Lifepak 600 Automated External Defibrillator

K011144510(k) TraditionalCleared Dec 3, 2001231-day reviewClass III21 CFR 870.5310Recalls 9

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Physio-Control Corp. · MKJ Automated External Defibrillators (Non-Wearable)

Cleared on Dec 3, 2001 after a 231-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 4 of the 29 devices in its family, counting itself, have a Class I recall. 28 later devices descend from it. It has been recalled 9 times.

Decided
Dec 3, 2001
Review
231 days
Class
Class III
Product code
MKJ
Recalls
9
Predicates
0

28 later devices descend from it, 2 citing it directly. 4 of the 29 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2004: 3 descendants cleared2003: 1 descendant cleared2013: 2 descendants cleared, 1 with a Class I recall12009: 1 descendant cleared2010: 1 descendant cleared2008: 3 descendants cleared2005: 1 descendant cleared2006: 3 descendants cleared, 1 with a Class I recall12007: 2 descendants cleared2011: 2 descendants cleared2014: 3 descendants cleared, 1 with a Class I recall12012: 2 descendants cleared, 1 with a Class I recall12015: 2 descendants cleared2024: 1 descendant cleared2025: 1 descendant cleared200020052010201520202025
  1. This deviceK011144 Lifepak 600 Automated External DefibrillatorMedtronic Physio-Control Corp. · 2001
  2. Descendant · cites it directly · 20 of its ownK033275 Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · Recalled
  3. Descendant · 2 generations on · 10 of its ownK040775 Lifepak 12Medtronic Physio-Control Corp. · 2004 · Recalled
  4. Descendant · 2 generations on · 9 of its ownK041459 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2004
  5. Descendant · 2 generations on · 7 of its ownK063510 Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · Recalled
  6. Descendant · cites it directly · 6 of its ownK040438 Powerheart Aed G3ACardiac Science, Inc. · 2004 · Recalled

Showing 5 of 28 descendants. See all descendants

Descendants cleared per year

2003: 12004: 32005: 12006: 32007: 22008: 32009: 12010: 12011: 22012: 22013: 22014: 32015: 22024: 12025: 1
2003high 3 (2004) · low 0 (2016)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 28 · See all
  • K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp. · 2003 · cites it directly20descendants
  • K040775Lifepak 12Medtronic Physio-Control Corp. · 2004 · 2 generations on10descendants
  • K041459Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2004 · 2 generations on9descendants
  • K063510Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc. · 2007 · 2 generations on7descendants
  • K040438Powerheart Aed G3ACardiac Science, Inc. · 2004 · cites it directly6descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K011144Lifepak 600 Automated External DefibrillatorMedtronic Physio-Control Corp.2001—
+1K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.20031
+2K040775Lifepak 12Medtronic Physio-Control Corp.20041
+2K041459Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.2004—
+2K063510Lifepak 12 Defibrillator/MonitorMedtronic Emergency Response Systems, Inc.20072
+1K040438Powerheart Aed G3ACardiac Science, Inc.200425

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200310
200430
200510
200631
200720
200830
200910
201010
201120
201221
201321
201431
201520
202410
202510
FIG. 1 — Lineage of K011144, 2001–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

  • Candidate predicates for a device like this, rankedin the report
  • Recall history across each candidate’s familyin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (28)

See all 28

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K033275Lifepak 12, 20, 500, CR PlusMedtronic Physio-Control Corp.200320
K040438Powerheart Aed G3ACardiac Science, Inc.20046

