510(k) · Cleared
Recalled · Class II · 2026Eopa Elongated One-Piece Arterial Cannula and Guidewire
Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly
Medtronic Perfusion Systems · DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Cleared on May 13, 2003 as substantially equivalent to K991066 (Medtronic Vascular, 1999) and 2 other predicates after a 42-day review. Its family tree reaches back 2 generations to 1988. 11 later devices descend from it. It has been recalled 6 times.
- Decided
- May 13, 2003
- Review
- 42 days
- Class
- Class II
- Product code
- DWF
- Recalls
- 6
- Predicates
- 3
Its longest line of descent runs back 2 generations to K880421 (1988). 4 ancestors in all. 11 later devices descend from it, 3 citing it directly.
- Root · 2 generations backK880421 Descending Arch CannulaDlp, Inc. · 1988
- 1 generation backK991066 Madtronic DLP Arterial CannulaeMedtronic Vascular · 1999
- This deviceK031037 Eopa Elongated One-Piece Arterial Cannula and GuidewireMedtronic Perfusion Systems · 2003
- Descendant · cites it directly · 6 of its ownK031518 Eopa Cap Elongated One-Piece Arterial Cannula with Pressure…Medtronic Vascular · 2003
- Descendant · cites it directly · 5 of its ownK111972 Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm…Medtronic, Inc. · 2011
- Descendant · 2 generations on · 4 of its ownK120988 DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous…Medtronic, Inc. · 2012 · Recalled
- Descendant · cites it directly · 3 of its ownK110980 Centrimag Return Cannula KitLevitronix, LLC · 2011
- Descendant · 2 generations on · 2 of its ownK152161 Thoratec CentriMag Return (Arterial) Cannula KitThoratec Corporation · 2015
Showing 5 of 11 descendants. See all descendants
Descendants cleared per year
Most influential descendants
Showing 5 of 11 · See all- K031518Eopa Cap Elongated One-Piece Arterial Cannula with Pressure Monitoring Line, Models…Medtronic Vascular · 2003 · cites it directly6descendants
- K111972Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm Cardioplegia Deliv.,Affinity…Medtronic, Inc. · 2011 · cites it directly5descendants
- K120988DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous Cannula, DLP Single…Medtronic, Inc. · 2012 · 2 generations on4descendants
- K110980Centrimag Return Cannula KitLevitronix, LLC · 2011 · cites it directly3descendants
- K152161Thoratec CentriMag Return (Arterial) Cannula KitThoratec Corporation · 2015 · 2 generations on2descendants
View FIG. 1 as a table
Line of descent and top descendants
| Generation | Number | Device | Applicant | Decided | Recalls |
|---|---|---|---|---|---|
| −2 | K880421 | Descending Arch Cannula | Dlp, Inc. | 1988 | — |
| −1 | K991066 | Madtronic DLP Arterial Cannulae | Medtronic Vascular | 1999 | — |
| 0 (this device) | K031037 | Eopa Elongated One-Piece Arterial Cannula and Guidewire | Medtronic Perfusion Systems | 2003 | — |
| +1 | K031518 | Eopa Cap Elongated One-Piece Arterial Cannula with Pressure Monitoring Line… | Medtronic Vascular | 2003 | — |
| +1 | K111972 | Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm Cardioplegia… | Medtronic, Inc. | 2011 | — |
| +2 | K120988 | DLP Single Stage Venous Cannula, DLP Right Angle Single Stage Venous Cannula… | Medtronic, Inc. | 2012 | 2 |
| +1 | K110980 | Centrimag Return Cannula Kit | Levitronix, LLC | 2011 | — |
| +2 | K152161 | Thoratec CentriMag Return (Arterial) Cannula Kit | Thoratec Corporation | 2015 | — |
Other ancestors by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 1989 | 1 | 0 |
| 2000 | 1 | 0 |
Descendants by year cleared
| Year cleared | Devices | With a Class I recall |
|---|---|---|
| 2003 | 1 | 0 |
| 2011 | 2 | 0 |
| 2012 | 1 | 0 |
| 2014 | 1 | 0 |
| 2015 | 1 | 0 |
| 2020 | 1 | 0 |
| 2024 | 1 | 0 |
| 2025 | 1 | 0 |
| 2026 | 2 | 0 |
Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.
Planning a device in DWF? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.
