510(k) Report

510(k) · Cleared

Cleared

Levitronix Centrimag Primary and Backup Consoles

K093832510(k) SpecialCleared Jan 13, 201030-day reviewClass II21 CFR 870.4380

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Levitronix, LLC · DWA Control, Pump Speed, Cardiopulmonary Bypass

Cleared on Jan 13, 2010 as substantially equivalent to K083340 (Levitronix, LLC, 2008) and 1 other predicate after a 30-day review. Its family tree reaches back 6 generations to 1985. 1 of the 12 devices in its family, counting itself, has a Class I recall.

Decided
Jan 13, 2010
Review
30 days
Class
Class II
Product code
DWA
Recalls
None
Predicates
2

Its longest line of descent runs back 6 generations to K852698 (1985). 11 ancestors in all. No later device cites it as a predicate yet. 1 of the 12 devices in this family, counting this one, has a Class I recall.

  1. Root · 6 generations backK852698 Bio Medicus Bio Pump Model BP-80Bio Medicus, Inc. · 1985
  2. 5 generations backK973011 BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc. · 1998
  3. 4 generations backK020271 Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003 · Recalled (Class I)
  4. 3 generations backK053630 Levitronix Centrimag Primary ConsoleLevitronix, LLC · 2006
  5. 2 generations backK081221 Levitronix Centrimag Primary Console, Thoratec Centrimag Primary…Levitronix, LLC · 2008
  6. 1 generation backK083340 Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary…Levitronix, LLC · 2008 · Recalled
  7. This deviceK093832 Levitronix Centrimag Primary and Backup ConsolesLevitronix, LLC · 2010

Class I recalls in this family

View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−6K852698Bio Medicus Bio Pump Model BP-80Bio Medicus, Inc.1985—
−5K973011BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc.1998—
−4K020271Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC2003Class I
−3K053630Levitronix Centrimag Primary ConsoleLevitronix, LLC2006—
−2K081221Levitronix Centrimag Primary Console, Thoratec Centrimag Primary ConsoleLevitronix, LLC2008—
−1K083340Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary ConsoleLevitronix, LLC20081
0 (this device)K093832Levitronix Centrimag Primary and Backup ConsolesLevitronix, LLC2010—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199010
199410
200510
200910
FIG. 1 — Lineage of K093832, 1985–2010Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DWA device

  1. #1 K221491 Anivia SG1000 Pump Consolescore 9.3 · 6 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K083340Levitronix Centrimag Primary Console; Thoratec Cantrimag Primary Console
Summary page 1
…the Food, Drug 'and Cosmetic Act. C. Predicate Devices: Levitronix CentriMag Primary Console -7 51 0(k) (K083340) Levitronix CentriMag Back-Up Console - 51 0(k) (K090004) Levitronix LLC Special 610(k): Labeling Ch…
Levitronix, LLC20081
*K090004Levitronix, Thoratec Centrimag Back-UP Consoles
Summary page 1
…and Cosmetic Act. C. Predicate Devices: Levitronix CentriMag Primary Console -7 51 0(k) (K083340) Levitronix CentriMag Back-Up Console - 51 0(k) (K090004) Levitronix LLC Special 610(k): Labeling Change Consolidated Centr[Mag* Primary C…
Levitronix, LLC20091

Also cleared against K083340: 2 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (1 Class I)

No FDA recall names this device.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K020271Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 25 years and passes through 3 companies; a new generation was cleared every 2.8 years on median.
  • 7 of the 11 ancestors have a recall of any class; 1 has a Class I recall.
  • 2 other devices also cite its predicate K083340.

How does its category compare?

All DWA devices

DWA Control, Pump Speed, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years. 2010, this device’s year, is highlighted.

FDA decisions per year · DWA

1in 2024

2007: 12008: 32009: 22010: 22011: 22013: 22019: 12020: 22023: 22024: 1
2007high 3 (2008) · low 0 (2012)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DWA

44in 2008

2008: 44 days median
2007high 44 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DWA

1in 2022

2010: 12012: 32013: 12017: 12021: 12022: 1, 1 Class I
2007high 3 (2012) · low 0 (2007)2026

Class I (1 since 2007) Class II/III not classified

Adverse event reports per year · DWA

362in 2025

2007: 102008: 102009: 172010: 1102011: 1472012: 1022013: 1002014: 982015: 1032016: 1092017: 1402018: 2002019: 3762020: 1462021: 1552022: 2052023: 4382024: 3892025: 362
2007high 438 (2023) · low 10 (2007)2025

MAUDE reports naming product code DWA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—010
2008344010
20092—017
20102—1110
20112—0147
20120—3102
20132—1100
20140—098
20150—0103
20160—0109
20170—1140
20180—0200
20191—0376
20202—0146
20210—1155
20220—1 (1)205
20232—0438
20241—0389
20250—0362
20260—0—

What is it indicated for?

The Levitronix CentriMag Primary and Back-Up Consoles are indicated for use with the Levitronix CentriMag Extracorporeal Blood Pumping System, The Levitronix CentriMag Extracorporeal Blood Pumping System is indicated to pump blood through the extracorporeal bypass circuit for extracorporeall circulatory support for periods appropriate to cardiopulmonary bypass (up to six hours). It is also indicated for use in extracorporeal circulatory support systems (for periods up to six hours) not requiring complete cardiopulmonary bypass (e.g., valvuloplasty, circulatory support during mitral valve reoperation, surgery of the vena cava or aorta, liver transplants etc).

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K093832: Levitronix Centrimag Primary and Backup Consoles.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K093832

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026