Oec 9800 E/CV+ Digital Mobile Imaging System

K024012510(k) TraditionalCleared Mar 20, 2003106-day reviewClass II21 CFR 892.1650Recalls 20 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Oec Medical Systems · JAA System, X-Ray, Fluoroscopic, Image-Intensified

Cleared on Mar 20, 2003 after a 106-day review. The FDA summary names its predicate without a number. 1 of the 60 devices in its family, counting itself, has a Class I recall. 59 later devices descend from it. It has been recalled 20 times, including a Class I recall.

Decided
Mar 20, 2003
Review
106 days
Class
Class II
Product code
JAA
Recalls
20 (1 Class I)
Predicates
0

59 later devices descend from it, 4 citing it directly. 1 of the 60 devices in this family, counting this one, has a Class I recall.

Descendants, by year cleared 2006: 2 descendants cleared2004: 2 descendants cleared2010: 1 descendant cleared2008: 1 descendant cleared2009: 3 descendants cleared2012: 2 descendants cleared2014: 3 descendants cleared2011: 1 descendant cleared2013: 3 descendants cleared2024: 4 descendants cleared2020: 5 descendants cleared2021: 5 descendants cleared2015: 2 descendants cleared2018: 2 descendants cleared2025: 7 descendants cleared2017: 5 descendants cleared2019: 3 descendants cleared2023: 3 descendants cleared2026: 3 descendants cleared2016: 1 descendant cleared2022: 1 descendant cleared200020052010201520202025
  1. This deviceK024012 Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems · 2003
  2. Descendant · cites it directly · 40 of its ownK041932 Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · Recalled
  3. Descendant · 2 generations on · 38 of its ownK073543 Oec 9900 EliteGe Oec Medical Systems · 2008 · Recalled
  4. Descendant · 2 generations on · 35 of its ownK082781 Oec 9900 EliteGe Healthcare Surgery · 2009 · Recalled
  5. Descendant · 3 generations on · 18 of its ownK120613 Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · Recalled
  6. Descendant · cites it directly · 18 of its ownK061203 Ziehm Vision R Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006 · Recalled

Showing 5 of 59 descendants. See all descendants

Descendants cleared per year

2004: 22006: 22008: 12009: 32010: 12011: 12012: 22013: 32014: 32015: 22016: 12017: 52018: 22019: 32020: 52021: 52022: 12023: 32024: 42025: 72026: 3
2004high 7 (2025) · low 0 (2005)2026

Most influential descendants

Showing 5 of 59 · See all
  • K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc. · 2004 · cites it directly40descendants
  • K073543Oec 9900 EliteGe Oec Medical Systems · 2008 · 2 generations on38descendants
  • K082781Oec 9900 EliteGe Healthcare Surgery · 2009 · 2 generations on35descendants
  • K120613Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc. · 2012 · 3 generations on18descendants
  • K061203Ziehm Vision R Digital Mobile Imaging SystemZiehm Imaging, Inc. · 2006 · cites it directly18descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K024012Oec 9800 E/CV+ Digital Mobile Imaging SystemGe Oec Medical Systems2003—
+1K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.20047
+2K073543Oec 9900 EliteGe Oec Medical Systems20085
+2K082781Oec 9900 EliteGe Healthcare Surgery20094
+3K120613Oec 9900 EliteGe Healthcare Surgery GE Oec Medical Systems, Inc.20127
+1K061203Ziehm Vision R Digital Mobile Imaging SystemZiehm Imaging, Inc.20064

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200420
200620
200810
200930
201010
201110
201220
201330
201430
201520
201610
201750
201820
201930
202050
202150
202210
202330
202440
202570
202630
FIG. 1 — Lineage of K024012, 2003–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

1 device in this family has a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JAA device

  1. #1 K253103 Nexus DRF Digital X-ray Imaging Systemscore 13.2 · 7 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—GE OEC Medical Systems - OEC 9800 Plus Mobile Digital Imaging System
Named without a submission number
—
*—Siemens Medical Solutions - Powermobil
Named without a submission number
—

What descends from it? (59)