Top candidate predicates for a new DWF device
- #1 K251258 MC2™ Two-Stage Venous Cannula; MC2X™…score 14.6 · 8 generations · no Class I recall
- #2 Held for the reportin the report
- #3 Held for the reportin the report
What was it cleared against? (3)
Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.
| Type | Number | Device | Decided |
|---|---|---|---|
| * | K991066 | Madtronic DLP Arterial CannulaeSummary page 1…EOPA Elongated One-Piece Arterial Cannula with Guidewire 20, 22 Fr. K991066 and KO00274 Extracorporeal Circuit with Bio-Active Surface K891687 === Pa… | 1999 |
| * | K000274 | Eopa Elongated One-Piece Arterial Cannula with Guidewire | 2000 |
| * | K891687 | Extracorporeal Circuit with Bio-Active SurfaceSummary page 1…20, 22 Fr. K991066 and KO00274 Extracorporeal Circuit with Bio-Active Surface K891687 Device Description: The EOPA Elongated One-Piece Arterial Cannulae have a flexible, thin wall wirewound body with… | 1989 |
What descends from it? (11)
See all 113 later devices cite it directly; ranked here by how many devices descend from each in turn.
Has anything in its family been recalled? (0 Class I)
Recalls of this device
| Date | Class | Recall | Root cause | Reason |
|---|---|---|---|---|
| Jun 24, 2026 | Class II | Z-2809-2026 | Under Investigation by firm | Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures. |
| Jun 24, 2026 | Class II | Z-2808-2026 | Under Investigation by firm | Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures. |
| Jun 24, 2026 | Class II | Z-2806-2026 | Under Investigation by firm | Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures. |
| Jun 24, 2026 | Class II | Z-2807-2026 | Under Investigation by firm | Medtronic received reports of pinhole leaks in the wire-wound body of 22 Fr EOPA 3D Arterial Cannulae and EOPA Elongated One-Piece Arterial Cannulae used during cardiopulmonary bypass procedures. |
| Dec 12, 2024 | Class II | Z-0985-2025 | Packaging | Incorrect labeling for seven manufactured lots of certain models of Arterial Cannulae. |
| Jul 24, 2015 | Class II | Z-2420-2015 | Packaging process control | EOPA Arterial Cannula Devices in this lot were shipped without the guidewire. |
The 5 latest of 6 recalls; the web page lists all.
No ancestor or descendant has a Class I recall.
What does this lineage say?
- The line of descent spans 15 years and passes through 3 companies; a new generation was cleared every 11.4 years on median.
- 1 of the 4 ancestors has a recall of any class.
- Of its descendants, K031518 (Medtronic Vascular, 2003) has gone furthest, with 6 descendants of its own.
How does its category compare?
All DWF devicesDWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.
FDA decisions per year · DWF
4in 2025
2026 to Oct 25, outlined.
Median 510(k) review, days · DWF
89in 2025
Receipt to decision; years with fewer than 3 clearances left blank.
Recalls per year · DWF
4in 2024
Class I (3 since 2007) Class II/III not classified
Adverse event reports per year · DWF
1,020in 2025
MAUDE reports naming product code DWF, not matched to any one device; reports are not rates.
View these four charts as a table
| Year | Decisions | Median 510(k) review, days | Recalls (Class I) | MAUDE reports |
|---|---|---|---|---|
| 2007 | 2 | — | 25 | 121 |
| 2008 | 6 | 95 | 25 | 126 |
| 2009 | 8 | 148 | 21 | 99 |
| 2010 | 3 | 62 | 26 | 197 |
| 2011 | 5 | 85 | 13 | 212 |
| 2012 | 10 | 110 | 37 | 233 |
| 2013 | 9 | 110 | 14 (1) | 238 |
| 2014 | 14 | 110 | 6 | 325 |
| 2015 | 9 | 46 | 3 (1) | 218 |
| 2016 | 3 | 32 | 3 | 144 |
| 2017 | 5 | 127 | 2 (1) | 101 |
| 2018 | 5 | 30 | 10 | 118 |
| 2019 | 3 | 120 | 2 | 179 |
| 2020 | 3 | 30 | 0 | 142 |
| 2021 | 5 | 358 | 3 | 162 |
| 2022 | 2 | — | 0 | 276 |
| 2023 | 2 | — | 0 | 470 |
| 2024 | 3 | 29 | 4 | 401 |
| 2025 | 4 | 89 | 0 | 1,020 |
| 2026 | 8 | 81 | 20 | — |
What is it indicated for?
This product is intended for use with cardiopulmonary bypass as an arterial return cannula.
Indications for use as stated in the FDA summary.
https://510kreport.com/device/K031037 · printed Nov 6, 2026
Cite this page
510(k) Report. “K031037: Eopa Elongated One-Piece Arterial Cannula and Guidewire.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031037
Report an extraction error
Are you the manufacturer?
Record history
- FDA decision record loaded May 2, 2026
- Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
- Lineage and recalls computed Nov 3, 2026
- Page last updated Nov 3, 2026