Extracorporeal Circuit with Bio-Active Surface

K891687510(k) TraditionalCleared May 31, 198970-day reviewClass II21 CFR 870.4350Recalls 10

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Medtronic Blood Systems, Inc. · DTZ Oxygenator, Cardiopulmonary Bypass

Cleared on May 31, 1989 after a 70-day review. FDA has not posted a summary or statement for it, so its predicates are not known. 4 of the 74 devices in its family, counting itself, have a Class I recall. 73 later devices descend from it. It has been recalled 10 times.

Decided
May 31, 1989
Review
70 days
Class
Class II
Product code
DTZ
Recalls
10
Predicates
0

73 later devices descend from it, 8 citing it directly. 4 of the 74 devices in this family, counting this one, have a Class I recall.

Descendants, by year cleared 2003: 7 descendants cleared, 1 with a Class I recall11998: 1 descendant cleared2011: 8 descendants cleared, 1 with a Class I recall11997: 1 descendant cleared2009: 6 descendants cleared2005: 5 descendants cleared2006: 4 descendants cleared2014: 3 descendants cleared, 1 with a Class I recall12012: 2 descendants cleared2000: 2 descendants cleared2008: 5 descendants cleared2010: 5 descendants cleared2020: 2 descendants cleared2002: 1 descendant cleared2019: 2 descendants cleared, 1 with a Class I recall12013: 3 descendants cleared2025: 4 descendants cleared2015: 2 descendants cleared2026: 4 descendants cleared2024: 3 descendants cleared2001: 1 descendant cleared2023: 2 descendants cleared1990200020102020
  1. This deviceK891687 Extracorporeal Circuit with Bio-Active SurfaceMedtronic Blood Systems, Inc. · 1989
  2. Descendant · cites it directly · 48 of its ownK973011 BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc. · 1998
  3. Descendant · 2 generations on · 21 of its ownK020271 Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003 · Recalled (Class I)
  4. Descendant · cites it directly · 15 of its ownK031827 Ultraflex Venous Cannula, Model 970XX and CB970XXMedtronic Perfusion Systems · 2003
  5. Descendant · cites it directly · 13 of its ownK031700 Medtronic Resting Heart System, Model AAR1000 and 711Medtronic Perfusion Systems · 2003 · Recalled
  6. Descendant · 2 generations on · 13 of its ownK991783 AB-180 XC System, Model AB-180XCCardiacassist, Inc. · 2000

Showing 5 of 73 descendants. See all descendants

Descendants cleared per year

1997: 11998: 12000: 22001: 12002: 12003: 72005: 52006: 42008: 52009: 62010: 52011: 82012: 22013: 32014: 32015: 22019: 22020: 22023: 22024: 32025: 42026: 4
1997high 8 (2011) · low 0 (1999)2026

Class I recalls in this family

Most influential descendants

Showing 5 of 73 · See all
  • K973011BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc. · 1998 · cites it directly48descendants
  • K020271Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003 · 2 generations on21descendants
  • K031827Ultraflex Venous Cannula, Model 970XX and CB970XXMedtronic Perfusion Systems · 2003 · cites it directly15descendants
  • K031700Medtronic Resting Heart System, Model AAR1000 and 711Medtronic Perfusion Systems · 2003 · cites it directly13descendants
  • K991783AB-180 XC System, Model AB-180XCCardiacassist, Inc. · 2000 · 2 generations on13descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
0 (this device)K891687Extracorporeal Circuit with Bio-Active SurfaceMedtronic Blood Systems, Inc.1989—
+1K973011BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc.1998—
+2K020271Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC2003Class I
+1K031827Ultraflex Venous Cannula, Model 970XX and CB970XXMedtronic Perfusion Systems2003—
+1K031700Medtronic Resting Heart System, Model AAR1000 and 711Medtronic Perfusion Systems20031
+2K991783AB-180 XC System, Model AB-180XCCardiacassist, Inc.2000—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
199710
199810
200020
200110
200210
200371
200550
200640
200850
200960
201050
201181
201220
201330
201431
201520
201921
202020
202320
202430
202540
202640
FIG. 1 — Lineage of K891687, 1989–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

4 devices in this family have a Class I recall. A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new DTZ device

  1. #1 K240934 Quantum Perfusion Blood Oxygenator ECC…score 14.3 · 12 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against?

FDA has not posted a summary or statement for it, so its predicates are not known.

What descends from it? (73)

See all 73

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K973011BPX80 Bio-Pump (BPX80)Medtronic Bio-Medicus, Inc.199848
K031827Ultraflex Venous Cannula, Model 970XX and CB970XXMedtronic Perfusion Systems200315
K031700Medtronic Resting Heart System, Model AAR1000 and 711Medtronic Perfusion Systems200313
K033416Select 3D Arterial Cannula with Carmeda Bioactive Surface and Select Cap Arterial…Medtronic Perfusion Systems200311
K031037Eopa Elongated One-Piece Arterial Cannula and GuidewireMedtronic Perfusion Systems200311
K031776MC2X Multi-Stage Venous CannulaMedtronic Perfusion Systems20037
K111972Affinity CP Centrifugal Pump, Resting Heart Sys.,Myotherm Cardioplegia Deliv.,Affinity…Medtronic, Inc.20115
K973475Carmeda Coated Cardiotherm Blood Cardioplegia Systems (Cbctxxx Series and Cbctxxxbr…Medtronic Vascular19970

