510(k) · Cleared
Recalled · Class I · 2019Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Levitronix, LLC · KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
Cleared on Mar 25, 2003 as substantially equivalent to K973011 (Medtronic Bio-Medicus, Inc., 1998) and 1 other predicate after a 421-day review. Its family tree reaches back 2 generations to 1985. 4 of the 27 devices in its family, counting itself, have a Class I recall. 21 later devices descend from it. It has been recalled 3 times, including a Class I recall.
- Decided
- Mar 25, 2003
- Review
- 421 days
- Class
- Class II
- Product code
- KFM
- Recalls
- 3 (1 Class I)
- Predicates
- 2
Its longest line of descent runs back 2 generations to K852698 (1985). 5 ancestors in all. 21 later devices descend from it, 10 citing it directly. 4 of the 27 devices in this family, counting this one, have a Class I recall.
- Root · 2 generations backK852698 Bio Medicus Bio Pump Model BP-80Bio Medicus, Inc. · 1985
- 1 generation backK973011 BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc. · 1998
- This deviceK020271 Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003
- Descendant · cites it directly · 11 of its ownK053630 Levitronix Centrimag Primary ConsoleLevitronix, LLC · 2006
- Descendant · cites it directly · 10 of its ownK081221 Levitronix Centrimag Primary Console, Thoratec Centrimag Primary…Levitronix, LLC · 2008
- Descendant · 2 generations on · 9 of its ownK083340 Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary…Levitronix, LLC · 2008 · Recalled
- Descendant · cites it directly · 4 of its ownK102129 Levitronix 2ND Generation Centrimag Primary Console, Levitronix…Levitronix, LLC · 2010
- Descendant · cites it directly · 2 of its ownK191557 CentriMag Acute Circulatory Support SystemThoratec Corporation (Now Part of Abbott) · 2019
Showing 5 of 21 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 21 · See all- K053630Levitronix Centrimag Primary ConsoleLevitronix, LLC · 2006 · cites it directly11descendants
- K081221Levitronix Centrimag Primary Console, Thoratec Centrimag Primary ConsoleLevitronix, LLC · 2008 · cites it directly10descendants
- K083340Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary ConsoleLevitronix, LLC · 2008 · 2 generations on9descendants
- K102129Levitronix 2ND Generation Centrimag Primary Console, Levitronix MonitorLevitronix, LLC · 2010 · cites it directly4descendants
- K191557CentriMag Acute Circulatory Support SystemThoratec Corporation (Now Part of Abbott) · 2019 · cites it directly2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K852698 | Bio Medicus Bio Pump Model BP-80 | Bio Medicus, Inc. | 1985 | — |
| −1 | K973011 | BPX80 Bio-Pump (BPX80) | Medtronic Bio-Medicus, Inc. | 1998 | — |
| 0 (this device) | K020271 | Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100 | Levitronix, LLC | 2003 | — |
| +1 | K053630 | Levitronix Centrimag Primary Console | Levitronix, LLC | 2006 | — |
| +1 | K081221 | Levitronix Centrimag Primary Console, Thoratec Centrimag Primary Console | Levitronix, LLC | 2008 | — |
| +2 | K083340 | Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary Console | Levitronix, LLC | 2008 | 1 |
| +1 | K102129 | Levitronix 2ND Generation Centrimag Primary Console, Levitronix Monitor | Levitronix, LLC | 2010 | — |
| +1 | K191557 | CentriMag Acute Circulatory Support System | Thoratec Corporation (Now Part of Abbott) | 2019 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
| 1990 | 1 | 0 |
| 1994 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2005 | 1 | 0 |
| 2006 | 1 | 0 |
| 2008 | 3 | 0 |
| 2009 | 2 | 0 |
| 2010 | 2 | 0 |
| 2011 | 2 | 1 |
| 2013 | 2 | 0 |
| 2014 | 1 | 1 |
| 2019 | 2 | 1 |
| 2020 | 1 | 0 |
| 2023 | 2 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new KFM device
- #1 K250326 Quantum Perfusion Centrifugal Blood Pump…score 11.4 · 7 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (2)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K973011 | BPX80 Bio-Pump (BPX80)Summary page 2…Device Name 1 Manufacturer I 510(k) No. Bio-Pump, Model BPX-80 I Medtronic K973011 Device Name I Manufacturer I 51U(K) NO. Bio-Console, Model 550… | 1998 |
