Nexgen Porous, Uncemented Femoral and Tibial Baseplate Components

K031061510(k) TraditionalCleared Oct 9, 2003189-day reviewClass II21 CFR 888.3565Recalls 24

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Cleared on Oct 9, 2003 as substantially equivalent to K933785 (Zimmer, Inc., 1995) and 2 other predicates after a 189-day review. Its family tree reaches back 2 generations to 1995. 193 later devices descend from it. It has been recalled 24 times.

Decided
Oct 9, 2003
Review
189 days
Class
Class II
Product code
MBH
Recalls
24
Predicates
3

Its longest line of descent runs back 2 generations to K933785 (1995). 4 ancestors in all. 193 later devices descend from it, 8 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1996: 1 other ancestor1999: 1 other ancestor2004: 2 descendants cleared2018: 11 descendants cleared2007: 2 descendants cleared2005: 1 descendant cleared2008: 1 descendant cleared2013: 12 descendants cleared2009: 2 descendants cleared2012: 10 descendants cleared2020: 17 descendants cleared2011: 5 descendants cleared2014: 8 descendants cleared2015: 4 descendants cleared2019: 7 descendants cleared2016: 3 descendants cleared2010: 2 descendants cleared2022: 12 descendants cleared2024: 19 descendants cleared2026: 17 descendants cleared2023: 15 descendants cleared2017: 4 descendants cleared2021: 17 descendants cleared2025: 22 descendants clearedK933785 · Next Generation Knee System · Zimmer, Inc. · 1995 · recalledK023211 · Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral Components · Zimmer, Inc. · 2002 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 8 descendants2 generations forward: 25 descendants3 generations forward: 41 descendants4 generations forward: 70 descendants5 generations forward: 35 descendants6 generations forward: 12 descendants7 generations forward: 2 descendants K023211 · Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral Components · Zimmer, Inc. · 2002 · recalledK933785 · Next Generation Knee System · Zimmer, Inc. · 1995 · recalled0+5
  1. Root · 2 generations backK933785 Next Generation Knee SystemZimmer, Inc. · 1995 · Recalled
  2. 1 generation backK023211 Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral…Zimmer, Inc. · 2002 · Recalled
  3. This deviceK031061 Nexgen Porous, Uncemented Femoral and Tibial Baseplate ComponentsZimmer, Inc. · 2003
  4. Descendant · cites it directly · 108 of its ownK072160 Nexgen Trabecular Metal Tibial TrayZimmer, Inc. · 2007 · Recalled
  5. Descendant · cites it directly · 102 of its ownK072619 Nexgen Complete Knee Solution LPS-Flex and Knee Gender Solutions…Zimmer, Inc. · 2007 · Recalled
  6. Descendant · cites it directly · 78 of its ownK081023 Evolis Total Knee SystemMedacta International S.A. · 2008 · Recalled
  7. Descendant · 2 generations on · 59 of its ownK121771 Zimmer Persona the Personalized Knee SystemZimmer, Inc. · 2012 · Recalled
  8. Descendant · cites it directly · 59 of its ownK043101 Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw…Zimmer, Inc. · 2005 · Recalled

Showing 5 of 193 descendants. See all descendants

Descendants cleared per year

2004: 22005: 12007: 22008: 12009: 22010: 22011: 52012: 102013: 122014: 82015: 42016: 32017: 42018: 112019: 72020: 172021: 172022: 122023: 152024: 192025: 222026: 17
2004high 22 (2025) · low 0 (2006)2026

