510(k) Report

510(k) · Cleared

Recalled · Class I · 2016

Oec Fluorotrak 9800 Plus

K022069510(k) TraditionalCleared Sep 17, 200284-day reviewClass II21 CFR 892.1650Recalls 18 (1 Class I)

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Ge Oec Medical Systems · OWB Interventional Fluoroscopic X-Ray System

Cleared on Sep 17, 2002 as substantially equivalent to K974355 (Ge Dec Medical Systems, 1998) and 1 other predicate after an 84-day review. Its family tree reaches back 4 generations to 1994. 5 of the 11 devices in its family, counting itself, have a Class I recall. 2 later devices descend from it. It has been recalled 18 times, including a Class I recall.

Decided
Sep 17, 2002
Review
84 days
Class
Class II
Product code
OWB
Recalls
18 (1 Class I)
Predicates
2

Its longest line of descent runs back 4 generations to K911783 (1994). 8 ancestors in all. 2 later devices descend from it, 1 citing it directly. 5 of the 11 devices in this family, counting this one, have a Class I recall.

Other ancestors, by year clearedDescendants, by year cleared 1998: 1 other ancestor1999: 1 other ancestor1996: 2 other ancestors, 1 with a Class I recall12004: 1 descendant cleared2006: 1 descendant clearedK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 1994×2: 2 generations cleared close together, K981998 (1998), K983529 (1998) K994270 · Instatrak 3000 System with Fluorotrak Module, Model IT3000 · Visualization Technology, Inc. · 2000 · Class I recall1990200020102020
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor3 generations back: 2 other ancestors, 1 with a Class I recall1 1 generation forward: 1 descendant2 generations forward: 1 descendant K994270 · Instatrak 3000 System with Fluorotrak Module, Model IT3000 · Visualization Technology, Inc. · 2000 · Class I recallK983529 · Instatrak 3000 · Visualization Technology, Inc. · 1998 · Class I recallK981998 · Pediatric Instatrak System · Visualization Technology, Inc. · 1998 · Class I recallK911783 · Allegro Viewing Wand · I.S.G. Technologies, Inc. · 19940
  1. Root · 4 generations backK911783 Allegro Viewing WandI.S.G. Technologies, Inc. · 1994
  2. 3 generations backK981998 Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · Recalled (Class I)
  3. 2 generations backK983529 Instatrak 3000Visualization Technology, Inc. · 1998 · Recalled (Class I)
  4. 1 generation backK994270 Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Recalled (Class I)
  5. This deviceK022069 Oec Fluorotrak 9800 PlusGe Oec Medical Systems · 2002
  6. Descendant · cites it directly · 1 of its ownK041931 Oec Olympus Mobile Fluoroscopy System with Integrated NavigationGe Oec Medical Systems, Inc. · 2004 · Recalled
  7. Descendant · 2 generations on · 0 of its ownK061953 9900 Plus Mobile Fluoroscopy System with 3D and Navigation OptionsGe Oec Medical Systems · 2006 · Recalled

Class I recalls in this family

Most influential descendants

Showing 2 of 2 · See all
  • K041931Oec Olympus Mobile Fluoroscopy System with Integrated NavigationGe Oec Medical Systems, Inc. · 2004 · cites it directly1descendant
  • K0619539900 Plus Mobile Fluoroscopy System with 3D and Navigation OptionsGe Oec Medical Systems · 2006 · 2 generations on0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−4K911783Allegro Viewing WandI.S.G. Technologies, Inc.1994—
−3K981998Pediatric Instatrak SystemVisualization Technology, Inc.1998Class I
−2K983529Instatrak 3000Visualization Technology, Inc.1998Class I
−1K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc.2000Class I
0 (this device)K022069Oec Fluorotrak 9800 PlusGe Oec Medical Systems2002—
+1K041931Oec Olympus Mobile Fluoroscopy System with Integrated NavigationGe Oec Medical Systems, Inc.20045
+2K0619539900 Plus Mobile Fluoroscopy System with 3D and Navigation OptionsGe Oec Medical Systems20061

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199621
199810
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200410
200610
FIG. 1 — Lineage of K022069, 1994–2006Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

5 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new OWB device

  1. #1 K251602 Alphenix, INFX-8000V/B, INFX-8000V/S, V9.6…score 18.3 · 18 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (2)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K974355Interventional Mobile Digital Imaging System (Herein Called Imdis)
Summary page 1
…pic Imaging System with Interactive Image Guided Surgical System The OEC FluoroTrak 9800 Plus is substantially equivalent to the OEC 9800 Mobile Digital Imaging System (K974355) marketed by GE OEC Medical Systems, Inc. and the InstaTrak 3000 System with FluoroTrak Module (K994270) marketed by Visualizatio…
Ge Dec Medical Systems19981
*K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000
Summary page 1
…ntially equivalent to the OEC 9800 Mobile Digital Imaging System (K974355) marketed by GE OEC Medical Systems, Inc. and the InstaTrak 3000 System with FluoroTrak Module (K994270) marketed by Visualization Technology, Inc. Page 1-SUM 510(k) Summary OEC FluoroTrak 9800 Plus Indications for…
Visualization Technology, Inc.2000Class I

Also cleared against K974355: 1 other device →

What descends from it? (2)

1 later device cites it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K041931Oec Olympus Mobile Fluoroscopy System with Integrated NavigationGe Oec Medical Systems, Inc.20041

