510(k) · Cleared
Recalled · Class II · 2025Boris Platform Diagnostic Ultrasound System
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Philips Ultrasound, Inc. · IYN System, Imaging, Pulsed Doppler, Ultrasonic
Cleared on Mar 13, 2003 as substantially equivalent to K961459 (Advanced Technology Laboratories, Inc., 1997) and 2 other predicates after a 30-day review. Its family tree reaches back 1 generation to 1997. 2 of the 539 devices in its family, counting itself, have a Class I recall. 535 later devices descend from it. It has been recalled 31 times.
- Decided
- Mar 13, 2003
- Review
- 30 days
- Class
- Class II
- Product code
- IYN
- Recalls
- 31
- Predicates
- 3
Its longest line of descent runs back 1 generation to K961459 (1997). 3 ancestors in all. 535 later devices descend from it, 6 citing it directly. 2 of the 539 devices in this family, counting this one, have a Class I recall.
- Root · 1 generation backK961459 Level 10 HdiAdvanced Technology Laboratories, Inc. · 1997
- This deviceK030455 Boris Platform Diagnostic Ultrasound SystemPhilips Ultrasound, Inc. · 2003
- Descendant · cites it directly · 528 of its ownK091970 Aixplorer Ultrasound SystemSupersonic Imagine · 2009
- Descendant · 2 generations on · 527 of its ownK102041 AixplorerSupersonic Imagine · 2010
- Descendant · 2 generations on · 525 of its ownK112255 AixplorerSupersonic Imagine · 2012
- Descendant · 3 generations on · 405 of its ownK121329 AixplorerSupersonic Imagine · 2012
- Descendant · 3 generations on · 312 of its ownK132274 AixplorerSupersonic Imagine · 2013
Showing 5 of 535 descendants. See all descendants
Descendants cleared per year
Class I recalls in this family
Most influential descendants
Showing 5 of 535 · See all- K091970Aixplorer Ultrasound SystemSupersonic Imagine · 2009 · cites it directly528descendants
- K102041AixplorerSupersonic Imagine · 2010 · 2 generations on527descendants
- K112255AixplorerSupersonic Imagine · 2012 · 2 generations on525descendants
- K121329AixplorerSupersonic Imagine · 2012 · 3 generations on405descendants
- K132274AixplorerSupersonic Imagine · 2013 · 3 generations on312descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −1 | K961459 | Level 10 Hdi | Advanced Technology Laboratories, Inc. | 1997 | — |
| 0 (this device) | K030455 | Boris Platform Diagnostic Ultrasound System | Philips Ultrasound, Inc. | 2003 | — |
| +1 | K091970 | Aixplorer Ultrasound System | Supersonic Imagine | 2009 | — |
| +2 | K102041 | Aixplorer | Supersonic Imagine | 2010 | — |
| +2 | K112255 | Aixplorer | Supersonic Imagine | 2012 | — |
| +3 | K121329 | Aixplorer | Supersonic Imagine | 2012 | — |
| +3 | K132274 | Aixplorer | Supersonic Imagine | 2013 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1999 | 1 | 0 |
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2009 | 3 | 0 |
| 2010 | 1 | 0 |
| 2011 | 1 | 0 |
| 2012 | 3 | 0 |
| 2013 | 10 | 0 |
| 2014 | 16 | 0 |
| 2015 | 30 | 0 |
| 2016 | 36 | 0 |
| 2017 | 49 | 0 |
| 2018 | 41 | 0 |
| 2019 | 32 | 0 |
| 2020 | 47 | 0 |
| 2021 | 45 | 0 |
| 2022 | 44 | 0 |
| 2023 | 38 | 0 |
| 2024 | 45 | 0 |
| 2025 | 54 | 2 |
| 2026 | 40 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
2 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new IYN device
- #1 K240850 EPIQ Series Diagnostic Ultrasound Systems…score 32.1 · 24 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K961459 | Level 10 HdiSummary page 1…eves that the Boris Platform system and transducers are substantially equivalent to the currently marketed Philips HDIB 5000 ultrasound system and transducers cleared in K961459, K991671, and K002003. 3) Device Description The Boris system is a general purpose, mobile, software controlled, diagnostic ult… | 1997 |
| * | K991671 | Hdi 5000 Diagnostic Ultrasound SystemSummary page 1…the Boris Platform system and transducers are substantially equivalent to the currently marketed Philips HDIB 5000 ultrasound system and transducers cleared in K961459, K991671, and K002003. 3) Device Description The Boris system is a general purpose, mobile, software controlled, diagnostic ultrasound s… | 1999 |
| * | K002003 | Hdi 5000 Ultrasound System with Extended Field of View ImagingSummary page 1…atform system and transducers are substantially equivalent to the currently marketed Philips HDIB 5000 ultrasound system and transducers cleared in K961459, K991671, and K002003. 3) Device Description The Boris system is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its fu… | 2000 |
What descends from it? (535)
See all 5356 later devices cite it directly; ranked here by how many devices descend from each in turn.