Most influential descendants

Showing 5 of 193 · See all
  • K072160Nexgen Trabecular Metal Tibial TrayZimmer, Inc. · 2007 · cites it directly108descendants
  • K072619Nexgen Complete Knee Solution LPS-Flex and Knee Gender Solutions Female (GSF) Porous…Zimmer, Inc. · 2007 · cites it directly102descendants
  • K081023Evolis Total Knee SystemMedacta International S.A. · 2008 · cites it directly78descendants
  • K121771Zimmer Persona the Personalized Knee SystemZimmer, Inc. · 2012 · 2 generations on59descendants
  • K043101Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem ExtensionsZimmer, Inc. · 2005 · cites it directly59descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−2K933785Next Generation Knee SystemZimmer, Inc.199583
−1K023211Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral ComponentsZimmer, Inc.20026
0 (this device)K031061Nexgen Porous, Uncemented Femoral and Tibial Baseplate ComponentsZimmer, Inc.2003—
+1K072160Nexgen Trabecular Metal Tibial TrayZimmer, Inc.20071
+1K072619Nexgen Complete Knee Solution LPS-Flex and Knee Gender Solutions Female (GSF)…Zimmer, Inc.20074
+1K081023Evolis Total Knee SystemMedacta International S.A.20081
+2K121771Zimmer Persona the Personalized Knee SystemZimmer, Inc.20122
+1K043101Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem…Zimmer, Inc.200522

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200420
200510
200720
200810
200920
201020
201150
2012100
2013120
201480
201540
201630
201740
2018110
201970
2020170
2021170
2022120
2023150
2024190
2025220
2026170
FIG. 1 — Lineage of K031061, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in MBH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new MBH device

  1. #1 K241597 Freedom® Total Knee System - Porous Tibial…score 19.9 · 8 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K933785Next Generation Knee System
Summary page 1
…888.3565 Predicate Devices: NexGen Complete Knee Solution, manufactured by Zimmer, K933785, cleared January 30, 1995, NexGen Cornplele Knee Solution Cniciate Retaining…
Zimmer, Inc.199583
*K023211Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral ComponentsZimmer, Inc.20026
*K960279Nexgen Compl. Knee Sol. Legacy Posterior Stabilized/Constrained Condylar Knee Femoral…
Summary page 1
…onstrained Condylar Knee (CCK) Femoral Components and Articular Surfaces, manufactured by Zimmer, K960279, cIeared April 26, 1996. Device Description: This submission is for the unce…
Zimmer, Inc.199621

Also cleared against K933785: 14 other devices →

What descends from it? (193)

See all 193

8 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K072160Nexgen Trabecular Metal Tibial TrayZimmer, Inc.2007108
K072619Nexgen Complete Knee Solution LPS-Flex and Knee Gender Solutions Female (GSF) Porous…Zimmer, Inc.2007102
K081023Evolis Total Knee SystemMedacta International S.A.200878
K043101Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem ExtensionsZimmer, Inc.200559
K042271Nexgen Complete Knee Solution LPS-Flex Prolong Highly Crosslinked Polyethylene Articular…Zimmer, Inc.200416
K122160Vanguard XP Knee SystemBiomet Manufacturing Corp20136
K041100Nexgen Porous, HA/TCP, Uncemented Femoral and Tibial Baseplate ComponentsZimmer, Inc.20046
K173057Zimmer Knee Joint Replacement Prostheses MR LabelingZimmer, Inc.20185

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Mar 28, 2024Class IIZ-1608-2024Process change controlOut of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
Nov 29, 2017Class IIZ-1039-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Nov 29, 2017Class IIZ-1030-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Nov 29, 2017Class IIZ-1038-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Nov 29, 2017Class IIZ-1040-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Nov 29, 2017Class IIZ-1041-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Nov 29, 2017Class IIZ-1048-2018PackagingThe LDPE bag packaging for various highly polished hip and knee implants may adhere to the highly polished surface, leaving residue or material from the LDPE bag on the implant after it is removed from the bag.
Apr 30, 2012Class IIZ-2092-2012Component design/selectionThe design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall…
Apr 30, 2012Class IIZ-2091-2012Component design/selectionThe design of the spring clip for the jaw assembly was updated in March 2011 due to fractures of the spring clip. Zimmer has received complaints for the spring clip fracturing on the updated design. The spring clip may fracture or fall…
Mar 14, 2011Not classifiedZ-1993-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1996-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-2174-2011Device DesignThe firm is initiating a recall due to the potiential for the spring clip to break off during use and fall into the surgical site. The firm has received 24 complaints of the spring breaking, four of which were reported that the patient…
Mar 14, 2011Not classifiedZ-1995-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1992-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1997-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1988-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-2173-2011Device DesignThe firm is initiating a recall due to the potiential for the spring clip to break off during use and fall into the surgical site. The firm has received 24 complaints of the spring breaking, four of which were reported that the patient…
Mar 14, 2011Not classifiedZ-1994-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1985-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1989-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1986-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1991-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1987-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…
Mar 14, 2011Not classifiedZ-1990-2011Device DesignThe firm is initiating a recall due to the potential for breakdown of the AlTiN PVD (black) coating during surgery and the coating could fall into the surgical site. The firm has received 22 reports of the black coating coming off the…

