510(k) Report

510(k) · Cleared

Recalled · 2009

Oxinium Femoral Heads

K030766510(k) SpecialCleared Jul 25, 2003136-day reviewClass II21 CFR 888.3350Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Smith & Nephew, Inc. · JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Cleared on Jul 25, 2003 after a 136-day review. The FDA summary names its predicate without a number. It has been recalled 2 times.

Decided
Jul 25, 2003
Review
136 days
Class
Class II
Product code
JDI
Recalls
2
Predicates
0

The FDA summary names its predicate without a number.

FIG. 1 — Lineage of K030766Source: FDA 510(k) database · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JDI? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JDI device

  1. #1 K253515 Zimmer Biomet OSS™ Orthopedic Salvage System…score 15.2 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*—femoral head components distributed by Smith & Nephew, Inc.
Named without a submission number
—

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 23, 2009Not classifiedZ-0389-2010OtherRadius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The…
Jan 12, 2009Not classifiedZ-2314-2009Process control46 mm Femoral head was mismarked and packaged as being 45 mm

No ancestor or descendant has a Class I recall.

How does its category compare?

All JDI devices

JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JDI

1in 2025

2007: 32008: 52009: 32011: 22012: 42013: 32014: 32015: 32016: 42017: 12018: 12019: 22020: 72021: 32022: 32023: 22024: 32025: 12026 (partial year): 2
2007high 7 (2020) · low 0 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JDI

59in 2024

2007: 81 days median2008: 96 days median2009: 65 days median2012: 167 days median2013: 176 days median2014: 98 days median2015: 190 days median2016: 166 days median2020: 247 days median2021: 162 days median2022: 69 days median2024: 59 days median
2007high 247 (2020) · low 59 (2024)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JDI

1in 2025

2007: 262008: 62009: 32010: 32011: 52012: 42013: 112014: 32015: 12016: 52017: 52018: 12019: 12020: 772025: 1
2007high 77 (2020) · low 0 (2021)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JDI

1,644in 2025

2007: 6962008: 8102009: 9642010: 2,0402011: 3,1602012: 5,1222013: 5,2062014: 4,9682015: 4,9062016: 3,9952017: 6,1352018: 6,7282019: 7,4072020: 4,0152021: 3,1702022: 1,9782023: 1,6352024: 1,8212025: 1,644
2007high 7,407 (2019) · low 696 (2007)2025

MAUDE reports naming product code JDI, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
200738126696
20085966810
20093653964
20100—32,040
20112—53,160
2012416745,122
20133176115,206
201439834,968
2015319014,906
2016416653,995
20171—56,135
20181—16,728
20192—17,407
20207247774,015
2021316203,170
202236901,978
20232—01,635
202435901,821
20251—11,644
20262—0—

What is it indicated for?

Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damagsd as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the…

Read the full statement (129 words)

Total hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damagsd as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; old, remote osteomyelitis with an extended drainage-free period; nonunion; femoral neck fracture and trochanteric fractures of the proximal femur with heads involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity. The Oxinium Femoral Heads are for single use only and can be used as part of cemented or uncemented total hip arthroplasty.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K030766: Oxinium Femoral Heads.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K030766

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Page last updated Nov 3, 2026