygen saturation and pulse rate alarm notifications to the caregiver.
Dec 12, 2013Class IIZ-0913-2014Process controlMedical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.
Dec 12, 2013Class IIZ-0912-2014Process controlMedical device adjustment knob may be loose and affect the accuracy of blended Air/Oxygen gas mixture to patients.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K122267Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2012 · Direct predicate · ◆ in the 2012 ancestors bar of FIG. 11 Class I recall
  • K101804Giraffe and Panda WarmersOhmeda Medical, A Division of Datex-Ohmeda, Inc. · 2010 · Direct predicate · ◆ in the 2010 ancestors bar of FIG. 11 Class I recall
  • K090697Modification to: Giraffe and Panda WarmerOhmeda Medical · 2009 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K070377Giraffe and Panda WarmerOhmeda Medical · 2007 · Direct predicate · ◆ on the line of descent in FIG. 11 Class I recall
  • K070210Giraffe and Panda T- Piece Resuscitation SystemOhmeda Medical · 2007 · 2 generations back · ◆ in the 2007 ancestors bar of FIG. 13 Class I recalls
  • K070247Giraffe and Panda Bag and Mask Resuscitation SystemOhmeda Medical · 2007 · 2 generations back · ◆ in the 2007 ancestors bar of FIG. 13 Class I recalls

What does this lineage say?

  • The line of descent spans 16 years and passes through 4 companies; a new generation was cleared every 1.6 years on median.
  • 10 of the 25 ancestors have a recall of any class; 6 have a Class I recall.
  • 1 other device also cites its predicate K122267.

How does its category compare?

All FMT devices

FMT Warmer, Infant Radiant: FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · FMT

1in 2025

2007: 22009: 12010: 12011: 12012: 42013: 12018: 12020: 22021: 12022: 12023: 12025: 1
2007high 4 (2012) · low 0 (2008)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · FMT

103in 2012

2012: 103 days median
2007high 103 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · FMT

4in 2019

2008: 12009: 22013: 22014: 12015: 12016: 12018: 42019: 4, 4 Class I2026 (partial year): 2, 2 Class I
2007high 4 (2018) · low 0 (2007)2026

Class I (6 since 2007) Class II/III not classified

Adverse event reports per year · FMT

359in 2025

2007: 212008: 1142009: 802010: 1062011: 792012: 542013: 552014: 792015: 562016: 562017: 462018: 792019: 1,4372020: 1,6052021: 4732022: 1532023: 1752024: 2572025: 359
2007high 1,605 (2020) · low 21 (2007)2025

MAUDE reports naming product code FMT, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20072—021
20080—1114
20091—280
20101—0106
20111—079
20124103054
20131—255
20140—179
20150—156
20160—156
20170—046
20181—479
20190—4 (4)1,437
20202—01,605
20211—0473
20221—0153
20231—0175
20240—0257
20251—0359
20260—2 (2)—

What is it indicated for?

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SPO 2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp0z) and…

Read the full statement (115 words)

Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment. An optional integrated SPO 2 monitoring feature may be used for continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (Sp0z) and pulse rate (measured by an SPOZ sensor). An optional integrated resuscitation system may be used to provide the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K123309: Giraffe and Panda Warmers.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K123309

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026