ClearedDinamap Procare Series Monitor, Models 100N, 200N, 300N, 400N
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Ge Medical Systems Information Technologies · DQA Oximeter
Cleared on Aug 1, 2002 as substantially equivalent to K014255 (Ge Medical Systems Information Technologies, 2002) and 1 other predicate after a 27-day review. Its family tree reaches back 5 generations to 1997. 1 of the 49 devices in its family, counting itself, has a Class I recall. 21 later devices descend from it.
- Decided
- Aug 1, 2002
- Review
- 27 days
- Class
- Class II
- Product code
- DQA
- Recalls
- None
- Predicates
- 3
Its longest line of descent runs back 5 generations to K962603 (1997). 27 ancestors in all, the oldest K896030 (1989). 21 later devices descend from it, 1 citing it directly. 1 of the 49 devices in this family, counting this one, has a Class I recall.
- Root · 5 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
- 4 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
- 3 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
- 2 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
- 1 generation backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
- This deviceK022193 Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies · 2002
- Descendant · cites it directly · 20 of its ownK073203 Carescape V100Ge Medical Systems Information Technologies · 2007
- Descendant · 2 generations on · 11 of its ownK102426 Carescape Model V100 Vital Signs MonitorGe Medical Systems Information Technologies · 2010
- Descendant · 2 generations on · 10 of its ownK120598 Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · Recalled
- Descendant · 3 generations on · 8 of its ownK130584 Monitor B40Ge Medical Systems China Co., Ltd. · 2013 · Recalled
- Descendant · 4 generations on · 7 of its ownK133576 Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · Recalled (Class I)
Showing 5 of 21 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
- K133576 Monitor B404 generations on · 2014
Most influential descendants
Showing 5 of 21 · See all- K073203Carescape V100Ge Medical Systems Information Technologies · 2007 · cites it directly20descendants
- K102426Carescape Model V100 Vital Signs MonitorGe Medical Systems Information Technologies · 2010 · 2 generations on11descendants
- K120598Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · 2 generations on10descendants
- K130584Monitor B40Ge Medical Systems China Co., Ltd. · 2013 · 3 generations on8descendants
- K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · 4 generations on7descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −5 | K962603 | Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables | Masimo Corp. | 1997 | — |
| −4 | K973887 | Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables | Masimo Corp. | 1998 | — |
| −3 | K982255 | Model 2000 Pulse Oximeter | Ivy Biomedical Systems, Inc. | 1998 | — |
| −2 | K991823 | N-395 Pulse Oximeter, Model N-395 | Nellcor Puritan Bennett, Inc. | 1999 | — |
| −1 | K012891 | Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and… | Nellcor Puritan Bennett, Inc. | 2002 | 1 |
| 0 (this device) | K022193 | Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400N | Ge Medical Systems Information Technologies | 2002 | — |
| +1 | K073203 | Carescape V100 | Ge Medical Systems Information Technologies | 2007 | — |
| +2 | K102426 | Carescape Model V100 Vital Signs Monitor | Ge Medical Systems Information Technologies | 2010 | — |
| +2 | K120598 | Procare Monitor B40 | Ge Medical Systems China Co., Ltd. | 2012 | 2 |
| +3 | K130584 | Monitor B40 | Ge Medical Systems China Co., Ltd. | 2013 | 2 |
| +4 | K133576 | Monitor B40 | Ge Medical Systems China Co., Ltd. | 2014 | Class I |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
| 1990 | 2 | 0 |
| 1991 | 4 | 0 |
| 1992 | 1 | 0 |
| 1994 | 1 | 0 |
| 1995 | 3 | 0 |
| 1996 | 3 | 0 |
| 1997 | 2 | 0 |
| 1999 | 3 | 0 |
| 2000 | 1 | 0 |
