Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem Extensions

K043101510(k) SpecialCleared Mar 7, 2005118-day reviewClass II21 CFR 888.3560Recalls 22

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Zimmer, Inc. · JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Cleared on Mar 7, 2005 as substantially equivalent to K933785 (Zimmer, Inc., 1995) and 2 other predicates after a 118-day review. Its family tree reaches back 3 generations to 1995. 59 later devices descend from it. It has been recalled 22 times.

Decided
Mar 7, 2005
Review
118 days
Class
Class II
Product code
JWH
Recalls
22
Predicates
3

Its longest line of descent runs back 3 generations to K933785 (1995). 5 ancestors in all. 59 later devices descend from it, 3 citing it directly.

Other ancestors, by year clearedDescendants, by year cleared 1999: 1 other ancestor1996: 1 other ancestor2018: 4 descendants cleared2009: 1 descendant cleared2013: 4 descendants cleared2014: 2 descendants cleared2010: 2 descendants cleared2011: 2 descendants cleared2022: 3 descendants cleared2012: 4 descendants cleared2020: 5 descendants cleared2024: 2 descendants cleared2026: 6 descendants cleared2019: 1 descendant cleared2017: 3 descendants cleared2023: 4 descendants cleared2021: 9 descendants cleared2016: 1 descendant cleared2025: 6 descendants clearedK933785 · Next Generation Knee System · Zimmer, Inc. · 1995 · recalledK023211 · Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral Components · Zimmer, Inc. · 2002 · recalledK031061 · Nexgen Porous, Uncemented Femoral and Tibial Baseplate Components · Zimmer, Inc. · 2003 · recalled1995200020052010201520202025
Other ancestors, by generationDescendants, by generation 1 generation back: 1 other ancestor2 generations back: 1 other ancestor 1 generation forward: 3 descendants2 generations forward: 19 descendants3 generations forward: 20 descendants4 generations forward: 9 descendants5 generations forward: 6 descendants6 generations forward: 2 descendants K031061 · Nexgen Porous, Uncemented Femoral and Tibial Baseplate Components · Zimmer, Inc. · 2003 · recalledK023211 · Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral Components · Zimmer, Inc. · 2002 · recalledK933785 · Next Generation Knee System · Zimmer, Inc. · 1995 · recalled0+5
  1. Root · 3 generations backK933785 Next Generation Knee SystemZimmer, Inc. · 1995 · Recalled
  2. 2 generations backK023211 Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral…Zimmer, Inc. · 2002 · Recalled
  3. 1 generation backK031061 Nexgen Porous, Uncemented Femoral and Tibial Baseplate ComponentsZimmer, Inc. · 2003 · Recalled
  4. This deviceK043101 Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw…Zimmer, Inc. · 2005
  5. Descendant · cites it directly · 52 of its ownK090988 GMK Total Knee SystemMedacta International S.A. · 2009 · Recalled
  6. Descendant · 2 generations on · 34 of its ownK102437 GMK Revision Knee SystemMedacta International S.A. · 2010
  7. Descendant · 2 generations on · 28 of its ownK103170 GMK Revision Knee SC LinersMedacta International S.A. · 2011
  8. Descendant · 2 generations on · 23 of its ownK113571 GMK Resufacing Patella Size 4Medacta International · 2012
  9. Descendant · 2 generations on · 20 of its ownK121416 GMK SphereMedacta International · 2012

Showing 5 of 59 descendants. See all descendants

Descendants cleared per year

2009: 12010: 22011: 22012: 42013: 42014: 22016: 12017: 32018: 42019: 12020: 52021: 92022: 32023: 42024: 22025: 62026: 6
2009high 9 (2021) · low 0 (2015)2026

Most influential descendants

Showing 5 of 59 · See all
  • K090988GMK Total Knee SystemMedacta International S.A. · 2009 · cites it directly52descendants
  • K102437GMK Revision Knee SystemMedacta International S.A. · 2010 · 2 generations on34descendants
  • K103170GMK Revision Knee SC LinersMedacta International S.A. · 2011 · 2 generations on28descendants
  • K113571GMK Resufacing Patella Size 4Medacta International · 2012 · 2 generations on23descendants
  • K121416GMK SphereMedacta International · 2012 · 2 generations on20descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−3K933785Next Generation Knee SystemZimmer, Inc.199583
−2K023211Nexgen Complete Knee Solution Cruciate Retaining (CR)-Flex Femoral ComponentsZimmer, Inc.20026
−1K031061Nexgen Porous, Uncemented Femoral and Tibial Baseplate ComponentsZimmer, Inc.200324
0 (this device)K043101Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem…Zimmer, Inc.2005—
+1K090988GMK Total Knee SystemMedacta International S.A.20091
+2K102437GMK Revision Knee SystemMedacta International S.A.2010—
+2K103170GMK Revision Knee SC LinersMedacta International S.A.2011—
+2K113571GMK Resufacing Patella Size 4Medacta International2012—
+2K121416GMK SphereMedacta International2012—

Other ancestors by year cleared

Year clearedDevicesWith a Class I recall
199610
199910

Descendants by year cleared

Year clearedDevicesWith a Class I recall
200910
201020
201120
201240
201340
201420
201610
201730
201840
201910
202050
202190
202230
202340
202420
202560
202660
FIG. 1 — Lineage of K043101, 1995–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in JWH? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new JWH device