Has anything in its family been recalled? (5 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 22, 2016Class IIZ-2629-2016Software design (manufacturing process)A firmware issue in the supplied LCD workstation monitor(s). Under some circumstances, the monitor(s) may appear dark with no video displayed.
Oct 3, 2014Class IIZ-0195-2015Component design/selectionGE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Apr 26, 2010Class IIZ-2241-2012Component change controlGE OEC recalled certain 9800 Fluoroscope & 9900 Elite Models because the Snubber Board did not conform to the IEC 60601-1 standard for clearance and creepage of voltage isolation.
Nov 6, 2009Class IIZ-2240-2012Device DesignGE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing…
Oct 10, 2008Class IIZ-2262-2012Device DesignGE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
Apr 15, 2008Not classifiedZ-1705-2008Device DesignUse of existing four-pedal footswitch on a different machine may cause various operational errors.
Feb 22, 2008Class IZ-2038-2012Software designPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The vertical lift column power supply in the mainframe C-arm of the device is…
Nov 19, 2007Class IIZ-2026-2012Device DesignPlease be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of…
Mar 1, 2007Not classifiedZ-0681-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0680-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0677-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Mar 1, 2007Not classifiedZ-0682-2007OtherDuring routine service, primary and/or secondary collimators may not have been properly reinstalled.
Feb 9, 2007Not classifiedZ-0557-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0556-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0558-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 9, 2007Not classifiedZ-0559-2007OtherDuring routine service, a cover may have been installed without the required proper lead shielding.
Feb 6, 2006Not classifiedZ-0709-06OtherSmall retaining pin may detach and fall into the surgical field during stereotactic surgery.
Dec 13, 2004Not classifiedZ-0692-5OtherEnhanced sterility testing revealed the sterilization efficacy of accessories ('Sensors') used with certain Navigation and Visualization systems may be compromised.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K994270Instatrak 3000 System with Fluorotrak Module, Model IT3000Visualization Technology, Inc. · 2000 · Direct predicate · ◆ on the line of descent in FIG. 12 Class I recalls
  • K983529Instatrak 3000Visualization Technology, Inc. · 1998 · 2 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K981998Pediatric Instatrak SystemVisualization Technology, Inc. · 1998 · 3 generations back · ◆ on the line of descent in FIG. 11 Class I recall
  • K960330InstatrakVisualization Technology, Inc. · 1996 · 3 generations back · ◆ in the 1996 ancestors bar of FIG. 11 Class I recall

What does this lineage say?

  • The line of descent spans 8 years and passes through 3 companies; a new generation was cleared every 2.5 years on median.
  • 7 of the 8 ancestors have a recall of any class; 4 have a Class I recall.
  • Of its descendants, K041931 (Ge Oec Medical Systems, Inc., 2004) has gone furthest, with 1 descendant of its own.
  • 1 other device also cites its predicate K974355.

How does its category compare?

All OWB devices

OWB Interventional Fluoroscopic X-Ray System: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · OWB

22in 2025

2007: 32008: 42009: 92010: 72011: 62012: 62013: 112014: 152015: 112016: 162017: 202018: 232019: 92020: 162021: 242022: 212023: 122024: 132025: 222026 (partial year): 15
2007high 24 (2021) · low 3 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · OWB

148in 2025

2007: 52 days median2008: 135 days median2009: 99 days median2010: 66 days median2011: 79 days median2012: 37 days median2013: 119 days median2014: 158 days median2015: 66 days median2016: 103 days median2017: 76 days median2018: 54 days median2019: 174 days median2020: 53 days median2021: 60 days median2022: 90 days median2023: 152 days median2024: 126 days median2025: 148 days median2026 (partial year): 221 days median
2007high 221 (2026) · low 37 (2012)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · OWB

42so far in 2026, the most since 2007

2008: 12013: 102014: 92015: 352016: 132017: 132018: 62019: 72020: 82021: 312022: 242023: 222024: 112025: 302026 (partial year): 42
2007high 42 (2026) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · OWB

14,008in 2025

2007: 12011: 12012: 12013: 272014: 1032015: 2,1342016: 4,0742017: 1,6042018: 1,4492019: 1,5582020: 1,5812021: 1,9812022: 2,9322023: 7,6042024: 7,9422025: 14,008
2007high 14,008 (2025) · low 0 (2008)2025

MAUDE reports naming product code OWB, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200735201
2008413510
200999900
201076600
201167901
201263701
2013111191027
2014151589103
20151166352,134
201616103134,074
20172076131,604
2018235461,449
2019917471,558
2020165381,581
20212460311,981
20222190242,932
202312152227,604
202413126117,942
2025221483014,008
20261522142—

What is it indicated for?

The OEC FluoroTrak 9800 Plus Mobile Imaging System provides the physician with fluoroscopic imaging during diagnostic, surgical and interventional procedures. The Fluorotrak surgical navigation feature is intended as an aid to the surgeon for locating anatomical structures anywhere on the human body during either open or percutaneous procedures. It is indicated for any medical condition that may benefit from the use of stereotactic surgery and which provides a reference to rigid anatomical structures such as sinus, skull, long bone or vertebra visible on fluoroscopic images.

Indications for use read by OCR from a scanned FDA summary, so letters may be misread. Check the PDF before quoting it.

Cite this page

510(k) Report. “K022069: Oec Fluorotrak 9800 Plus.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K022069

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026