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 8 years and passes through 1 company; a new generation was cleared every 7.7 years on median.
  • 4 of the 4 ancestors have a recall of any class.
  • Of its descendants, K072160 (Zimmer, Inc., 2007) has gone furthest, with 108 descendants of its own.
  • 14 other devices also cite its predicate K933785.

How does its category compare?

All MBH devices

MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · MBH

8in 2025

2007: 102008: 32009: 32010: 42011: 52012: 52013: 62014: 92015: 82016: 42017: 42018: 62019: 12020: 72021: 62022: 72023: 62024: 102025: 82026 (partial year): 7
2007high 10 (2007) · low 1 (2019)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · MBH

98in 2025

2007: 59 days median2008: 70 days median2009: 209 days median2010: 84 days median2011: 84 days median2012: 90 days median2013: 138 days median2014: 114 days median2015: 100 days median2016: 74 days median2017: 115 days median2018: 89 days median2020: 112 days median2021: 48 days median2022: 57 days median2023: 94 days median2024: 84 days median2025: 98 days median2026 (partial year): 153 days median
2007high 209 (2009) · low 48 (2021)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · MBH

1in 2024

2009: 22011: 152012: 1, 1 Class I2013: 4, 4 Class I2016: 12017: 92018: 12019: 12020: 22023: 22024: 12026 (partial year): 4
2007high 15 (2011) · low 0 (2007)2026

Class I (5 since 2007) Class II/III not classified

Adverse event reports per year · MBH

1,561in 2025

2007: 272008: 212009: 892010: 1212011: 2472012: 3702013: 3082014: 3852015: 9552016: 1,0342017: 1,0072018: 1,1912019: 1,3942020: 1,0892021: 1,5392022: 1,2612023: 1,2632024: 1,4972025: 1,561
2007high 1,561 (2025) · low 21 (2008)2025

MAUDE reports naming product code MBH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071059027
2008370021
20093209289
20104840121
201158415247
20125901 (1)370
201361384 (4)308
201491140385
201581000955
201647411,034
2017411591,007
201868911,191
20191—11,394
2020711221,089
202164801,539
202275701,261
202369421,263
2024108411,497
202589801,561
202671534—

What is it indicated for?

This device is indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic &tis, polyarthritis. - Collagen disorders, andor avascular necrosis of the femoral condyle. - Post-traumatic loss of joint configuration, particularly when there is potellofemoral erosion, dysfunction or prior patellectomy. - Moderate valgus, varus, or flexion deformities. - The salvage of previously failed surgical…

Read the full statement (120 words)

This device is indicated for patients with severe knee pain and disability due to:

- Rheumatoid arthritis, osteoarthritis, traumatic &tis, polyarthritis.

- Collagen disorders, andor avascular necrosis of the femoral condyle.

- Post-traumatic loss of joint configuration, particularly when there is potellofemoral erosion, dysfunction or prior patellectomy.

- Moderate valgus, varus, or flexion deformities.

- The salvage of previously failed surgical attempts or For a knee in which satisfactors stabiIity in flexion cannot be obtained at the time of surgery,

CR,LPS and CR-Flex porous coaled femoral or tibial baseplate components may be used cemented or uncernented (biological fixation). All other femoral, tibial baseplate md dl-polyethylene patella components (including H m C P coated) are indicated for cemented USE onIy .

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031061: Nexgen Porous, Uncemented Femoral and Tibial Baseplate Components.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031061

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026