510(k) Report

510(k) · Cleared

Recalled · Class II · 2019

GI Monitor and GI Monitor Calibrators on the Access Immunoassay Systems, Model 387687, 387688

K033038510(k) TraditionalCleared Dec 29, 200391-day reviewClass II21 CFR 866.6010Recalls 2

Data to Oct 25, 2026 (FDA decisions) Oct 12, 2026 (recalls) Sep 30, 2026 (MAUDE) refreshed weekly, last Nov 3, 2026

Beckman Coulter, Inc. · NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer

Cleared on Dec 29, 2003 as substantially equivalent to K020566 (Fujirebio Diagnostics,Inc., 2002) after a 91-day review. Its family tree reaches back 1 generation to 2002. It has been recalled 2 times.

Decided
Dec 29, 2003
Review
91 days
Class
Class II
Product code
NIG
Recalls
2
Predicates
1

Its longest line of descent runs back 1 generation to K020566 (2002). No later device cites it as a predicate yet.

K020566 · Fujirebio Diagnostics CA 19-9 Ria · Fujirebio Diagnostics,Inc. · 2002200020022004
  1. Root · 1 generation backK020566 Fujirebio Diagnostics CA 19-9 RiaFujirebio Diagnostics,Inc. · 2002
  2. This deviceK033038 GI Monitor and GI Monitor Calibrators on the Access Immunoassay…Beckman Coulter, Inc. · 2003
View FIG. 1 as a table

Line of descent and top descendants

GenerationNumberDeviceApplicantDecidedRecalls
−1K020566Fujirebio Diagnostics CA 19-9 RiaFujirebio Diagnostics,Inc.2002—
0 (this device)K033038GI Monitor and GI Monitor Calibrators on the Access Immunoassay Systems, Model…Beckman Coulter, Inc.2003—
FIG. 1 — Lineage of K033038, 2002–2003Source: FDA 510(k) summaries, read by 510(k) Report · How lineage is built

Get the monthly Device Review Tracker

Notable FDA decisions, review times and new Class I recalls. See an issue.

Sources: FDA 510(k) database (decision fields), the FDA summary (predicates and indications), FDA recall and enforcement reports, and MAUDE; processed by 510(k) Report. Verify on FDA.gov · Methodology and accuracy.

Planning a device in NIG? A 510(k) Report weighs predicate choices, recall history and review times for a new device like this one.

Top candidate predicates for a new NIG device

  1. #1 K253528 Atellica IM CA 19-9 IIscore 12.1 · 2 generations · no Class I recall
  2. #2 Held for the reportin the report
  3. #3 Held for the reportin the report
Get a 510(k) Report

What was it cleared against? (1)

Devices the submitter claimed this one is substantially equivalent to (*), and reference devices cited for a specific test or feature (†), as written in the FDA summary.

TypeNumberDeviceApplicantDecidedRecalls
*K020566Fujirebio Diagnostics CA 19-9 Ria
Summary page 1
…Cancer Predicate Device Fujirebio Diagnostics CA 19-9 RIA Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355 510(k) Number: k020566 Device Description The Access GI Monitor reagents, calibrators, and the Access Immunoassay Analyzers (Access, Access 2,…
Fujirebio Diagnostics,Inc.2002—

Also cleared against K020566: 4 other devices →

What descends from it?

No later submission names this one as a predicate yet.

Has anything in its family been recalled? (0 Class I)

Recalls of this device

DateClassRecallRoot causeReason
Apr 30, 2019Class IIZ-2241-2020OtherThe recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687…
May 9, 2017Class IIZ-2638-2017Device DesignThe Access immunoassays are susceptible to biotin interference.

No ancestor or descendant has a Class I recall.

What does this lineage say?

  • Its one ancestor has no recall.
  • 4 other devices also cite its predicate K020566.

How does its category compare?

All NIG devices

NIG System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer: FDA decisions, review times, recalls and adverse event reports over 20 years.

FDA decisions per year · NIG

1in 2023

2011: 12019: 12020: 12023: 12026 (partial year): 1
2007high 1 (2011) · low 0 (2007)2026

2026 to Oct 25, outlined.

Median 510(k) review, days · NIG

Too few NIG 510(k)s in any year since 2007 to give a median.

Recalls per year · NIG

2in 2016

2010: 12011: 12012: 12016: 2
2007high 2 (2016) · low 0 (2007)2026

Class I (0 since 2007) Class II/III not classified

Adverse event reports per year · NIG

98in 2025

2007: 62008: 92009: 112010: 122011: 82012: 1,7852013: 1822014: 1172015: 292016: 692017: 982018: 582019: 1352020: 932021: 692022: 832023: 1292024: 1212025: 98
2007high 1,785 (2012) · low 6 (2007)2025

MAUDE reports naming product code NIG, not matched to any one device; reports are not rates.

View these four charts as a table
YearDecisionsMedian 510(k) review, daysRecalls (Class I)MAUDE reports
20070—06
20080—09
20090—011
20100—112
20111—18
20120—11,785
20130—0182
20140—0117
20150—029
20160—269
20170—098
20180—058
20191—0135
20201—093
20210—069
20220—083
20231—0129
20240—0121
20250—098
20261—0—

What is it indicated for?

The Access GI Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 19-9 antigen levels in human serum and plasma using the Access Immunoassay Systems. This device is indicated for use in the measurement of CA 19-9 antigen to aid in the management of pancreatic cancer patients. The test is useful as an aid in monitoring of disease status in those patients having confirmed pancreatic cancer whose…

Read the full statement (114 words)

The Access GI Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 19-9 antigen levels in human serum and plasma using the Access Immunoassay Systems. This device is indicated for use in the measurement of CA 19-9 antigen to aid in the management of pancreatic cancer patients. The test is useful as an aid in monitoring of disease status in those patients having confirmed pancreatic cancer whose serum CA 19-9 antigen levels exceed 10 U/mL, the cut-off value for individuals who are Lewis blood group antigen negative. Serial testing for patient CA 19-9 antigen concentrations should be used in conjunction with other clinical methods used for monitoring pancreatic cancer.

Indications for use as stated in the FDA summary.

Cite this page

510(k) Report. “K033038: GI Monitor and GI Monitor Calibrators on the Access Immunoassay Systems, Model 387687, 387688.” Lineage record from FDA 510(k) data and the FDA summary. Updated Nov 3, 2026. Accessed Nov 6, 2026. https://510kreport.com/device/K033038

Report an extraction error

Tell us what is wrong; we check it against the FDA summary.

Are you the manufacturer?

Send a correction or an update to this page, for example a newer model or a public statement about a recall.

Record history
  • FDA decision record loaded May 2, 2026
  • Summary read Oct 13, 2026 (extraction version 9eaa42aa01ed)
  • Lineage and recalls computed Nov 3, 2026
  • Page last updated Nov 3, 2026