510(k) Report

510(k) · Cleared

Recalled · Class II · 2018

Monitor B40

K130584510(k) TraditionalCleared Jun 26, 2013113-day reviewClass II21 CFR 870.1025Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Medical Systems China Co., Ltd. · MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms)

Cleared on Jun 26, 2013 as substantially equivalent to K120598 (Ge Medical Systems China Co., Ltd., 2012) and 2 other predicates after a 113-day review. Its family tree reaches back 8 generations to 1997. 5 of the 80 devices in its family, counting itself, have a Class I recall. 8 later devices descend from it. It has been recalled 2 times.

Decided
Jun 26, 2013
Review
113 days
Class
Class II
Product code
MHX
Recalls
2
Predicates
3

Its longest line of descent runs back 8 generations to K962603 (1997). 71 ancestors in all, the oldest K896030 (1989). 8 later devices descend from it, 1 citing it directly. 5 of the 80 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 2013: 1 other ancestor, 1 with a Class I recall12010: 1 other ancestor2009: 2 other ancestors2006: 3 other ancestors2005: 2 other ancestors2004: 5 other ancestors, 2 with a Class I recall22003: 4 other ancestors2007: 1 other ancestor2002: 8 other ancestors, 1 with a Class I recall12000: 8 other ancestors1994: 3 other ancestors2001: 3 other ancestors1999: 3 other ancestors1998: 1 other ancestor1996: 4 other ancestors1995: 3 other ancestors1990: 2 other ancestors1997: 2 other ancestors1993: 1 other ancestor1992: 1 other ancestor1991: 4 other ancestors1989: 1 other ancestor2014: 1 descendant cleared, 1 with a Class I recall12015: 2 descendants cleared2020: 1 descendant cleared2017: 1 descendant cleared2022: 1 descendant cleared2024: 1 descendant cleared2025: 1 descendant clearedReference device: K061907 · Modification to: Datex-Ohmeda S/5 Entropy Module, E-Entropy and Accessories · GE Healthcare · 2008K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999×2: 2 generations cleared close together, K012891 (2002), K022193 (2002) K073203 · Carescape V100 · Ge Medical Systems Information Technologies · 2007K120598 · Procare Monitor B40 · Ge Medical Systems China Co., Ltd. · 2012 · recalled1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 2 other ancestors, 1 with a Class I recall12 generations back: 7 other ancestors, 1 with a Class I recall13 generations back: 12 other ancestors, 1 with a Class I recall14 generations back: 10 other ancestors, 1 with a Class I recall15 generations back: 14 other ancestors6 generations back: 9 other ancestors7 generations back: 9 other ancestors 1 generation forward: 1 descendant, 1 with a Class I recall12 generations forward: 2 descendants3 generations forward: 2 descendants4 generations forward: 1 descendant5 generations forward: 2 descendants K120598 · Procare Monitor B40 · Ge Medical Systems China Co., Ltd. · 2012 · recalledK073203 · Carescape V100 · Ge Medical Systems Information Technologies · 2007K022193 · Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400N · Ge Medical Systems Information Technologies · 2002K012891 · Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and Cables (Aka Accessories) · Nellcor Puritan Bennett, Inc. · 2002 · recalledK991823 · N-395 Pulse Oximeter, Model N-395 · Nellcor Puritan Bennett, Inc. · 1999K982255 · Model 2000 Pulse Oximeter · Ivy Biomedical Systems, Inc. · 1998K973887 · Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and Cables · Masimo Corp. · 1998K962603 · Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & Cables · Masimo Corp. · 1997−50+5
  1. Root · 8 generations backK962603 Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp. · 1997
  2. 7 generations backK973887 Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and…Masimo Corp. · 1998
  3. 6 generations backK982255 Model 2000 Pulse OximeterIvy Biomedical Systems, Inc. · 1998
  4. 5 generations backK991823 N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc. · 1999
  5. 4 generations backK012891 Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax…Nellcor Puritan Bennett, Inc. · 2002 · Recalled
  6. 3 generations backK022193 Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies · 2002
  7. 2 generations backK073203 Carescape V100Ge Medical Systems Information Technologies · 2007
  8. 1 generation backK120598 Procare Monitor B40Ge Medical Systems China Co., Ltd. · 2012 · Recalled
  9. This deviceK130584 Monitor B40Ge Medical Systems China Co., Ltd. · 2013
  10. Descendant · cites it directly · 7 of its ownK133576 Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · Recalled (Class I)
  11. Descendant · 2 generations on · 5 of its ownK151063 Monitor B40Ge Medical Systems China Co., Ltd. · 2015 · Recalled
  12. Descendant · 3 generations on · 4 of its ownK171580 Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd. · 2017 · Recalled
  13. Descendant · 2 generations on · 4 of its ownK143676 B40i Patient MonitorGE Healthcare Finland Oy · 2015 · Recalled
  14. Descendant · 3 generations on · 3 of its ownK201941 Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc. · 2020

