510(k) Report

510(k) · Cleared

Cleared

Admire

K133646510(k) TraditionalCleared Jun 20, 2014205-day reviewClass II21 CFR 892.1750

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Siemens Medical Systems, Inc. · JAK System, X-Ray, Tomography, Computed

Cleared on Jun 20, 2014 as substantially equivalent to K103424 (Siemens Medical Solutions USA, Inc., 2011) after a 205-day review. Its family tree reaches back 6 generations to 2001. 1 later device descends from it.

Decided
Jun 20, 2014
Review
205 days
Class
Class II
Product code
JAK
Recalls
None
Predicates
1

Its longest line of descent runs back 6 generations to K013561 (2001). 11 ancestors in all, the oldest K982349 (1998). 1 later device descends from it, 1 citing it directly.

  1. Root · 6 generations backK013561 Lightspeed 4.0 CT Scanner SystemGE Medical Systems · 2001 · Recalled
  2. 5 generations backK030420 Lightspeed 5.0 Computed Tomography SystemGE Medical Systems · 2003 · Recalled
  3. 4 generations backK040372 Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC · 2004 · Recalled
  4. 3 generations backK052216 Somatom DefinitionSiemens Medical Systems, Inc. · 2005
  5. 2 generations backK082220 Somatom Flash DSSiemens Medical Solutions USA, Inc. · 2008
  6. 1 generation backK103424 SafireSiemens Medical Solutions USA, Inc. · 2011
  7. This deviceK133646 AdmireSiemens Medical Systems, Inc. · 2014
  8. Descendant · cites it directly · 0 of its ownK153360 AdmireSiemens Medi Cal Solutions, Inc. · 2016

Most influential descendants

Showing 1 of 1 · See all
  • K153360AdmireSiemens Medi Cal Solutions, Inc. · 2016 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K013561Lightspeed 4.0 CT Scanner SystemGE Medical Systems20015
−5K030420Lightspeed 5.0 Computed Tomography SystemGE Medical Systems20034
−4K040372Lightspeed 7.0 CT Scanner SystemGe Medical Systems, LLC20045
−3K052216Somatom DefinitionSiemens Medical Systems, Inc.2005—
−2K082220Somatom Flash DSSiemens Medical Solutions USA, Inc.2008—
−1K103424SafireSiemens Medical Solutions USA, Inc.2011—
0 (this device)K133646AdmireSiemens Medical Systems, Inc.2014—
+1K153360AdmireSiemens Medi Cal Solutions, Inc.2016—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199810
200110
200410
200610
200810

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201610
FIG. 1 — Lineage of K133646, 2001–2016Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JAK? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAK device

  1. #1 K230223 iCAC Devicescore 23.5 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K103424Safire
Summary page 2
…Class 11 Product Code: 90JAK Legally Marketed Predicate Devices Trade Name: SAF IRE 510(k)#: K103424 Clearance Date: November 22, 2011 Classification Name: System, X-Ray, Tomography, Computed Classi…
Siemens Medical Solutions USA, Inc.2011—

What descends from it? (1)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K153360AdmireSiemens Medi Cal Solutions, Inc.20160

Has anything in its family been recalled? (0 Class I)

No FDA recall names this device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 13 years and passes through 4 companies; a new generation was cleared every 2.6 years on median.
  • 6 of the 11 ancestors have a recall of any class.
  • Of its descendants, K153360 (Siemens Medi Cal Solutions, Inc., 2016) has gone furthest, with 0 descendants of its own.

How does its category compare?

All JAK devices

JAK System, X-Ray, Tomography, Computed: FDA decisions, review times, recalls and adverse event reports over 20 years. 2014, this device’s year, is highlighted.

FDA decisions per year · JAK

27in 2025

2007: 142008: 262009: 152010: 72011: 162012: 282013: 172014: 302015: 302016: 252017: 292018: 242019: 282020: 302021: 202022: 302023: 272024: 262025: 272026 (partial year): 21
2007high 30 (2014) · low 7 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAK

141in 2025

2007: 27 days median2008: 42 days median2009: 31 days median2010: 92 days median2011: 155 days median2012: 121 days median2013: 132 days median2014: 200 days median2015: 140 days median2016: 129 days median2017: 96 days median2018: 114 days median2019: 94 days median2020: 112 days median2021: 145 days median2022: 131 days median2023: 139 days median2024: 141 days median2025: 141 days median2026 (partial year): 167 days median
2007high 200 (2014) · low 27 (2007)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAK

17in 2025

2007: 112008: 162009: 292010: 512011: 342012: 232013: 112014: 102015: 282016: 442017: 332018: 102019: 52020: 82021: 152022: 92023: 92024: 202025: 172026 (partial year): 23
2007high 51 (2010) · low 5 (2019)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAK

219in 2025

2007: 762008: 1202009: 1052010: 1192011: 2052012: 1922013: 4142014: 3082015: 4542016: 3652017: 3382018: 5432019: 2332020: 1772021: 2612022: 1992023: 3102024: 2122025: 219
2007high 543 (2018) · low 76 (2007)2025

MAUDE reports naming product code JAK, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200714271176
2008264216120
2009153129105
201079251119
20111615534205
20122812123192
20131713211414
20143020010308
20153014028454
20162512944365
2017299633338
20182411410543
201928945233
2020301128177
20212014515261
2022301319199
2023271399310
20242614120212
20252714117219
20262116723—

What is it indicated for?

ADMIRE is a CT reconstruction software. The end user can choose to apply either ADMIRE or the weighted filter back-projection (WFBP) to the acquired raw data. Depending on the clinical task, Patient size, anatomical location, and clinical practice, the use of ADMIRE can help to reduce radiation dose while maintaining pixel noise, low contrast detectability and high contrast resolution. Phantom measurements showed that high contrast resolution and pixel…

Read the full statement (119 words)

ADMIRE is a CT reconstruction software. The end user can choose to apply either ADMIRE or the weighted filter back-projection (WFBP) to the acquired raw data. Depending on the clinical task, Patient size, anatomical location, and clinical practice, the use of ADMIRE can help to reduce radiation dose while maintaining pixel noise, low contrast detectability and high contrast resolution. Phantom measurements showed that high contrast resolution and pixel noise are equivalent between full dose WFBP images and reduced dose ADMIRE images. Additionally, ADMIRE can reduce spiral artifacts by using iterations going back and forth between image space and raw data space.

Images reconstructed with ADMIRE are not intended to be evaluated with syngo Osteo CT or syngo Calcium Scoring.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K133646: Admire.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K133646

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026