Has anything in its family been recalled? (4 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Aug 8, 2018Class IIZ-2495-2020Nonconforming Material/ComponentAffected products failed a sterilization test.
Feb 4, 2016Class IIIZ-1023-2016Component design/selectionMedtronic is notifying customers who have or may receive specific lots of Medtronic Perfusion Tubing Packs that include the potentially affected cuvettes manufactured by Terumo Cardiovascular Systems (Terumo CVS). Terumo Cardiovascular…
Aug 18, 2015Class IIZ-2736-2015Nonconforming Material/ComponentHemocor High Performance Hemoconcentrators demonstrate low ultrafiltration performance that is below product specification. These potentially affected units have been distributed as stand-alone devices and in specific lots of Medtronic…
Jul 17, 2013Class IIZ-2105-2013Process controlMedtronic was notified that Edwards Lifesciences has initiated a product recall for the Rigid Suction Wand, model S099B. The Wands are included within certain Perfusion Tubing Packs that are manufactured and distributed by Medtronic.
Nov 19, 2012Class IIZ-0618-2013Package design/selectionMedtronic is initiating an Urgent Medical Device Customer Notification. We have confirmed that certain Medtronic Intersept Tubing Packs contain packaging trays that are susceptible to damage that can compromise product sterility.
Dec 14, 2011Not classifiedZ-0813-2012Labeling designInstructions for Use for the DLP Single Stage Venous Cannulae are in the process of being updated. The current precaution statement referencing ventricular assist procedures will be identified as a contraindication in the updated…
May 9, 2008Not classifiedZ-1920-2008Nonconforming Material/ComponentMedtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently…
Aug 1, 2007Not classifiedZ-1227-2007OtherMedtronic Custom Tubing Packs might have open header bag seals.
Jan 2, 2007Not classifiedZ-0583-2007OtherOne lot of Custom Perfusion Sets, Catalog 2T10R1 was incorrectly assembled. The one way vacuum relief valve on the 102 inch line is attached backwards, thus preventing flow.
Sep 15, 2004Not classifiedZ-1489-04OtherSpiral tip from DLP VAD cannula may be loose and potentially detach during use.

Relatives with a Class I recall

Class I is FDA’s most serious recall class. A recall in the family tree is context for this device, not evidence about it.

  • K020271Levitronix Centrimag Extracorporeal Blood Pumping System, Model L-100Levitronix, LLC · 2003 · 2 generations on · ◆ in the 2003 descendants bar of FIG. 11 Class I recall
  • K102726Cardiohelp SystemMaquet Cardiopulmonary, AG · 2011 · 3 generations on · ◆ in the 2011 descendants bar of FIG. 11 Class I recall
  • K133598Cardiohelp SystemMaquet Cardiopulmonary, AG · 2014 · 4 generations on · ◆ in the 2014 descendants bar of FIG. 11 Class I recall
  • K183623LifeSPARC Pump, LifeSPARC ControllerCardiacassist, Inc. · 2019 · 4 generations on · ◆ in the 2019 descendants bar of FIG. 12 Class I recalls

What does this lineage say?

  • Of its descendants, K973011 (Medtronic Bio-Medicus, Inc., 1998) has gone furthest, with 48 descendants of its own.

How does its category compare?

All DTZ devices

DTZ Oxygenator, Cardiopulmonary Bypass: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · DTZ

2in 2025

2007: 42008: 32009: 92010: 22011: 42012: 92013: 92014: 52015: 92016: 32017: 22018: 32019: 52020: 52021: 52022: 22023: 52024: 32025: 22026 (partial year): 3
2007high 9 (2009) · low 2 (2010)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · DTZ

124in 2024

2007: 55 days median2008: 88 days median2009: 210 days median2011: 160 days median2012: 90 days median2013: 28 days median2014: 28 days median2015: 134 days median2016: 62 days median2018: 72 days median2019: 90 days median2020: 93 days median2021: 336 days median2023: 145 days median2024: 124 days median2026 (partial year): 137 days median
2007high 336 (2021) · low 28 (2013)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · DTZ

1in 2025

2007: 72008: 52010: 12011: 132012: 142014: 22016: 12018: 2, 2 Class I2019: 12020: 32021: 1, 1 Class I2022: 62025: 1
2007high 14 (2012) · low 0 (2009)2026

Class I (3 since 2007) Class II/III not classified

Adverse event reports per year · DTZ

976in 2025

2007: 1342008: 1052009: 522010: 1192011: 1902012: 2152013: 3462014: 7642015: 6582016: 4432017: 3882018: 3202019: 4892020: 4422021: 5952022: 5592023: 7382024: 5352025: 976
2007high 976 (2025) · low 52 (2009)2025

MAUDE reports naming product code DTZ, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20074557134
20083885105
20099210052
20102—1119
2011416013190
201299014215
20139280346
20145282764
201591340658
20163621443
20172—0388
20183722 (2)320
20195901489
20205933442
202153361 (1)595
20222—6559
202351450738
202431240535
20252—1976
202631370—
Cite this page

510(k) Report. “K891687: Extracorporeal Circuit with Bio-Active Surface.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 4, 2026. https://510kreport.com/device/K891687

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026