Showing 5 of 8 descendants. See all descendants

Descendants cleared per year

2014: 12015: 22017: 12020: 12022: 12024: 12025: 1
2014high 2 (2015) · low 0 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 8 · See all
  • K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · cites it directly7descendants
  • K151063Monitor B40Ge Medical Systems China Co., Ltd. · 2015 · 2 generations on5descendants
  • K171580Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd. · 2017 · 3 generations on4descendants
  • K143676B40i Patient MonitorGE Healthcare Finland Oy · 2015 · 2 generations on4descendants
  • K201941Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc. · 2020 · 3 generations on3descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−8K962603Masimo Set MS-1 Pulse Oximeter & the Lnop Series of Sensors & CablesMasimo Corp.1997—
−7K973887Masimo Set MS-1P Pulse Oximeter and the Lnop Series of Sensors and CablesMasimo Corp.1998—
−6K982255Model 2000 Pulse OximeterIvy Biomedical Systems, Inc.1998—
−5K991823N-395 Pulse Oximeter, Model N-395Nellcor Puritan Bennett, Inc.1999—
−4K012891Oximax Pulse Oximetry System with N-595 Pulse Oximeter and Oximax Sensors and…Nellcor Puritan Bennett, Inc.20021
−3K022193Dinamap Procare Series Monitor, Models 100N, 200N, 300N, 400NGe Medical Systems Information Technologies2002—
−2K073203Carescape V100Ge Medical Systems Information Technologies2007—
−1K120598Procare Monitor B40Ge Medical Systems China Co., Ltd.20122
0 (this device)K130584Monitor B40Ge Medical Systems China Co., Ltd.2013—
+1K133576Monitor B40Ge Medical Systems China Co., Ltd.2014Class I
+2K151063Monitor B40Ge Medical Systems China Co., Ltd.20151
+3K171580Monitor B125, Monitor B105Ge Medical Systems (China) Co., Ltd.20171
+2K143676B40i Patient MonitorGE Healthcare Finland Oy20151
+3K201941Monitor B125, Monitor B105Ge Medical Systems Information Technologies, Inc.2020—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
198910
199020
199140
199210
199310
199430
199530
199640
199720
199810
199930
200080
200130
200281
200340
200452
200520
200630
200710
200920
201010
201311

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201411
201520
201710
202010
202210
202410
202510
FIG. 1 — Lineage of K130584, 1997–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MHX device

  1. #1 K251702 IntelliVue Patient Monitor MX750; IntelliVue…score 23.0 · 40 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K120598Procare Monitor B40
Summary page 1
…alarm) Primary Product Code: pMHX Secondary Product Codes: - BzQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, , : DSK, GWQ, FLL, NHO; NHP, NHQ " ‘Predicate Device(s):. K120598 Monitor B40 . ' K 102239 CARESCAPE Monitor B650 K123195 CARESCAPETM Respiratory Modules , Device Description: T =…
Ge Medical Systems China Co., Ltd.20122
*K102239Carescape Monitor B650
Summary page 1
…ode: pMHX Secondary Product Codes: - BzQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, , : DSK, GWQ, FLL, NHO; NHP, NHQ " ‘Predicate Device(s):. K120598 Monitor B40 . ' K 102239 CARESCAPE Monitor B650 K123195 CARESCAPETM Respiratory Modules , Device Description: T =% . monitor that is devel…
GE Healthcare Finland Oy20108
*K123195Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and Accessories
Summary page 1
…es: - BzQ, CBR, CBS, CBQ, CCK, CCL, DXN, DQA, DRT, DSB, , : DSK, GWQ, FLL, NHO; NHP, NHQ " ‘Predicate Device(s):. K120598 Monitor B40 . ' K 102239 CARESCAPE Monitor B650 K123195 CARESCAPETM Respiratory Modules , Device Description: T =% . monitor that is developed based on the predicate Mon…
GE Healthcare Finland Oy2013Class I
†K061907Modification to: Datex-Ohmeda S/5 Entropy Module, E-Entropy and AccessoriesGE Healthcare2008—

Also cleared against K120598: 1 other device →

What descends from it? (8)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K133576Monitor B40 Class IGe Medical Systems China Co., Ltd.20147

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Sep 11, 2018Class IIZ-0584-2019Software designPatient monitors may restart due to network overload caused by network configuration.
Oct 11, 2013Class IIZ-0489-2014Software designGE Healthcare has recently become aware of a potential safety issue due to ECG filter settings associated with the B20/B40 V1, B30 and B40 V2 Patient Monitors. The B20, B30 and B40 Patient Monitors set the ECG filter setting incorrectly…