| * | K941921 | Bio-Medicus Bio-Console Models 540, 540A,B,C,D,E, & 550Summary page 2…Device Name I Manufacturer I 51U(K) NO. Bio-Console, Model 550 I Medtronic K941921 D. Device Description System - The Levitronix CentriMag@Extracorporeal Blood Pumping System is a fault-t… | 1994 |
What descends from it? (21)
See all 2110 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K053630 | Levitronix Centrimag Primary Console | 2006 | 11 |
| K081221 | Levitronix Centrimag Primary Console, Thoratec Centrimag Primary Console | 2008 | 10 |
| K102129 | Levitronix 2ND Generation Centrimag Primary Console, Levitronix Monitor | 2010 | 4 |
| K191557 | CentriMag Acute Circulatory Support System | 2019 | 2 |
| K110493 | Tandemheart Pump | 2011 | 2 |
| K102726 | Cardiohelp System Class I | 2011 | 2 |
| K090051 | Levitronix Pedimag Blood Pump, Model 201-90052 | 2009 | 2 |
| K051209 | Levitronix Centrimag Back-UP Console | 2005 | 2 |
| K131179 | Centrimag Primary Console (2ND Generation); Mag Monitor | 2013 | 0 |
| K073631 | Orqis Extracorporeal Blood Pumping System, Model BPS-100 | 2008 | 0 |
Has anything in its family been recalled? (4 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Aug 22, 2019 | Class I | Z-0221-2020 | Process change control | Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI). |
| Sep 5, 2018 | Class II | Z-0103-2019 | Device Design | Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection. |
| Mar 17, 2008 | Not classified | Z-1901-2008 | Device Design | Stops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit |
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · Cites it directly · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
- K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 2 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
- K183623LifeSPARC Pump, LifeSPARC ControllerCardiacassist, Inc. · 2019 · 2 generations on · ◆ in the 2019 descendants bar of FIG. 12 Class I recalls
What does this lineage say?
- The line of descent spans 18 years and passes through 3 companies; a new generation was cleared every 12.4 years on median.
- 3 of the 5 ancestors have a recall of any class.
- Of its descendants, K053630 (Levitronix, LLC, 2006) has gone furthest, with 11 descendants of its own.
- 9 other devices also cite its predicate K973011.
How does its category compare?
All KFM devicesKFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · KFM
2in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · KFM
30in 2023
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · KFM
2in 2024
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · KFM
249in 2025
MAUDE reports naming product code KFM, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 0 | — | 0 | 60 |
| 2008 | 4 | 194 | 2 | 62 |
| 2009 | 5 | 184 | 4 | 85 |
| 2010 | 3 | 109 | 2 | 81 |
| 2011 | 7 | 16 | 1 | 57 |
| 2012 | 0 | — | 5 | 82 |
| 2013 | 2 | — | 0 | 104 |
| 2014 | 0 | — | 0 | 347 |
| 2015 | 0 | — | 0 | 299 |
| 2016 | 0 | — | 0 | 156 |
| 2017 | 1 | — | 0 | 121 |
| 2018 | 0 | — | 1 | 176 |
| 2019 | 3 | 147 | 3 (1) | 311 |
| 2020 | 3 | 30 | 0 | 437 |
| 2021 | 3 | 534 | 0 | 446 |
| 2022 | 0 | — | 0 | 294 |
| 2023 | 3 | 30 | 0 | 346 |
| 2024 | 0 | — | 2 | 296 |
| 2025 | 2 | — | 0 | 249 |
| 2026 | 0 | — | 0 | — |
What is it indicated for?
The Levitronix CentriMag@Extracorporeal Blood Pumping System is indicated to pump
blood through the extracorporeal bypass circuit for extracorporeal circulatory support for
periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for
use in extracorporeal circulatory support systems (for periods up to six hours) not
requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support
during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K020271 · printed Nov 6, 2026
Cite this page
510(k) Report. “K020271: Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K020271
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026