510(k) · Cleared
Recalled · 2009Oxinium Femoral Heads
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Smith & Nephew, Inc. · JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Cleared on Jul 25, 2003 after a 136-day review. The FDA summary names its predicate without a number. It has been recalled 2 times.
- Decided
- Jul 25, 2003
- Review
- 136 days
- Class
- Class II
- Product code
- JDI
- Recalls
- 2
- Predicates
- 0
The FDA summary names its predicate without a number.
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in JDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new JDI device
- #1 K253515 Zimmer Biomet OSS™ Orthopedic Salvage System…score 15.2 · 12 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (1)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | — | femoral head components distributed by Smith & Nephew, Inc. Named without a submission number |
What descends from it?
No later submission names this one as a predicate yet.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jan 23, 2009 | Not classified | Z-0389-2010 | Other | Radius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The… |
| Jan 12, 2009 | Not classified | Z-2314-2009 | Process control | 46 mm Femoral head was mismarked and packaged as being 45 mm |
No ancestor or descendant has a Class I recall.
How does its category compare?
All JDI devicesJDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · JDI
1in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · JDI
59in 2024
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · JDI
1in 2025
Class I (0 since 2007) Class II/III not classified
Adverse event reports per year · JDI
1,644in 2025
MAUDE reports naming product code JDI, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 3 | 81 | 26 | 696 |
| 2008 | 5 | 96 | 6 | 810 |
| 2009 | 3 | 65 | 3 | 964 |
| 2010 | 0 | — | 3 | 2,040 |
| 2011 | 2 | — | 5 | 3,160 |
| 2012 | 4 | 167 | 4 | 5,122 |
| 2013 | 3 | 176 | 11 | 5,206 |
| 2014 | 3 | 98 | 3 | 4,968 |
| 2015 | 3 | 190 | 1 | 4,906 |
| 2016 | 4 | 166 | 5 | 3,995 |
| 2017 | 1 | — | 5 | 6,135 |
| 2018 | 1 | — | 1 | 6,728 |
| 2019 | 2 | — | 1 | 7,407 |
| 2020 | 7 | 247 | 77 | 4,015 |
| 2021 | 3 | 162 | 0 | 3,170 |
| 2022 | 3 | 69 | 0 | 1,978 |
| 2023 | 2 | — | 0 | 1,635 |
| 2024 | 3 | 59 | 0 | 1,821 |
| 2025 | 1 | — | 1 | 1,644 |
| 2026 | 2 | — | 0 | — |
What is it indicated for?
Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damagsd as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the…
Read the full statement (129 words)
Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damagsd as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Oxinium Femoral Heads are for single use only and can be used as part of cemented or uncemented total hip arthroplasty.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K030766 · printed Nov 5, 2026
Cite this page
510(k) Report. “K030766: Oxinium Femoral Heads.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K030766
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Page last updated Nov 3, 2026