510(k) Report

510(k) · Cleared

Recalled · Class II · 2024

Bayer Advia Centaur CA 19-9 Assay

K031393510(k) TraditionalCleared Jun 24, 200353-day reviewClass II21 CFR 866.6010Recalls 8

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Bayer Healthcare, LLC · NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer

Cleared on Jun 24, 2003 as substantially equivalent to K020566 (Fujirebio Diagnostics,Inc., 2002) after a 53-day review. Its family tree reaches back 1 generation to 2002. 2 later devices descend from it. It has been recalled 8 times.

Decided
Jun 24, 2003
Review
53 days
Class
Class II
Product code
NIG
Recalls
8
Predicates
1

Its longest line of descent runs back 1 generation to K020566 (2002). 2 later devices descend from it, 2 citing it directly.

Descendants, by year cleared 2011: 1 descendant cleared2026: 1 descendant clearedK020566 · Fujirebio Diagnostics CA 19-9 Ria · Fujirebio Diagnostics,Inc. · 2002200020052010201520202025
Descendants, by generation 1 generation forward: 2 descendants K020566 · Fujirebio Diagnostics CA 19-9 Ria · Fujirebio Diagnostics,Inc. · 20020
  1. Root · 1 generation backK020566 Fujirebio Diagnostics CA 19-9 RiaFujirebio Diagnostics,Inc. · 2002
  2. This deviceK031393 Bayer Advia Centaur CA 19-9 AssayBayer Healthcare, LLC · 2003
  3. Descendant · cites it directly · 0 of its ownK253528 Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc. · 2026
  4. Descendant · cites it directly · 0 of its ownK100375 Dimension Vista Loci CA 19-9 Flex Reagent Cartridge, and Dimension…Siemens Healthcare Diagnostics, Inc. · 2011

Most influential descendants

Showing 2 of 2 · See all
  • K253528Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc. · 2026 · cites it directly0descendants
  • K100375Dimension Vista Loci CA 19-9 Flex Reagent Cartridge, and Dimension LOCI7 Calibrator with…Siemens Healthcare Diagnostics, Inc. · 2011 · cites it directly0descendants
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K020566Fujirebio Diagnostics CA 19-9 RiaFujirebio Diagnostics,Inc.2002—
0 (this device)K031393Bayer Advia Centaur CA 19-9 AssayBayer Healthcare, LLC2003—
+1K253528Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc.2026—
+1K100375Dimension Vista Loci CA 19-9 Flex Reagent Cartridge, and Dimension LOCI7…Siemens Healthcare Diagnostics, Inc.2011—

Descendants by year cleared

Year clearedDevicesWith a Class I recall
201110
202610
FIG. 1 — Lineage of K031393, 2002–2026Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NIG device

  1. #1 K253528 Atellica IM CA 19-9 IIscore 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020566Fujirebio Diagnostics CA 19-9 Ria
Summary page 3
…em. 6. Substantial Equivalence The ADVIA Centaur CA 19-9 Assay is substantially equivalent to the Fujirebio Diagnostics CA 19-9 RIA predicate device cleared under K020566. Both products are based on the use of the same monoclonal antibody (Mab), 1116NS19-9, which is intended for use in the qua…
Fujirebio Diagnostics,Inc.2002—

Also cleared against K020566: 4 other devices →

What descends from it? (2)

2 later devices cite it directly; ranked here by how many devices descend from each in turn.

NumberDeviceApplicantDecidedDescendants
K253528Atellica IM CA 19-9 IISiemens Healthcare Diagnostics, Inc.20260
K100375Dimension Vista Loci CA 19-9 Flex Reagent Cartridge, and Dimension LOCI7 Calibrator with…Siemens Healthcare Diagnostics, Inc.20110

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Jul 26, 2024Class IIZ-3102-2024Under Investigation by firmPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across…
Jul 26, 2024Class IIZ-3103-2024Under Investigation by firmPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across…
Jul 26, 2024Class IIZ-3100-2024Under Investigation by firmPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across…
Jul 26, 2024Class IIZ-3101-2024Under Investigation by firmPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across…
Jul 26, 2024Class IIZ-3104-2024Under Investigation by firmPositive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across…
Oct 19, 2016Class IIZ-0612-2017Process designCA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.
Oct 19, 2016Class IIZ-0613-2017Process designCA 19-9 assay on the ADVIA Centaur CP does not meet the correlation to the CA 19-9 assay on the ADVIA Centaur/XP/XPT systems as stated in the Instructions for Use.
Nov 14, 2012Class IIZ-0508-2013Device DesignOperating range of the ADVIA Centaur systems is 18C to 30C (64.4F to 86.0F), a change in room temperature may increase or decrease assay results, depending on the specific assay and ADVIA Centaur system used.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • Of its descendants, K253528 (Siemens Healthcare Diagnostics, Inc., 2026) has gone furthest, with 0 descendants of its own.
  • 4 other devices also cite its predicate K020566.

How does its category compare?

All NIG devices

NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NIG

1in 2023

2011: 12019: 12020: 12023: 12026 (partial year): 1
2007high 1 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NIG

Too few NIG 510(k)s in any year since 2007 to give a median.

Recalls per year · NIG

2in 2016

2010: 12011: 12012: 12016: 2
2007high 2 (2016) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NIG

98in 2025

2007: 62008: 92009: 112010: 122011: 82012: 1,7852013: 1822014: 1172015: 292016: 692017: 982018: 582019: 1352020: 932021: 692022: 832023: 1292024: 1212025: 98
2007high 1,785 (2012) · low 6 (2007)2025

MAUDE reports naming product code NIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—09
20090—011
20100—112
20111—18
20120—11,785
20130—0182
20140—0117
20150—029
20160—269
20170—098
20180—058
20191—0135
20201—093
20210—069
20220—083
20231—0129
20240—0121
20250—098
20261—0—

What is it indicated for?

The ADVIA Centaur® CA 19-9 Assay is an in vitro immunoassay for the quantitative measurement of the CA 19-9 tumor-associated antigen, in human serum, using the ADVIA Centaur System. This assay is indicated for the serial measurement of CA 19-9 to aid in the management of patients diagnosed with cancers of the exocrine pancreas. The test is useful as an aid in monitoring of disease status in those patients having confirmed pancreatic cancer who have…

Read the full statement (129 words)

The ADVIA Centaur® CA 19-9 Assay is an in vitro immunoassay for the quantitative measurement of the CA 19-9 tumor-associated antigen, in human serum, using the ADVIA Centaur System. This assay is indicated for the serial measurement of CA 19-9 to aid in the management of patients diagnosed with cancers of the exocrine pancreas. The test is useful as an aid in monitoring of disease status in those patients having confirmed pancreatic cancer who have levels of serum CA 19-9 at some point in their disease process exceeding the median concentration determined for the apparently healthy cohort. CA 19-9 values must be interpreted in conjunction with all other clinical and laboratory data before a medical decision is determined. This assay is not intended for use on any other system.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K031393: Bayer Advia Centaur CA 19-9 Assay.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K031393

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026