See all 59

4 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041932Oec Olympus Fluoroscopic Imaging SystemGe Oec Medical Systems, Inc.200440
K061203Ziehm Vision R Digital Mobile Imaging SystemZiehm Imaging, Inc.200618
K041931Oec Olympus Mobile Fluoroscopy System with Integrated NavigationGe Oec Medical Systems, Inc.20041
K092713Spectre Wireless Encrypted Footswitch and Hand Switch SystemTransamerican Medical Imaging20100

Has anything in its family been recalled? (1 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 26, 2010Class IIZ-2241-2012Component change controlGE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Nov 6, 2009Class IIZ-2240-2012Device DesignGE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing…
Oct 10, 2008Class IIZ-2262-2012Device DesignGE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
Aug 12, 2008Not classifiedZ-1885-2008Radiation Control for Health and Safety ActBeam limitation may be non-compliant on some X-ray units.
Aug 12, 2008Not classifiedZ-1884-2008Radiation Control for Health and Safety ActBeam limitation may be non-compliant on some X-ray units.
Apr 15, 2008Not classifiedZ-1705-2008Device DesignUse of existing four-pedal footswitch on a different machine may cause various operational errors.
Feb 22, 2008Class IZ-2038-2012Software designPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is…
Nov 19, 2007Class IIZ-2026-2012Device DesignPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of…
Jul 6, 2007Not classifiedZ-0990-2007OtherThe Air Kerma Rate displayed value and cumulative air kerma displayed value could exceed the allowable error under certain imaging conditions.
Jun 12, 2007Not classifiedZ-0849-2007OtherX-ray systems were not configured to display Air Kerma Rate and Cumulative Air Kerma.
Mar 1, 2007Not classifiedZ-0681-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0680-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0677-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0682-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Feb 9, 2007Not classifiedZ-0557-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0556-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0558-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0559-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Aug 16, 2006Not classifiedZ-1302-06OtherX-ray systems could provide output which exceeds the 20 R/minute limit.
May 31, 2006Not classifiedZ-0001-2007Process controlWeld defect- the weld between the post body and the screw was not properly formed and can fail, even when minimal force is used.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Of its descendants, K041932 (Ge Oec Medical Systems, Inc., 2004) has gone furthest, with 40 descendants of its own.

How does its category compare?

All JAA devices

JAA System, X-Ray, Fluoroscopic, Image-Intensified: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JAA

2in 2025

2007: 12008: 62010: 12011: 42012: 12013: 62014: 62015: 32016: 22017: 12018: 32019: 62020: 42021: 62022: 12023: 22024: 32025: 22026 (partial year): 1
2007high 6 (2008) · low 0 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JAA

89in 2024

2008: 18 days median2011: 48 days median2013: 129 days median2014: 276 days median2015: 197 days median2018: 30 days median2019: 49 days median2020: 39 days median2021: 104 days median2024: 89 days median
2007high 276 (2014) · low 18 (2008)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JAA

4in 2025

2007: 352008: 212009: 182010: 62011: 72012: 22013: 32014: 12015: 12016: 72017: 32018: 12019: 22020: 22021: 22022: 52023: 92024: 22025: 4
2007high 35 (2007) · low 1 (2014)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JAA

33in 2025

2007: 9,5172008: 9,4842009: 12,0402010: 5,8032011: 14,0002012: 11,6842013: 13,7062014: 10,2652015: 2,9222016: 352017: 562018: 362019: 492020: 392021: 352022: 342023: 412024: 252025: 33
2007high 14,000 (2011) · low 25 (2024)2025

MAUDE reports naming product code JAA, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071—359,517
2008618219,484
20090—1812,040
20101—65,803
2011448714,000
20121—211,684
20136129313,706
20146276110,265
2015319712,922
20162—735
20171—356
2018330136
2019649249
2020439239
20216104235
20221—534
20232—941
2024389225
20252—433
20261—0—

What is it indicated for?

The OEC 9800 E/CV+ Digital Mobile Imaging System is designed to provide fluoroscopic and spot-film imaging of the patient during diagnostic, surgical and interventional procedures. It includes features designed specifically for diagnostic and interventional cardiac imaging procedures, and is also intended for use in cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. It may also be used for other imaging applications at the physician’s discretion.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K024012: Oec 9800 E/CV+ Digital Mobile Imaging System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K024012

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026