  1. #1 K221479 Persona Personalized Knee Systemscore 22.4 · 11 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (3)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K933785Next Generation Knee System
Summary page 1
…NexGen Precoat Stemmed Tibial Plates and Stem Extensions, manufactured by Zimmer, Inc., K933785, cleared January 30, 1995. NexGen Locking Screws and Taper Stem Plugs,…
Zimmer, Inc.199583
*K991581Nexgen Complete Knee Solution Legacy Posterior Stabilized (LPS); LPS-Flex Fixed Bearing…
Summary page 1
…red January 30, 1995. NexGen Locking Screws and Taper Stem Plugs, manufactured by Zimmer, Inc., K991581, cleared July 30, 1999. NexGen Porous, Uncemented Femor…
Zimmer, Inc.19997
*K031061Nexgen Porous, Uncemented Femoral and Tibial Baseplate Components
Summary page 1
…exGen Porous, Uncemented Femoral and Tibial Baseplate Components, manufactured by Zimmer, Inc., K031061, cleared October 3, 2003. Device Description: The NexGen MIS Tibial Components are part of the…
Zimmer, Inc.200324

Also cleared against K933785: 14 other devices →

What descends from it? (59)

See all 59

3 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K090988GMK Total Knee SystemMedacta International S.A.200952
K173057Zimmer Knee Joint Replacement Prostheses MR LabelingZimmer, Inc.20185
K133630GMK Short StemMedacta International20131

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jan 11, 2016Class IIZ-1148-2016PackagingAs a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
May 19, 2014Class IIZ-1935-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1936-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1933-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1931-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1932-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1934-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1938-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1937-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
May 19, 2014Class IIZ-1939-2014Equipment maintenanceZimmer is initiating a voluntary recall of specific identified lots due to the potential that the threads may be out of specification.
Oct 11, 2013Class IIZ-1023-2014PackagingZimmer recently conducted a review of all historic packaging validations at its Zimmer Manufacturing B.V. (ZMBV) facility in Ponce, Puerto Rico. Based on this review, it was determined that the packaging operations conducted in the…
Apr 26, 2010Not classifiedZ-2415-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2414-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2411-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2416-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2413-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2408-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2418-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2417-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2409-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2410-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.
Apr 26, 2010Not classifiedZ-2412-2010OtherThe firm has received complaints of loosening of the implanted device requiring revision surgery. There have been 114 MDRs filed all reported that the device loosened and the patient required additional surgery to replace the device.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • The line of descent spans 10 years and passes through 1 company; a new generation was cleared every 1.4 years on median.
  • 5 of the 5 ancestors have a recall of any class.
  • Of its descendants, K090988 (Medacta International S.A., 2009) has gone furthest, with 52 descendants of its own.
  • 14 other devices also cite its predicate K933785.

How does its category compare?

All JWH devices

JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · JWH

20in 2025

2007: 182008: 172009: 112010: 192011: 252012: 332013: 312014: 332015: 272016: 182017: 252018: 252019: 282020: 222021: 242022: 172023: 202024: 192025: 202026 (partial year): 12
2007high 33 (2012) · low 11 (2009)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · JWH

55in 2025

2007: 74 days median2008: 81 days median2009: 168 days median2010: 85 days median2011: 87 days median2012: 85 days median2013: 128 days median2014: 98 days median2015: 87 days median2016: 88 days median2017: 118 days median2018: 90 days median2019: 133 days median2020: 59 days median2021: 61 days median2022: 56 days median2023: 57 days median2024: 89 days median2025: 55 days median2026 (partial year): 63 days median
2007high 168 (2009) · low 55 (2025)2026

Receipt to decision; years with fewer than 3 clearances left blank.

Recalls per year · JWH

2in 2025

2007: 142008: 102009: 1092010: 322011: 272012: 852013: 312014: 132015: 112016: 172017: 612018: 142019: 52020: 22021: 82022: 122023: 672024: 72025: 22026 (partial year): 4
2007high 109 (2009) · low 2 (2020)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · JWH

9,306in 2025

2007: 1,3052008: 1,4832009: 1,8702010: 3,2862011: 5,1332012: 5,2032013: 6,1352014: 6,7222015: 6,8812016: 7,5862017: 12,2632018: 14,0712019: 14,9872020: 12,5922021: 9,4082022: 8,3642023: 8,9472024: 10,8282025: 9,306
2007high 14,987 (2019) · low 1,305 (2007)2025

MAUDE reports naming product code JWH, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20071874141,305
20081781101,483
2009111681091,870
20101985323,286
20112587275,133
20123385855,203
201331128316,135
20143398136,722
20152787116,881
20161888177,586
2017251186112,263
201825901414,071
201928133514,987
20202259212,592
2021246189,408
20221756128,364
20232057678,947
20241989710,828
2025205529,306
202612634—

What is it indicated for?

This device is indicated for patients with severe knee pain and disability due to:

- Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.

- Collagen disorders and/or avascular necrosis of the femoral condyle.

- Post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy

- Moderate valgus, varus or flexion deformities.

- The salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.

This device is intended for cemented use only.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K043101: Nexgen Complete Knee Solution Mis Tibial Components, Locking Screw and Stem Extensions.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 5, 2026. https://510kreport.com/device/K043101

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026