510(k) Report

510(k) · Cleared

Recalled · Class II · 2020

Axiom Aristos FX (Multi Purpose Radiography System)

K013826510(k) TraditionalCleared Feb 1, 200274-day reviewClass II21 CFR 892.1680Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Systems, Inc. · KPR System, X-Ray, Stationary

Cleared on Feb 1, 2002 as substantially equivalent to K983732 (Siemens Medical Solutions USA, Inc., 1999) and 1 other predicate after a 74-day review. Its family tree reaches back 3 generations to 1981. 81 later devices descend from it. It has been recalled 8 times.

Decided
Feb 1, 2002
Review
74 days
Class
Class II
Product code
KPR
Recalls
8
Predicates
2

Its longest line of descent runs back 3 generations to K803130 (1981). 6 ancestors in all. 81 later devices descend from it, 1 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1995: 1 other ancestor1992: 1 other ancestor1988: 1 other ancestor2006: 1 descendant cleared2015: 5 descendants cleared2008: 4 descendants cleared2007: 1 descendant cleared2010: 4 descendants cleared2013: 3 descendants cleared2009: 1 descendant cleared2011: 3 descendants cleared2018: 5 descendants cleared2022: 5 descendants cleared2019: 6 descendants cleared2014: 4 descendants cleared2012: 3 descendants cleared2017: 2 descendants cleared2024: 4 descendants cleared2016: 5 descendants cleared2020: 7 descendants cleared2025: 5 descendants cleared2023: 4 descendants cleared2021: 5 descendants cleared2026: 4 descendants clearedK803130 · Multix U. Bucky Table X-Ray Unit · Siemens Corp. · 1981 · recalledK971452 · Multix Top/Pro Radiographic X-Ray Table · Siemens Medical Solutions USA, Inc. · 1997 · recalledK983732 · Thorax FD and Multix FD · Siemens Medical Solutions USA, Inc. · 1999 · recalled19801990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 1 other ancestor 1 generation forward: 1 descendant2 generations forward: 4 descendants3 generations forward: 16 descendants4 generations forward: 20 descendants5 generations forward: 21 descendants6 generations forward: 9 descendants7 generations forward: 3 descendants8 generations forward: 2 descendants9 generations forward: 2 descendants10 generations forward: 3 descendants K983732 · Thorax FD and Multix FD · Siemens Medical Solutions USA, Inc. · 1999 · recalledK971452 · Multix Top/Pro Radiographic X-Ray Table · Siemens Medical Solutions USA, Inc. · 1997 · recalledK803130 · Multix U. Bucky Table X-Ray Unit · Siemens Corp. · 1981 · recalled0+5+10
  1. Root · 3 generations backK803130 Multix U. Bucky Table X-Ray UnitSiemens Corp. · 1981 · Recalled
  2. 2 generations backK971452 Multix Top/Pro Radiographic X-Ray TableSiemens Medical Solutions USA, Inc. · 1997 · Recalled
  3. 1 generation backK983732 Thorax FD and Multix FDSiemens Medical Solutions USA, Inc. · 1999 · Recalled
  4. This deviceK013826 Axiom Aristos FX (Multi Purpose Radiography System)Siemens Medical Systems, Inc. · 2002
  5. Descendant · cites it directly · 80 of its ownK061054 Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · Recalled
  6. Descendant · 2 generations on · 78 of its ownK062623 Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · Recalled
  7. Descendant · 2 generations on · 43 of its ownK081722 YSIOSiemens Medical Solutions USA, Inc. · 2008 · Recalled
  8. Descendant · 3 generations on · 32 of its ownK121513 Multix FusionSiemens Medical Solution · 2012 · Recalled
  9. Descendant · 3 generations on · 23 of its ownK133259 Ysio MaxSiemens Medical Solutions USA, Inc. · 2014 · Recalled

Showing 5 of 81 descendants. See all descendants

Descendants cleared per year

2006: 12007: 12008: 42009: 12010: 42011: 32012: 32013: 32014: 42015: 52016: 52017: 22018: 52019: 62020: 72021: 52022: 52023: 42024: 42025: 52026: 4
2006high 7 (2020) · low 1 (2006)2026

Most influential descendants

Showing 5 of 81 · See all
  • K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc. · 2006 · cites it directly80descendants
  • K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc. · 2007 · 2 generations on78descendants
  • K081722YSIOSiemens Medical Solutions USA, Inc. · 2008 · 2 generations on43descendants
  • K121513Multix FusionSiemens Medical Solution · 2012 · 3 generations on32descendants
  • K133259Ysio MaxSiemens Medical Solutions USA, Inc. · 2014 · 3 generations on23descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K803130Multix U. Bucky Table X-Ray UnitSiemens Corp.19811
−2K971452Multix Top/Pro Radiographic X-Ray TableSiemens Medical Solutions USA, Inc.19978
−1K983732Thorax FD and Multix FDSiemens Medical Solutions USA, Inc.19991
0 (this device)K013826Axiom Aristos FX (Multi Purpose Radiography System)Siemens Medical Systems, Inc.2002—
+1K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.20063
+2K062623Axiom Luminos DRFSiemens Medical Solutions USA, Inc.200720
+2K081722YSIOSiemens Medical Solutions USA, Inc.200811
+3K121513Multix FusionSiemens Medical Solution20124
+3K133259Ysio MaxSiemens Medical Solutions USA, Inc.20147