| Number | Device | Decided | Descendants |
|---|---|---|---|
| K091970 | Aixplorer Ultrasound System | 2009 | 528 |
| K132304 | Epiq Diagnostic Ultrasound System | 2013 | 211 |
| K091804 | CX50 Version 2.0 Diagnostic Ultrasound System and Transducers | 2009 | 57 |
| K123754 | CX50 3.0 Diagnostic Ultrasound System | 2012 | 53 |
| K173090 | Aplio i900/i800/i700/i600 Diagnostic Ultrasound System, V2.4 | 2018 | 21 |
| K083877 | Flying Diagnostic Ultrasound System | 2009 | 0 |
Has anything in its family been recalled? (2 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Sep 5, 2025 | Class III | Z-1625-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1621-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1619-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Sep 5, 2025 | Class III | Z-1620-2026 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2390-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2371-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2368-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2377-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2367-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2379-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2378-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2381-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2402-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2365-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2376-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2382-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2359-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2358-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2357-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2366-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2391-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2353-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| Jul 3, 2025 | Class III | Z-2363-2025 | Device Design | To provide clarification and labelling to define the useful life of ultrasound transducers in the field. |
| May 28, 2025 | Class II | Z-2050-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2056-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2051-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| May 28, 2025 | Class II | Z-2057-2025 | Process control | Ultrasound transducer devices were refurbished beyond their useful life. |
| Aug 30, 2019 | Class II | Z-1844-2020 | Process control | The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of… |
| Aug 30, 2019 | Class II | Z-1842-2020 | Process control | The firm has discovered an issue associated with some Transesophageal (TEE) transducers that were not properly programmed during manufacturing. As a result, the system is unable to correctly measure the temperature on the distal tip of… |
| Mar 27, 2011 | Not classified | Z-2062-2011 | Other | Users of the Philips model S8-3t Micro TEE Transducer have reported that significant image quality degradation may occur after a number of use/cleaning cycles. If users do not inspect the transducer for such damage prior to each use… |
| Nov 2, 2006 | Not classified | Z-0422-2007 | Other | Acoustic output intensity may exceed maximum specified limits and the patient contact temperature may exceed 43 degree C (when the system is used with the S12-4 transducer in CW-mode) |
The 5 latest of 31 recalls; the web page lists all.
Relatives with a Class I recall
Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.
What does this lineage say?
How does its category compare?
All IYN devicesIYN System, Imaging, Pulsed Doppler, Ultrasonic: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · IYN
57in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · IYN
124in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · IYN
10in 2025
Class I (1 since 2007) Class II/III not classified
Adverse event reports per year · IYN
610in 2025
MAUDE reports naming product code IYN, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 13 | 26 | 16 | 3 |
| 2008 | 25 | 18 | 12 | 2 |
| 2009 | 18 | 29 | 8 | 9 |
| 2010 | 29 | 67 | 10 | 17 |
| 2011 | 39 | 46 | 8 | 18 |
| 2012 | 43 | 37 | 3 | 20 |
| 2013 | 51 | 60 | 8 | 18 |
| 2014 | 56 | 39 | 2 | 48 |
| 2015 | 54 | 55 | 3 | 116 |
| 2016 | 60 | 56 | 2 | 145 |
| 2017 | 64 | 83 | 0 | 146 |
| 2018 | 59 | 85 | 7 | 85 |
| 2019 | 47 | 79 | 0 | 59 |
| 2020 | 57 | 72 | 4 | 67 |
| 2021 | 46 | 105 | 8 | 213 |
| 2022 | 53 | 105 | 10 | 186 |
| 2023 | 44 | 115 | 4 | 229 |
| 2024 | 50 | 119 | 1 | 266 |
| 2025 | 57 | 124 | 10 | 610 |
| 2026 | 44 | 107 | 2 (1) | — |
What is it indicated for?
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K030455 · printed Nov 6, 2026
Cite this page
510(k) Report. “K030455: Boris Platform Diagnostic Ultrasound System.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K030455