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K123195Carescape Respiratory Modules, E-Sco, E-Scov, E-Scaiov and AccessoriesGE Healthcare Finland Oy · 2013 · Direct predicate · ◆ in the 2013 ancestors bar of FIG. 13 Class I recalls
  • K043276Datex-Ohmeda S/5TM FM with L-FICU04 and L-FICU04A Software and N-Fcrec ModuleDatex-Ohmeda · 2004 · 2 generations back · ◆ in the 2004 ancestors bar of FIG. 11 Class I recall
  • K040831Datex-Ohmeda Trusat Pulse Oximeter and Accessories, Model 3500Datex-Ohmeda, Inc. · 2004 · 3 generations back · ◆ in the 2004 ancestors bar of FIG. 13 Class I recalls
  • K021955Trutrak Plus Enhancements to the Datex-Ohmeda 3800 Series & 3900 Series Oximeters & AccessoriesDatex-Ohmeda, Inc. · 2002 · 4 generations back · ◆ in the 2002 ancestors bar of FIG. 12 Class I recalls
  • K133576Monitor B40Ge Medical Systems China Co., Ltd. · 2014 · Cites it directly · ◆ in the 2014 descendants bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 16 years and passes through 5 companies; a new generation was cleared every 11 months on median.
  • 22 of the 71 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K133576 (Ge Medical Systems China Co., Ltd., 2014) has gone furthest, with 7 descendants of its own.
  • 1 other device also cites its predicate K120598.

How does its category compare?

All MHX devices

MHX Monitor, Physiological, Patient (With Arrhythmia Detection Or Alarms): FDA decisions, review times, recalls and adverse event reports over 20 years. 2013, this device’s year, is highlighted.

FDA decisions per year · MHX

8in 2025

2007: 152008: 162009: 132010: 202011: 122012: 162013: 212014: 132015: 142016: 132017: 142018: 112019: 72020: 112021: 52022: 152023: 82024: 52025: 82026 (partial year): 14
2007high 21 (2013) · low 5 (2021)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MHX

192in 2025

2007: 96 days median2008: 58 days median2009: 57 days median2010: 90 days median2011: 81 days median2012: 88 days median2013: 120 days median2014: 248 days median2015: 203 days median2016: 149 days median2017: 134 days median2018: 203 days median2019: 243 days median2020: 196 days median2021: 195 days median2022: 189 days median2023: 229 days median2024: 181 days median2025: 192 days median2026 (partial year): 205 days median
2007high 248 (2014) · low 57 (2009)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MHX

21so far in 2026, the most since 2007

2007: 32008: 22009: 72010: 132011: 162012: 112013: 142014: 62015: 52016: 72017: 62018: 122019: 62022: 62023: 32024: 32025: 152026 (partial year): 21
2007high 21 (2026) · low 0 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · MHX

3,761in 2025

2007: 1312008: 2322009: 3822010: 9552011: 9492012: 6762013: 1,9972014: 1,6092015: 1,2842016: 1,1052017: 1,1152018: 2,1572019: 1,4622020: 1,7142021: 2,4232022: 1,4702023: 2,2542024: 2,8142025: 3,761
2007high 3,761 (2025) · low 131 (2007)2025

MAUDE reports naming product code MHX, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200715963131
200816582232
200913577382
2010209013955
2011128116949
2012168811676
201321120141,997
20141324861,609
20151420351,284
20161314971,105
20171413461,115
201811203122,157
2019724361,462
20201119601,714
2021519502,423
20221518961,470
2023822932,254
2024518132,814
20258192153,761
20261420521—

What is it indicated for?

The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport. The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner. The Monitor B40 is not intended for use during MRI. The Monitor B40 can be a stand-alone…

Read the full statement (161 words)

The Monitor B40 is a portable multi-parameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.

The Monitor B40 is intended for use under the direct supervision of a licensed health care practitioner.

The Monitor B40 is not intended for use during MRI.

The Monitor B40 can be a stand-alone monitor or interfaced to other devices via a network.

The Monitor B40 monitors and displays: ECG (including ST segment, arrhythmia detection), invasive blood pressure, heart/pulse rate, osciliometric non-invasive blood pressure (systolic, diastolic and mean arterial pressure), functional oxygen saturation (Sp0O2) and pulse rate via continuous monitoring (including monitoring during conditions of clinical patient motion or low perfusion), temperature with a reusable or disposable electronic thermometer for continual monitoring.

Esophageal/Nasopharyngeal/Tympanic/Rectal/Bladder/Axillary/Skin/Airway/Room/Myocardial/Core/Surface temperature, impedance respiration, respiration rate, airway gases (C0O2, 02, N20, anesthetic agents, anesthetic agent identification and resplratory rate) and Entropy.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K130584: Monitor B40.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K130584

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026