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198810
199210
199510

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
200710
200840
200910
201040
201130
201230
201330
201440
201550
201650
201720
201850
201960
202070
202150
202250
202340
202440
202550
202640
FIG. 1 — Lineage of K013826, 1981–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KPR? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KPR device

  1. #1 K242119 Innovision-Exiiscore 17.0 · 9 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K983732Thorax FD and Multix FD
Summary page 1
…cate device. 4. Substantial Equivalence The MPRS stationary x-ray system is a modified version of the current, commercially available SIEMENS Medical Systems: Thorax FD K983732, cleared on 4/21/99 Multix FD K983732, cleared on 4/21/99 The table is a modified version of the Koordinat M , K951176, cleared…
Siemens Medical Solutions USA, Inc.19991
*K951176Koordinat M
Summary page 1
…ally available SIEMENS Medical Systems: Thorax FD K983732, cleared on 4/21/99 Multix FD K983732, cleared on 4/21/99 The table is a modified version of the Koordinat M , K951176, cleared on 3/16/95 Information that substantiates this claim of equivalence is provided throughout this 510(k) submission and s…
Siemens Medical Solutions USA, Inc.1995—

Also cleared against K983732: 3 other devices →

What descends from it? (81)

See all 81

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K061054Axiom Aristos FX PlusSiemens Medical Solutions USA, Inc.200680

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 6, 2020Class IIZ-1764-2020Radiation Control for Health and Safety ActIt was determined that in rare cases persons can receive low dose of unintended radiation. The system performed exposures without the exposure button being pressed. Exposure will be released for less than 500ms, therefore, the overall…
Apr 14, 2015Class IIZ-1583-2015Device DesignIf the cable cart and the corrugated hose are unable to move freely, the cable holder could drop down and hit patients or system operators.
Feb 17, 2011Not classifiedZ-1959-2011OtherDuring product monitoring, firm discovered a possible hazard to patients when using the detector alpha rotation of all Axiom Aristos FX and FX Plus systems with serial numbers up to and including 1216. When using the detector alpha…
Aug 25, 2008Not classifiedZ-0163-2009Equipment maintenanceTube support arm or the detecting mounting hardware may become loose at the joint with the telescope.
Jun 10, 2005Not classifiedZ-0989-05Otherpotential issue (system may lock up) when repeating patient registrations
Jun 10, 2005Not classifiedZ-0990-05Otherpotential issue (system may lock up) when repeating patient registrations
Jun 10, 2005Not classifiedZ-0988-05Otherpotential issue (system may lock up) when repeating patient registrations
Apr 19, 2003Not classifiedZ-0984-03Othersoftware malfunction. May indicate that the image is flipped when it is not.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 21 years and passes through 3 companies; a new generation was cleared every 2.8 years on median.
  • 4 of the 6 ancestors have a recall of any class.
  • Of its descendants, K061054 (Siemens Medical Solutions USA, Inc., 2006) has gone furthest, with 80 descendants of its own.
  • 3 other devices also cite its predicate K983732.

How does its category compare?

All KPR devices

KPR System, X-Ray, Stationary: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KPR

8in 2025

2007: 122008: 152009: 92010: 92011: 92012: 112013: 52014: 102015: 182016: 112017: 62018: 152019: 122020: 162021: 62022: 82023: 92024: 112025: 82026 (partial year): 6
2007high 18 (2015) · low 5 (2013)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KPR

155in 2025

2007: 31 days median2008: 48 days median2009: 35 days median2010: 126 days median2011: 152 days median2012: 58 days median2013: 163 days median2014: 108 days median2015: 81 days median2016: 84 days median2017: 75 days median2018: 30 days median2019: 33 days median2020: 39 days median2021: 55 days median2022: 43 days median2023: 76 days median2024: 162 days median2025: 155 days median2026 (partial year): 197 days median
2007high 197 (2026) · low 30 (2018)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · KPR

7in 2025

2007: 102008: 102009: 52010: 32011: 52012: 12013: 52014: 82015: 12016: 42017: 32018: 42019: 12020: 12022: 22023: 42024: 12025: 7
2007high 10 (2007) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KPR

39in 2025

2007: 1202008: 952009: 612010: 822011: 1032012: 742013: 242014: 342015: 592016: 182017: 312018: 362019: 772020: 412021: 372022: 272023: 392024: 342025: 39
2007high 120 (2007) · low 18 (2016)2025

MAUDE reports naming product code KPR, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007123110120
200815481095
2009935561
20109126382
201191525103
20121158174
20135163524
201410108834
20151881159
20161184418
2017675331
20181530436
20191233177
20201639141
2021655037
2022843227
2023976439
202411162134
20258155739
202661970—

What is it indicated for?

MPRS is a dedicated x-ray system with a flat panel detector, which allows the acquisition of x-ray exposures without the use of conventional film/screen systems. The MPRS allows radiographic exposures of the whole body including skull, spinal column, chest, abdomen, extremities, excluding mammography. Radiographic exposures may be taken with the patient in the sitting, standing, or supine positions. The intended use and indications for use of this modified device as described in its labeling have not changed from its predicate device.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K013826: Axiom Aristos FX (Multi Purpose Radiography System).” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K013826

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 30, 2026 (extraction version 2d9447fa9ea4)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026