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Feb 8, 2013Class IIZ-1082-2013Component design/selectionLIFEPAK CR Plus or LIFEPAK Express Automated External Defibrillators (AEDs) are recalled due to the potential to experience an early internal battery depletion issue.
Jul 31, 2009Not classifiedZ-2030-2009Nonconforming Material/ComponentThe K1 relay on the Analog printed circuit board assembly may short due to moisture ingress in a highly humid environment. K1 shorting may affect the ECG amplitude causing the Shock Advisory System(SAS) to not analyze the ECG rhythm…
Aug 28, 2008Not classifiedZ-2341-2008Employee errorThis recall is being conducted due to the device being configured with the incorrect software for semi-automatic instead of fully automatic use. When the device is needed for a cardiac arrest emergency, the device will require that the…
Jul 24, 2008Not classifiedZ-0149-2009Nonconforming Material/ComponentThe device may not power on although it indicates it is ready for use and would not be able to provide defibrillation therapy.
Jul 24, 2008Not classifiedZ-0150-2009Nonconforming Material/ComponentThe device may not power on although it indicates it is ready for use and would not be able to provide defibrillation therapy.
Mar 2, 2007Not classifiedZ-0671-2007Software designLIFEPAK CR Plus defibrillators with software version 1.16 or lower may miscalculate the HLC battery capacity and result in the CHARGE-PAK battery charger icon/indicator to display and falsely indicate battery needs replacement.
Oct 5, 2006Not classifiedZ-0836-2007Nonconforming Material/ComponentPremature battery depletion due to electrical component failure
Oct 5, 2006Not classifiedZ-0837-2007Nonconforming Material/ComponentPremature battery depletion due to electrical component failure
May 23, 2003Not classifiedZ-0940-03OtherDefibrillation therapy may be delayed or not delivered for persons needing defibrillation therapy.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K130454Lifepak 20 Defibrillator/Monitor, Lifepak 20E Defibrillator/MonitorPhysio-Control, Inc. · 2013 · 2 generations on · ◆ in the 2013 descendants bar of FIG. 11 Class I recall
  • K122600Lifepak 1000 DefibrillatorPhysio-Control, Inc. · 2012 · 3 generations on · ◆ in the 2012 descendants bar of FIG. 11 Class I recall
  • K052465Reduced Energy Pediatric Electrode Pads for the Samaritan Aed and Pad, Models Sde 251, SPP 351Heartsine Technologies, Inc. · 2006 · 3 generations on · ◆ in the 2006 descendants bar of FIG. 11 Class I recall
  • K142430LIFEPAK 15 monitor/defibrillatorPhysio-Control, Inc. · 2014 · 3 generations on · ◆ in the 2014 descendants bar of FIG. 13 Class I recalls

What does this lineage say?

  • Of its descendants, K033275 (Medtronic Physio-Control Corp., 2003) has gone furthest, with 20 descendants of its own.

How does its category compare?

All MKJ devices

MKJ Automated External Defibrillators (Non-Wearable): FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MKJ

48in 2025

2007: 72008: 132009: 82010: 42011: 112012: 102013: 72014: 112015: 62017: 42018: 202019: 252020: 522021: 412022: 192023: 242024: 272025: 482026 (partial year): 31
2007high 52 (2020) · low 0 (2016)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MKJ

156in 2015

2007: 69 days median2008: 141 days median2009: 165 days median2010: 250 days median2011: 36 days median2012: 197 days median2013: 191 days median2014: 196 days median2015: 156 days median
2007high 250 (2010) · low 36 (2011)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MKJ

1in 2025

2007: 122008: 242009: 372010: 352011: 72012: 5, 1 Class I2013: 42014: 24, 15 Class I2015: 12016: 2, 1 Class I2017: 6, 4 Class I2018: 32019: 6, 2 Class I2020: 22021: 42022: 32023: 112024: 52025: 12026 (partial year): 3
2007high 37 (2009) · low 1 (2015)2026

Class I (23 since 2007) Class II/III not classified

Adverse event reports per year · MKJ

13,315in 2025

2007: 3,9582008: 4,7522009: 6,7822010: 7,3992011: 8,1702012: 9,7322013: 14,9552014: 15,3992015: 16,1292016: 17,0622017: 16,6842018: 16,5912019: 16,8612020: 16,1972021: 16,3002022: 14,8672023: 13,7662024: 15,1092025: 13,315
2007high 17,062 (2016) · low 3,958 (2007)2025

MAUDE reports naming product code MKJ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
2007769123,958
200813141244,752
20098165376,782
20104250357,399
2011113678,170
2012101975 (1)9,732
20137191414,955
20141119624 (15)15,399
20156156116,129
20160—2 (1)17,062
20174—6 (4)16,684
201820—316,591
201925—6 (2)16,861
202052—216,197
202141—416,300
202219—314,867
202324—1113,766
202427—515,109
202548—113,315
202631—3—
Cite this page

510(k) Report. “K011144: Lifepak 600 Automated External Defibrillator.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K011144

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026