| 2002 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2007 | 1 | 0 |
| 2010 | 1 | 0 |
| 2012 | 1 | 0 |
| 2013 | 2 | 0 |
| 2014 | 2 | 1 |
| 2015 | 3 | 0 |
| 2016 | 1 | 0 |
| 2017 | 3 | 0 |
| 2018 | 1 | 0 |
| 2019 | 1 | 0 |
| 2020 | 2 | 0 |
| 2022 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DQA device
- #1 K221282 Empatica Health Monitoring Platformscore 20.1 · 38 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary; numbers the summary cites that cannot be its predicate are marked – and kept off the family tree.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K014255 | Dinamap Procare Series Monitor; Model 100, 200, 300, 400 | 2002 |
| * | K012891 | Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables… | 2002 |
| Kept off the family tree | K955846 | FDA’s K955846 is Model 2082 Temp Plus Iii Thermometer/Model 2087 Temp Plus Iii Wall Mount… Our reading of the summary gave K955846, FDA’s number for Model 2082 Temp Plus Iii Thermometer/Model 2087 Temp Plus… (Y). Its FDA name has none of the distinctive words the summary uses for the device it names, “Alaris Medical System”. Kept off the family tree. Summary page 3…DINAMAP ProCare Series KO14255 Pulse Oximetrv I Nellcor N-595 Pulse Oximeter I KO12891 1 Temperature Alaris Medical System K95 5 846 NIBP DINAMAP ProCare Series KO14255… | 1996 |
What descends from it? (21)
See all 211 later device cites it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K073203 | Carescape V100 | 2007 | 20 |
Has anything in its family been recalled? (1 Class I)
No FDA recall names this device.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
- K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
What does this lineage say?
- The line of descent spans 5 years and passes through 4 companies; a new generation was cleared every 10 months on median.
- 4 of the 27 ancestors have a recall of any class.
- Of its descendants, K073203 (Ge Medical Systems Information Technologies, 2007) has gone furthest, with 20 descendants of its own.
How does its category compare?
All DQA devicesDQA Oximeter: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DQA
15in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DQA
182in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DQA
3in 2024
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · DQA
399in 2025
MAUDE reports naming product code DQA, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 20 | 122 | 6 | 325 |
| 2008 | 22 | 97 | 5 | 312 |
| 2009 | 35 | 117 | 29 | 281 |
| 2010 | 29 | 127 | 4 | 185 |
| 2011 | 15 | 195 | 1 | 164 |
| 2012 | 14 | 120 | 0 | 322 |
| 2013 | 15 | 182 | 0 | 468 |
| 2014 | 11 | 175 | 2 | 1,124 |
| 2015 | 13 | 158 | 1 | 2,082 |
| 2016 | 14 | 246 | 0 | 1,775 |
| 2017 | 9 | 200 | 0 | 1,190 |
| 2018 | 18 | 260 | 32 | 694 |
| 2019 | 19 | 158 | 0 | 548 |
| 2020 | 15 | 126 | 0 | 426 |
| 2021 | 10 | 174 | 0 | 553 |
| 2022 | 20 | 265 | 0 | 478 |
| 2023 | 13 | 317 | 0 | 392 |
| 2024 | 14 | 285 | 3 | 545 |
| 2025 | 15 | 182 | 0 | 399 |
| 2026 | 8 | 187 | 0 | — |
What is it indicated for?
The DINAMAPB ProCare Series 100N-400N Monitor is intended to monitor a single adult, pediatric or neonatal patient's vital signs at the bedside or during intra-hospital transport. Vital signs parameters include non-invasive blood pressure (systolic, diastolic, and mean arterial pressure), pulse rate and/or temperature and/or oxygen saturation (pulse oximetry). The portable device is designed for use in numerous clinical settings in various hospital departments such as emergency, radiology, recovery, medical/surgical, labor and delivery, endoscopy, cardiac step-down. It can also be used in satellite areas, physicians' offices, or alternate care settings.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K022193 · printed Nov 4, 2026
Cite this page
510(k) Report. “K022193: Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400N.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K022193
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026