510(k) Report

510(k) · Cleared

Recalled · Class II · 2017

Modification to Radial Head Implant

K023604510(k) SpecialCleared Nov 27, 200230-day reviewClass II21 CFR 888.3170Recalls 18

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Avanta Orthopaedics, Inc. · KWI Prosthesis, Elbow, Hemi-, Radial, Polymer

Cleared on Nov 27, 2002 as substantially equivalent to K982288 (Avanta Orthopaedics, Inc., 1998) and 1 other predicate after a 30-day review. Its family tree reaches back 2 generations to 1995. 19 later devices descend from it. It has been recalled 18 times.

Decided
Nov 27, 2002
Review
30 days
Class
Class II
Product code
KWI
Recalls
18
Predicates
2

Its longest line of descent runs back 2 generations to K944507 (1995). 3 ancestors in all. 19 later devices descend from it, 2 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 2000: 1 other ancestor2013: 1 descendant cleared2006: 1 descendant cleared2010: 1 descendant cleared2016: 1 descendant cleared2017: 1 descendant cleared2020: 3 descendants cleared2018: 1 descendant cleared2023: 2 descendants cleared2024: 4 descendants cleared2021: 2 descendants cleared2025: 1 descendant cleared2022: 1 descendant clearedK944507 · Metallic Radial Head Implants · Wrightmedicaltechnologyinc · 1995K982288 · Radial Head Implant · Avanta Orthopaedics, Inc. · 1998 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor 1 generation forward: 2 descendants2 generations forward: 3 descendants3 generations forward: 5 descendants4 generations forward: 6 descendants5 generations forward: 2 descendants6 generations forward: 1 descendant K982288 · Radial Head Implant · Avanta Orthopaedics, Inc. · 1998 · recalledK944507 · Metallic Radial Head Implants · Wrightmedicaltechnologyinc · 19950+5
  1. Root · 2 generations backK944507 Metallic Radial Head ImplantsWrightmedicaltechnologyinc · 1995
  2. 1 generation backK982288 Radial Head ImplantAvanta Orthopaedics, Inc. · 1998 · Recalled
  3. This deviceK023604 Modification to Radial Head ImplantAvanta Orthopaedics, Inc. · 2002
  4. Descendant · cites it directly · 18 of its ownK060438 RHSTornier · 2006
  5. Descendant · cites it directly · 16 of its ownK130533 Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S. · 2013 · Recalled
  6. Descendant · 2 generations on · 15 of its ownK153452 Salto XT, Salto TalarisTornier S.A.S. · 2016 · Recalled
  7. Descendant · 3 generations on · 8 of its ownK192994 Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc. · 2020
  8. Descendant · 4 generations on · 2 of its ownK210390 Apex 3D Total Ankle Replacement SystemParagon 28, Inc. · 2021

Showing 5 of 19 descendants. See all descendants

Descendants cleared per year

2006: 12010: 12013: 12016: 12017: 12018: 12020: 32021: 22022: 12023: 22024: 42025: 1
2006high 4 (2024) · low 0 (2007)2026

Most influential descendants

Showing 5 of 19 · See all
  • K060438RHSTornier · 2006 · cites it directly18descendants
  • K130533Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S. · 2013 · cites it directly16descendants
  • K153452Salto XT, Salto TalarisTornier S.A.S. · 2016 · 2 generations on15descendants
  • K192994Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc. · 2020 · 3 generations on8descendants
  • K210390Apex 3D Total Ankle Replacement SystemParagon 28, Inc. · 2021 · 4 generations on2descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K944507Metallic Radial Head ImplantsWrightmedicaltechnologyinc1995—
−1K982288Radial Head ImplantAvanta Orthopaedics, Inc.19981
0 (this device)K023604Modification to Radial Head ImplantAvanta Orthopaedics, Inc.2002—
+1K060438RHSTornier2006—
+1K130533Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S.201319
+2K153452Salto XT, Salto TalarisTornier S.A.S.20168
+3K192994Paragon 28 APEX 3D Total Ankle Replacement SystemParagon 28, Inc.2020—
+4K210390Apex 3D Total Ankle Replacement SystemParagon 28, Inc.2021—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
200010

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200610
201010
201310
201610
201710
201810
202030
202120
202210
202320
202440
202510
FIG. 1 — Lineage of K023604, 1995–2025Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in KWI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new KWI device

  1. #1 K260624 Arthrex Radial Head Replacement Systemscore 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K982288Radial Head Implant
Summary page 1
…87 KWI Original Predicate Device: Original 5 1Ok application Avanta Orthopaedics (K982288, K002644). Registration Number: 2030506 Owner Operator Number: 9001389 Devic…
Avanta Orthopaedics, Inc.19981
*K002644Radial Head Implant
Summary page 1
…87 KWI Original Predicate Device: Original 5 1Ok application Avanta Orthopaedics (K982288, K002644). Registration Number: 2030506 Owner Operator Number: 9001389 Device Descrip…
Avanta Orthopaedics, Inc.200021

Also cleared against K982288: 4 other devices →

What descends from it? (19)

See all 19

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K060438RHSTornier200618
K130533Salto Talaris Total Ankle Prosthesis, RHSTornier S.A.S.201316

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Oct 24, 2017Class IIZ-2686-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2715-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2719-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2713-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2714-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2687-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2689-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2691-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2692-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2718-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2717-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2712-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2721-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2716-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2720-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2711-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2690-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.
Oct 24, 2017Class IIZ-2688-2018Under Investigation by firmRevision rate identified as higher than other proximal radial head replacement devices based on literature review. Product may not perform at expected level post-surgery.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 7 years and passes through 2 companies; a new generation was cleared every 4.2 years on median.
  • 2 of the 3 ancestors have a recall of any class.
  • Of its descendants, K060438 (Tornier, 2006) has gone furthest, with 18 descendants of its own.
  • 4 other devices also cite its predicate K982288.

How does its category compare?

All KWI devices

KWI Prosthesis, Elbow, Hemi-, Radial, Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · KWI

1in 2024

2010: 12012: 12013: 12014: 12017: 12019: 12020: 12021: 12022: 12023: 12024: 12026 (partial year): 1
2007high 1 (2010) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · KWI

Too few KWI 510(k)s in any year since 2007 to give a median.

Recalls per year · KWI

2in 2022

2013: 12014: 12017: 12022: 2
2007high 2 (2022) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · KWI

118in 2025

2007: 152008: 162009: 142010: 132011: 102012: 162013: 312014: 292015: 1032016: 1382017: 3602018: 2022019: 2142020: 1182021: 442022: 442023: 772024: 1242025: 118
2007high 360 (2017) · low 10 (2011)2025

MAUDE reports naming product code KWI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—015
20080—016
20090—014
20101—013
20110—010
20121—016
20131—131
20141—129
20150—0103
20160—0138
20171—1360
20180—0202
20191—0214
20201—0118
20211—044
20221—244
20231—077
20241—0124
20250—0118
20261—0—

What is it indicated for?

Avanta Orthopaedics Radial Head implant is intended for replacement of the proximal end of the radius:

Replacement of the radial head for degenerative, or post-traumatic disabilities presenting pain, crepitation and decreased motion at the radiohumeral and or proximal radio-ulnar joint with:

joint destruction or subluxation visible on x-ray

resistance to conservative treatment

Primary replacement after fracture of the radial head

Symptomatic sequelae after radial head resection

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K023604: Modification to Radial Head Implant.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K023604

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026